Display, operation and front side settings........7 Back side settings and terminal connections......9 Further operating conditions ............ 10 Conductivity measuring cells for standard transmitters ... 11 Product range HEYL-NEOMERIS and contact details .... 14 Gebr. Heyl Vertriebsgesellschaft für innovative Wasseraufbereitung mbH...
D200 Description 1. Description Device for the measurement of the electrical conductivity of aqueous solutions using a two-electrode conductivity cell. Designed for wall mounting installation. Applications: Demineralisation, reverse osmosis, desalination, cooling water recirc., phase separation Operation at 230 V AC or 24 V DC.
Page 4
D200 Technical Data Gebr. Heyl Vertriebsgesellschaft für innovative Wasseraufbereitung mbH...
Page 5
D200 Technical Data Gebr. Heyl Vertriebsgesellschaft für innovative Wasseraufbereitung mbH...
D200 Display, operation and front side settings 4. Display, operation and front side settings Display: Conductivity in µS/cm Display top line: Display bottom line: Without temperature compensation: Limit values 1 and 2 With connected Pt100 sensor and enabled Sensor temperature or...
Page 8
D200 Display, operation and front side settings Relay settings: The relay output modes of operation is switchable via DIP switch 6 at the back side (see chapter Back side settings and terminal connections). DIP switch position off (bottom): Relays drop at limit value exceedance...
D200 Back side settings and terminal connections 5. Back side settings and terminal connections Temp. range Temp. zero point 21 22 Terminal connections: Shield Sensor cable shield Conductivity measuring cell Conductivity measuring cell Pt100 Pt100 0-10 V Output 0 – 10 V 4-20 mA Output 4 –...
D200 Further operating conditions Potentiometer settings (all are pre-adjusted!): Zero point conductivity measurement Range conductivity measurement Setting for temperature measuring at 0 ° C (Pt 100 = 100.0 Ohm) Setting for range, e.g. at 95 ° C (Pt100 = 136.8 Ohm)
D200 Conductivity measuring cells for standard transmitters 7. Conductivity measuring cells for standard transmitters Conductivity measuring cells, ½“ and ¾“: Cell constant K: 0.1 or 1.0 (±5%), depending on type • Electrode body material: PP (M101, M110, M201, M210) oder PVDF •...
Page 12
D200 Conductivity measuring cells for standard transmitters Pt100 with ¾ M210 / Pt100 ¾ M210 880211 ¾“ Screw-in thread / Pt100 and magnetic valve plug with Screw-in ½ M401 880262 ½“ thread magnetic valve plug ½ M401 Pt100 with ½ M401 880263 ½“...
Page 13
D200 Conductivity measuring cells for standard transmitters Combinations overview: Conductivity measuring cells and transmitters Cell Measuring Reference Measuring Reference constant measuring cell Instru cell Instru (± 5 %) range ments ments with Pt100 with Pt100 ½ M101 880206 ¾ M201 / Pt100...
Product range HEYL-NEOMERIS and contact details and reacts already at thin films. Does not depend on conductivity, consistency or draught in the environment. 8. Product range HEYL-NEOMERIS and contact details TESTOMAT® Instruments, TITROMAT® Instruments ONLINE ANALYSIS Accessories and spares TESTOMAT®-/ TITROMAT®-Instuments INSTRUMENTS Reagents TESTOMAT®-/TITROMAT®-Instruments, Reagents BOB Instruments...