Product Description - Probasics Zzz-PAP Owner's Manual

Integrated heated humidifier

Advertisement

5.
Connect the other end of the flexible tubing to your mask.
6.
Adjust the heater temperature to the desired setting by the
control knob. The 1-6 setting corresponds to the temperature
as follows:
1 = approximately 86°F
3 = approximately 109°F
5 = approximately 124°F
The system is ready for use. Consult the CPAP user's manual for operating instructions.
Danger: In order to avoid the destruction or damage to the CPAP equipment or even the hazard of
electric shock, when the water chamber is connected to the outlet port of CPAP, DO NOT fill
water from the outlet port of the water chamber. Users should not overfill the water chamber.
Always ensure the water level is no higher than the marking for maximum volume.
5
Cleaning
5.1
Heater
Clean the heated humidifier base by wiping it with a damp cloth. Allow the heater base to
air dry before connecting the power cord.
Inspect the heater and power cord for any damage and replace it if necessary.
5.2
Water Chamber
1.
Power off the device and allow the heater and water to cool.
2.
Disconnect the flexible tubing from the water chamber. Press down on the water chamber
and slide it out of the heater. Empty the remaining water.
3.
Use the extraction tool to remove the chamber base. Match the indent
on the extraction tool to the remark on the water chamber. Use the
dotted line on the tool as the pivot and flip it over to remove the base.
4.
Use a mild detergent to wash all the chamber parts. Rinse all the parts
with clear water and allow them to air dry. The components are also
dishwasher safe.
5.
All items of the chamber are subject to normal wear and tear and may eventually be
replaced. Replace the chamber assembly as necessary.
2 = approximately 102°F
4 = approximately 117°F
6 = approximately 140°F
7
2

Product description

Baffler
Flow Direction
Water Chamber
Removable Base Plate
Locking Tab
Fill Line
Extraction Tool
Heater Plate
Switch
Heater Plate
Power ON Indicator
Water Chamber:
The water chamber contains the water for humidification. Its removable base
plate allows itself ease for cleaning.
Symbol indicates the maximum level of water that water chamber can contain.
Fill Line:
Flow Direction:
Symbol Indicates the air flow direction
Baffler:
The baffler can prevent water spillage due to improper shaking.
Open the water chamber for cleaning.
Extraction Tool:
Heater Plate:
Warms the water in the water chamber.
Heater Plate Switch:
When the switch is pressed against the water chamber, the heater plate will
start to work.
Control Knob:
This control knob, graduated from 1 to 6, is used to adjust the heating
temperature of the heater plate from 86°F to 140°F.
When lit, this indicates the humidifier is powered on and ready for operation.
Power On Indicator
(Green LED):
Heater On Indicator
When lit, this indicates the heater base is warming the chamber.
(Yellow LED):
Inlet Port with Silicon
Connect to the outlet port of the CPAP device.
Connector:
Connect the flexible tubing here.
Outlet port:
AC Power Cord:
Connect to an AC outlet providing main power to the heated humidifier.
Power Socket (Inlet):
Connect the AC power cord here.
Connect the heated humidifier and the CPAP device.
Power Jumper Cord:
Power Socket (Outlet):
Connect the power jumper cord here.
Power Switch:
Power the Zzz-PAP heated humidifier ON or OFF.
Silicon Connector
Outlet Port
Inlet Port
Power Socket (Outlet)
Heater Platform
Power Switch
Power Cord
Power Socket (Inlet)
Heater ON Indicator
Control Knob
4
Power
Jumper Cord