Fully fuzzy auto digital wrist blood pressure monitor
Hide thumbs Also See for GP-6220:

Advertisement

Fully Fuzzy Auto Digital Wrist
Blood Pressure Monitor
tensio control
GB
INSTRUCTION MANUAL
Model GP-6220
Manufactured by:
Geratherm Medical AG
Fahrenheitstraße 1
98716 Geschwenda
Germany
Phone: +49 (0) 36205 98-0
Fax: +49 (0) 36205 98-116
E-Mail: info@geratherm.com
www.geratherm.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GP-6220 and is the answer not in the manual?

Questions and answers

Summary of Contents for GERATHERM GP-6220

  • Page 1 Fully Fuzzy Auto Digital Wrist Blood Pressure Monitor tensio control INSTRUCTION MANUAL Model GP-6220 Manufactured by: Geratherm Medical AG Fahrenheitstraße 1 98716 Geschwenda Germany Phone: +49 (0) 36205 98-0 Fax: +49 (0) 36205 98-116 E-Mail: info@geratherm.com www.geratherm.com...
  • Page 2: Warranty

    CAUTION 1. Consult your physician to verify your blood pressure range before using this device at home. 2. If the cuff causes any discomfort during measurement, press the power key to turn off the device immediately. 3. Pull off the velcro strap to detach the cuff if the pressure is over 300 mm Hg and the cuff does not deflate automatically.
  • Page 3 INFORMATION YOU SHOULD KNOW BEFORE OPERATING THE DEVICE What is blood pressure? A force is created by the heart as the ventricle forcibly ejects blood into the blood vessels and through the vascular system. Another force is created by the arteries as they resist the blood flow. Blood pressure is the result of these two forces.
  • Page 4 Fluctuation and variation in blood pressure The following chart shows possible blood pressure fluctuations during a 24-hour period. Hypertension Boundary Normal The following factors will Lower influence blood pressure measurement results and cause variations: Bathing, Breathing, Conversation, Meals, Smoking, Exercise, Drinking Alcohol, Mo- ving, Thoughts, Mental tension, Temperature...
  • Page 5: About The Device

    ABOUT THE DEVICE Memory Botton LCD display Battery Cover Power Button The cuff is designed to fit wrist sizes between 13.6 and 19.5 cm (5.35 to 7.68 inch). 4 - GB...
  • Page 6 ACHTUNG EXPLANATION OF DISPLAYS Date Memory Systolic Battery Indicator Diastolic Pulse Indicator Pulse Rate IPD (Irregular Pulse Detection) ACHTUNG PREPARATION FOR USE Installing/Replacing Batteries 1. Press lightly on the arrow and slide the battery cover in the direction of the arrow. 2.
  • Page 7 Attaching the Pressure Cuff 1. Free the left wrist of any clothing etc. and wrap the cuff directly around it. The display should be positioned on the palm side of the wrist. 2. Fasten the cuff. Make sure it is not too tight. The cuff‘s upper edge should be approxima- tely 1 cm (0.39 inch) from the palm line.
  • Page 8: How To Operate The Device

    ACHTUNG HOW TO OPERATE THE DEVICE 1. Setting the Clock Enter the clock setting mode by pressing and holding the MEMORY button for over three seconds. The display will show a flashing year. Press the POWER button to adjust the year when it is flashing. Then click the MEMORY button again, a „month „...
  • Page 9 3. Recalling Stored Values 3.1 Press and release the MEMORY key to recall the last reading. (Example: no. 35 in the memory). The display will show the stored pressure and pulse rate in alternation. 3.2 Press the MEMORY button to go to previous stored values. 4.
  • Page 10: Care And Maintenance

    ACHTUNG BETRIEB DES GERÄTES ACHTUNG CARE AND MAINTENANCE Do not drop Do not modify or disas- the device. semble the body or the wrist cuff. Use a cloth moistened Do not twist with water or neutral the wrist cuff. detergent to clean the body of the device and thenwipe it dry.
  • Page 11 ACHTUNG BETRIEB DES GERÄTES ACHTUNG ERROR MESSAGES Message Correction The pressure measured was lower than 20 mm Hg. Please measure again. The pressure measured was higher than 300 mm Hg. Please measure again. Pumping error. Please check cuff and try again. The pressure can not be measured due to signal noise.
  • Page 12: Specifications

    ACHTUNG SPECIFICATIONS Display System Liquid Crystal Display Measuring Method Oscillometric Power Source 2 alkaline “AAA“ type batteries (1.5 V) Measuring Range 20 - 300 mm Hg; 40-200 beats/minute Accuracy ± 3 mm Hg ± 5% pulse rate Automatic with integrated Inflation Pressure electrical pump Exhaust...
  • Page 13 Specifications are subject to change for improvement without notice. Calibra- tion is recommended every two years. 12 - GB...