Conditions
Ambient temperature: 20/30
Load: Unloaded
Distribution data
Temperature of the cycle in each area (SV=-30
220V50Hz
Middle of
Maximum Minimum
-27.8
-29.9
-28.1
-31.3
-28.1
-31.0
-28.1
-31.4
-28.9
-30.6
-29.2
-32.5
-28.9
-32.0
-28.9
-31.9
-28.8
-32.1
-29.6
-31.1
-29.7
-32.6
-28.9
-30.6
-29.5
-31.5
-28.6
-30.3
-29.9
-32.0
Average
-
-
-30.1
220V50Hz
Middle of
Maximum Minimum
-26.2
-28.1
-27.2
-26.7
-29.4
-28.1
-26.7
-29.1
-27.9
-26.8
-29.5
-28.2
-27.7
-29.0
-28.4
-28.2
-31.0
-29.6
-28.0
-30.2
-29.1
-27.8
-30.3
-29.1
-27.6
-30.7
-29.2
-28.5
-30.0
-29.3
-28.5
-31.6
-30.1
-27.6
-29.4
-28.5
-28.5
-31.4
-30.0
-27.3
-29.1
-28.2
-28.8
-31.9
-30.4
Average
-
-
-28.9
Note:This data does not represent a guarantee of product performance.
air temperature)
Ambient temperature 20
Differential Maximum Minimum
cycle
-28.9
±1.1
-27.8
-29.8
-29.7
±1.6
-28.3
-31.2
-29.6
±1.5
-28.3
-31.0
-29.8
±1.7
-28.3
-31.3
-29.8
±0.9
-29.1
-30.6
-30.9
±1.7
-29.4
-32.4
-30.5
±1.6
-29.1
-32.0
-30.4
±1.5
-29.1
-31.9
-30.5
±1.7
-28.9
-32.2
-30.4
±0.8
-29.6
-31.0
-31.2
±1.5
-29.7
-32.7
-29.8
±0.9
-28.9
-30.7
-30.5
±1.0
-29.5
-31.7
-29.5
±0.9
-28.6
-30.4
-31.0
±1.1
-30.0
-32.6
-
-
Ambient temperature 30
Differential Maximum Minimum
cycle
±1.0
-26.2
-28.0
±1.4
-26.7
-29.4
±1.2
-26.8
-29.1
±1.4
-26.8
-29.5
±0.7
-27.6
-29.0
±1.4
-28.1
-31.0
±1.1
-27.9
-30.3
±1.3
-27.8
-30.4
±1.6
-27.6
-30.8
±0.8
-28.5
-30.0
±1.6
-28.5
-31.6
±0.9
-27.5
-29.3
±1.5
-28.4
-31.4
±0.9
-27.2
-29.1
±1.6
-28.8
-31.8
-
-
Unit:
230V50Hz
Middle of
Differential
cycle
-28.8
±1.0
-29.8
±1.5
-29.7
±1.4
-29.8
±1.5
-29.9
±0.8
-30.9
±1.5
-30.6
±1.5
-30.5
±1.4
-30.6
±1.7
-30.3
±0.7
-31.2
±1.5
-29.8
±0.9
-30.6
±1.1
-29.5
±0.9
-31.3
±1.3
-
-30.2
-
Unit:
230V50Hz
Middle of
Differential
cycle
-27.1
±0.9
-28.1
±1.4
-28.0
±1.2
-28.2
±1.4
-28.3
±0.7
-29.6
±1.5
-29.1
±1.2
-29.1
±1.3
-29.2
±1.6
-29.3
±0.8
-30.1
±1.6
-28.4
±0.9
-29.9
±1.5
-28.2
±1.0
-30.3
±1.5
-
-28.8
-
29
Amount of power consumption
Amount of power consumption when driving at cycle
Ambient temp 20
50Hz
220V
1.47
230V
1.50
1.55
240V
Note:This data does not represent a guarantee
of product performance.
(SV=-30
Unit kWh/day
Ambient temp 30
50Hz
1.87
1.91
2.08