Niked Medical Mark 5 Nuvo M5C5 User Manual

Oxygen concentrator

Advertisement

Quick Links

USER'S GUIDE
Mark 5 Nuvo
M5C5

OXYGEN CONCENTRATOR

[Original language is English ]
Federal Law (US) restricts this device to sale
by, or on the order of, a licensed physician.
This oxygen concentrator should be used
only under the supervision of a licensed
physician.
GLOSSARY OF SYMBOLS..........................................1
GENERAL SAFETY GUIDELINES ............................2
I. DESCRIPTION ...........................................................2
I.1.Front panel (Fig. I. 1)............................................. 3
I.2.Rear panel (Fig. I. 2)....................................................3
II. STARTING-UP / INSTALLATION .........................3
II.1.Use in direct oxygen therapy......................................3
III.CLEANING-MAINTENANCE ...............................4
III.1.Cleaning ....................................................................4
III.2.Everyday disinfection....................................................4
III.3.Maintenance..............................................................4
: ON (power switched on)
: Off (power switched off).
: Type B device
: Class II protection
: Do not expose to open  ames.
2010-2224F Rev C
CONTENTS
IV.
USEFUL INFORMATION................................... 4
IV.1. Accessories and spare parts..................................... 4
patient ..................................................................... 4
IV.3. Operating principles ................................................ 4
IV.4. Alarms - Safety devices........................................... 5
IV.5. Indicator light function .......................................... 5
IV.6. Technical characteristics.......................................... 5
IV.7. Standards ................................................................. 6
IV.8. Method for disposing of waste ................................ 6
IV.9. Method for disposing of device............................... 6

GLOSSARY OF SYMBOLS

: Do not use oil or grease.
: Technical information.
: Consult the accompanying documents.
: Keep in the vertical position.
: Fragile - handle with care.
May 2011
Danger: Do not smoke when using oxygen or when
near this device.
1

Advertisement

Table of Contents
loading

Summary of Contents for Niked Medical Mark 5 Nuvo M5C5

  • Page 1: Table Of Contents

    USER'S GUIDE Mark 5 Nuvo M5C5 OXYGEN CONCENTRATOR [Original language is English ] Federal Law (US) restricts this device to sale by, or on the order of, a licensed physician. This oxygen concentrator should be used Danger: Do not smoke when using oxygen or when only under the supervision of a licensed near this device.
  • Page 2: General Safety Guidelines

    GENERAL SAFETY GUIDELINES Only persons who have read and understood this entire manual should be allowed to operate the MARK 5 Nuvo [M5C5]. USE OF OXYGEN CONFORMITY WITH IEC60601-1 (§ 6.8.2 B): Oxygen is not a  ammable gas, but it accelerates the "The manufacturer, assembler, installer or distributor are not...
  • Page 3: I.1.Front Panel (Fig. I. 1)

    The tube between the cannula and the MARK 5 Nuvo [M5C5] should be limited to 60 feet (20 meters) long, in order to ensure that the oxygen  ow rate remains within speci...
  • Page 4: Iii.cleaning-Maintenance

    The equipment supplier performs periodic maintenance ope- rations to assure continued reliable service from the MARK Onlythe outside of the MARK 5 Nuvo [M5C5] is to be cleaned, 5 Nuvo [M5C5]. with a soft, dry cloth or, if necessary, a damp sponge, then thoroughly dried with wipes and an alcohol based solution.
  • Page 5: Iv.3. Operating Principles

    Accuracy of  ow supplied: • Electrical protection of the MARK 5 Nuvo [M5C5] The  ow supplied is equal to the  ow set on the  owmeter, A 5A circuit breaker is incorporated into the front cabinet accurate to within ±...
  • Page 6: Iv.7. Standards

    The work must be carried out by suitably trained technicians certi ed by the manufacturer. All waste from the MARK 5 Nuvo [M5C5] (patient circuit,  lter, etc.) must be disposed of using the methods appropriate Use original spare parts only (see Pg. 7).
  • Page 7 IV. 10. Troubleshooting. Observations Possible Causes Solutions The I-O (ON/OFF) button is in the Power cable is not correctly Check the cable connection. ON position but the device does not plugged in. operate. Check the circuit breaker on the front of the unit under the  owmeter. Power failure.
  • Page 8 MMainain Main Maintenance This i Maintenance Items; Nidek Medical Products, Inc. 3949 Valley East Industrial Drive Birmingham, Alabama 35217 U.S.A. Tel: 205-856-7200 Fax: 205-856-0533 May 2011 2010-2224F Rev C...

Table of Contents