while 'browsing' through menus). Please see the ES3 Calibration Guide for more
detailed information regarding headset management and calibration.
Specification of headset type should only be done during calibration (i.e., when
selecting which type of earphone is being calibrated). Specification of the Primary
headset type is done via the menu sequence ⌐S
⌐G
⌐P
ETUP
ENERAL
ERFORM
C
(password) ⌐C
P
⌐H
T
{TDH-39/EAR-5A}.
ALIBRATION
ALIBRATE
RIMARY
EADSET
YPE
Specification of the Secondary headset type is done via the menu sequence
⌐S
⌐G
⌐P
C
(password) ⌐C
S
ETUP
ENERAL
ERFORM
ALIBRATION
ALIBRATE
ECONDARY
⌐H
T
{TDH-39/EAR-5A/None}. If no secondary headset is equipped,
EADSET
YPE
'None' should be selected.

CALIBRATION

Micro Audiometrics recommends that audiometers be calibrated annually. About
one month before the annual calibration due date, the ES3 will begin displaying a
reminder at power up that the due date is approaching. This reminder will be
cleared once the instrument is calibrated. If the calibration due date passes and
the instrument has not been calibrated, a caution will be displayed regarding
potential test data invalidation.
Note: If an ES3S audiometer is equipped with the Dual Headset Option, separate
calibration due reminders will occur.
The last calibration date and calibration due dates may be viewed at any time via
the menu sequence ⌐S
⌐G
⌐C
D
. Please see the ES3
ETUP
ENERAL
ALIBRATION
ATES
Calibration Guide for more detailed information regarding headset management
and calibration.