Conditioning the Battery

NOTE: The definition of conditioning the battery is simply charging the battery for a full 
six hour cycle before the initial use of the product.
1. Disconnect the battery from the compressor.
2. Connect the power cable (AC) or auto adapter cable (DC) to the battery.
3. Charge the battery for a full six hour cycle.
The orange light will indicate if the battery is properly charging. Battery charging 
to full capacity is dependent upon many factors. Conditioning the battery requires 
at least 6 hours. Failure to condition the battery prior to its first use may result in a 
reduced battery life expectancy. To assure maximum life performance of the battery 
after conditioning, leave the battery connected to a power source when not in use. 
The battery operation time depends upon the charge condition of the battery, the 
temperature, the storage time of the battery prior to use, the amount of medication 
used for treatment, and the number of charge/discharge cycles. The manufacturer 
of the battery recommends the battery be conditioned every one or two months.

Charging the Battery

NOTE: For this procedure, refer to FIGURE 1.
1. Press in the release button on the front of the battery pack and separate the 
battery from the compressor (Detail "A"). 
2. Connect the power cable (AC) to the AC adapter and cable provided with your 
STRATOS Portable Plus Compressor (Detail "B").
3. Connect the AC adapter and cable to the battery (Detail "A").
4. Plug the power cable into the appropriate power source.
The ORANGE light will turn on indicating that the charging cycle has begun. The 
battery will reach full charge after at least six hours. Completion of the charge cycle 
is indicated by a GREEN battery light. Full charge will provide up to ninety 
minutes of operation.
NOTE: In order to ensure maximum performance of the battery, keep the battery plugged 
in and charging when not in use. This will not overcharge or damage the battery.
NOTE: The battery must be plugged into a power source for the battery to charge. The 
battery will not charge when it is attached to the compressor nebulizer system.
NOTE: Never leave the battery connected to either power cable when the cable is 
disconnected from the power source. This will cause the battery to discharge.
Hook
Battery
STRATOS™Portable Plus Battery Pack
DETAIL "A" -
BATTERY
Release
Button
Connect
Adapter
HERE
FIGURE 1 Charging the Battery
2
DETAIL "B" -
ADAPTER
AC Adapter
and Cable
To Battery
Power
Cable
To Outlet
Part No 1150754

Advertisement

loading

This manual is also suitable for:

Stratos aerosol irc1731Stratos portable plus