MAT 4-807 User Manual

Mat 4-807 cordless drill system user’s manual

Advertisement

Quick Links

CORDLESS SYSTEM
Easy to use
Ergonomic design
Simple, compact battery packs with high capacity
Small, robust drill with an extensive range of attachments
Steam-sterilisable
CORDLESS DRILL SYSTEM
User's Manual No. 6.1 ( GB )
EDICAL ADVANCED TECHNOLOGY
M
1253

Advertisement

Table of Contents
loading

Summary of Contents for MAT 4-807

  • Page 1 EDICAL ADVANCED TECHNOLOGY 1253 CORDLESS DRILL SYSTEM User’s Manual No. 6.1 ( GB ) CORDLESS SYSTEM Easy to use Ergonomic design Simple, compact battery packs with high capacity Small, robust drill with an extensive range of attachments Steam-sterilisable...
  • Page 2: Intended Use

    Intended use: The Drill 4-807 is used in bone surgery for drilling and tapping, inserting and removing screws and placing Kirschner wires. The drill has a cannulated drive shaft, so that drill wires can be briefly inserted and fed through the drill (diameter of cannulation: 3.2mm) for safe working.
  • Page 3: Maintenance

    Sterilisation: The Cordless Drill 4-807 with end cap and funnel can be sterilised at up to 2 bar at 134°C in the steam steriliser. It is advisable to place the equipment on a cloth or wrap it in a cloth to prevent external or internal damage. To ensure proper sterilisation, the end cap must be screwed onto the Cordless Drill 4-807.
  • Page 4: Transport And Storage Conditions

    To prevent damage! The Cordless Drill 4-807 and attachments must be transported in such a way that no damage can occur. This particularly means en- suring suitable packaging. The same attention must be paid to correct storage. This device must only be kept in a dry, closed place! Safety instruction: Remove the battery from the Cordless Drill 1.
  • Page 5: Technical Data

    It is hereby certified that the Cordless Drill 4-807/ 4-808 has been tested and is interference-suppressed in compliance with the EMC provisions of EN60601-1-2. This Cordless Drill 4-807 / 4-808 has been tested by our notified body for a Class II.a product with internal or external energy source, as per 93/42 EEC Annex II.3...
  • Page 6: Key To The Symbols

    Disposal: MAT GmbH & Co. KG will dispose of the Cordless Drill 4-807 and atttachments (without any environmental contamination)in accordance with the provisions of MED. UMW MAT GmbH & Co. KG will deal with disposal! Key to the symbols: Symbol 1 Observe instructions in User´s Manual!
  • Page 7 OVERVIEW: 6. Clockwise rotation 7. Main switch and reverse rotation 1. Cordless Drill 2. Battery Compatment 3. Rechargeable Battery 4. End cap 5. Chuck with Drill Bit 6. Clockwise Rotation 7. Main switch and reverse Rotation Important To obtain full power from the drill during the surgical operation, always use freshly charged batteries. Preparation Step 1.
  • Page 8 Overview of attachments: NOTE: All the clutches and chucks for the Cordless Drill system are mounted in the same way! • Step 1. Slide the clutch collar (1) in the direction indicated by the arrow • Slide the chuck (2) onto the shaft (3) as far as it will go •...
  • Page 9 Standard rechargeable battery 3.) REF. 4 – 811 Funnel for inserting Rechargeable Battery 4-809 or 4-810 into Cordless Drill 4-807 (to avoid contamination of the Cordless Drill. If it is not used, there may be a loss of sterility) 4.) REF. 4 – 814...
  • Page 10: Care And Maintenance

    Cleaning, disinfection and sterilisation of attachments: (except Rechargeable Battery 4-809/4-810 and Desktop Charger 4-814/4-816, because these products are not categorised as medical devices ) CLEANING AND DISINFECTION: CARE AND MAINTENANCE Care and maintenance are vital to keep the attachments in good condition and fully functional! Causes of corrosion with stainless steel used for instruments.
  • Page 11 1. MECHANICAL CLEANING: Mechanical cleaning is the preferred method. Strictly follow the instructions supplied by the manufacturer of the cleaning machine. Prepare the instruments and attachments for mechanical cleaning. Multipart instruments and attachments that can be dismantled must be taken apart or opened for cleaning. All tips and delicate parts should be cleaned manually. Do not overfill wire trays; place heavy parts on the bottom of the tray.
  • Page 12: Fault Rectification

    Only store in the original packaging, in a dry, closed room. Never sterilise the BATTERY 4-809/4-810 or the Charger 4-814/4-816. Only use compatible BATTERIES 4-809 and 4-810 from MAT GmbH&Co.KG. Combination with other makes is forbidden. Do not immerse Charger 4-814/4-816 or BATTERY 4-809/4-810 in liquid.
  • Page 13: Safety Instruction

    (3) of the Charger 4-814/4-816 against moisture.Never immerse in liquid or clean under flowing water! Repair: MAT GmbH & Co. KG only accepts responsibility for the safety, reliability and performance of the Charger 4-814/4-816 and BATTERY 4-809/4-810if: • the Charger 4-814/4-816 and BATTERY 4-809/4-810 are used in compliance with the instructions in the User’s Manual.
  • Page 14 Technical Data: Type of equipment: Chargers 4-814 and 4-816 Rated voltage 240 V and 115 V Rated frepuency 50 HZ Rated umput 14 VA Rated output voltage 12 V Rated output current 1100 mA max. Protection class Degree of protection, as per DIN EN 529 IPX0 Function The Chargers 4-814 and 4-816 are controlled and monitored by a microcontroller.
  • Page 15 Notice:...
  • Page 16 MAT GmbH & Co.KG Friedrich-Wöhler Straße 10 78576 Emmingen-Liptingen, Germany Phone: +49 (0)7465 909 188 · Fax : +49 (0)7465 909 024 EDICAL ADVANCED TECHNOLOGY www.mat-medical.com · info@mat-medical.com Cordless Drill System User’s Manual No. 6 ( GB ) © MAT Revision 06.04.2010...