Replacing The Battery - Invacare RLS6 Assembly, Installation And Operating Instructions

Invacare reliant scale assembly, installation and operating instructions
Table of Contents

Advertisement

NOTE: The patientʹs weight will continue to be seen in the display window.
NOTE: The display will turn off automatically after a two minute period of non‐use [no changes in weight exceeding 
five pounds (two kilograms)]. You can not adjust the time delay for automatic shut off.
NOTE: After the display has turned off, the weight may be recalled by pressing the ON/OFF button.
NOTE: The unit can be turned off by pressing the ON/OFF button a second time. 

Replacing the Battery

NOTE: For this procedure, refer to FIGURE 4.
NOTE: The scale is powered by a nine volt alkaline battery that should provide approximately 1500 readings before 
needing replacement.
When battery replacement is needed, LO BAT will appear on the display. Perform the following:
1. Slide the battery door open in the direction of the arrow.
2. Remove existing battery.
3. Install new battery.
4. Reinstall the battery door.
Calibrate the Reliant Scale
NOTE: For this procedure, refer to FIGURE 5 on page 7.
NOTE: The Reliant Scale will be pre‐calibrated at the factory with the load cell. Should it be necessary to re‐calibrate 
the scale, follow the instructions outlined below.
1. The patient and the sling MUST be removed from the scale to properly calibrate the Reliant Scale. Refer 
to Operation in the Patient Lift Owner's Manual for further instructions.
2. For sling detachment instructions, refer to Lifting the Patient in the Patient Lift Owner's Manual.
3. Turn the scale display power on by pressing the ON/OFF key.
4. Rotate the Swivel Bar until the back of the scale display is visible.
5. Insert the tip of a paper clip or similar tool into the small hole on the back of the scale display and press 
the CAL button. 
NOTE: Once the CAL button is pressed, "CAL1" will be seen in the display window.
6. Press the UNITS key. The display reads CAL2.
Reliant Scale RLS6
9V Battery
Battery
Door
Arrow
FIGURE 4 Replacing the Battery
Assembled Scale
6
Part No 1114832

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents