B-K Medical Type 8661 User Manual

Endovaginal endfire transducer

Advertisement

Quick Links

Endovaginal Endfire Transducer
Type 8661
User Guide
English BB0613-G
August 2006

Advertisement

Table of Contents
loading

Summary of Contents for B-K Medical Type 8661

  • Page 1 Endovaginal Endfire Transducer Type 8661 User Guide English BB0613-G August 2006...
  • Page 2 Your opinions are important to us. You are of course always welcome to contact us via your B-K Medical representative or by contacting us directly. © 2006 B-K Medical...
  • Page 3: Table Of Contents

    Endovaginal Endfire Transducer Type 8661 Introduction Scanning Plane General Information Caring for the Transducer Cleaning and Disinfection Starting Scanning Connecting the Transducer Changing Frequency Using a Transducer Cover Using the Transducer Control Button Changing Orientation Endovaginal Scanning Puncture Facilities Puncture Attachment UA1242...
  • Page 5: Introduction

    Endovaginal Endfire Transducer Type 8661 Endovaginal Endfire Transducer Type 8661 Introduction This is the user guide for Endovaginal Endfire Transducer Type 8661 and must be used together with Transducer Care, Cleaning & Safety which contains important safety informa- tion. 8661 is an endovaginal endfire transducer suitable for gynecological examinations.
  • Page 6: Caring For The Transducer

    Transducer Care, Cleaning & Safety. Cleaning and Disinfection To ensure the best results when using B-K Medical equipment, it is important to maintain a strict regular cleaning routine. Full details of cleaning and disinfection procedures can be found in the Transducer Care, Cleaning &...
  • Page 7: Connecting The Transducer

    8661 • Starting Scanning Connecting the Transducer WARNING Keep all plugs and sockets absolutely dry at all times. The transducer is connected to the scanner using the array Transducer Socket on the scan- ner. To connect, the transducer plug’s locking lever should first be in a horizontal position. Align the plug to the scanner socket and insert securely.
  • Page 8: Endovaginal Scanning

    Puncture Facilities Puncture and biopsy is possible with Type 8661. The puncture attachment is illustrated in the following pages with a brief description of its use and operating instructions. The use of probe sheaths during invasive applications is recommended.
  • Page 9: Performing Puncture And Biopsy

    Caution Ensure that the tip of the needle guide lies close to the front end of the probe, if not the needle guide may be damaged and require realignment by B-K Medical technicians. Caution Ensure UA1242 is positioned correctly.
  • Page 10: Cleaning After Puncture And Biopsy

    8661 • Performing Puncture and Biopsy Press the scanner Puncture or Biopsy control button to superimpose a puncture line on the scan image. If more than one puncture line is available, refer to the applicable scanner user guide for instructions on how to change which one appears. Move the transducer until the puncture line transects the target.
  • Page 11: Disposal

    8661 • Disposal Disposal When the transducer is scrapped at the end of its life, national rules for the relevant mate- rial in each individual land must be followed. Within the EU, when you discard the trans- ducer, you must send it to appropriate facilities for recovery and recycling. See the applicable scanner user guide for further details.