Performance Verification Test Record; Items Tested; Equipment Used; Test Record - LeCroy DA18000 Instruction Manual

Differential amplifier
Table of Contents

Advertisement

Appendix A
A

PERFORMANCE VERIFICATION TEST RECORD

This record can be used to record the results of measurements made during the perfor-
mance verification of the DA18000 Differential Amplifier.
Photocopy this page and record the results on the copy. File the completed record as
required by applicable internal quality procedures.
The section in the test record corresponds to the parameters tested in the performance veri-
fication procedure. The numbers preceding the individual data records correspond to the
steps in the procedure that require the recording of data. Results to be recorded in the col-
umn labeled "Test Result" are the actual specification limit check. The test limits are included
in all of these steps. Other measurements and the results of intermediate calculations that
support the limit check are to be recorded in the column labeled "Intermediate Results".
Permission is granted to reproduce these pages for the purpose of recording test results.

ITEMS TESTED

DA18000

EQUIPMENT USED

OSCILLOSCOPE
OSCILLOSCOPE
ST-20 TDR
ST-20 TDR

TEST RECORD

Step
Description
1. Rise Time
1-m Source rise time, t
1-o
System rise time, t
1-q
Probe Rise Time, t
2. Voltage Gain Accuracy
2-d
Scope Accuracy
2-i
System Accuracy
2-j
Probe Accuracy (< 2%)
DA18000-OM-E Rev A
Item
Serial Number
Model
Serial Number
Intermediate data
______________ ps
source
______________ ps
system
(< 19 ps)
r
______________V
______________ V
§ § §
Appendix A
Calibration Due
Date
Test Result
_________ ps
_________%
A-1