VXI VT1509A User Manual

Eight-channel fixed filter and amplifier signal conditioning plug-on

Advertisement

Quick Links

bus
Eight-Channel
Fixed Filter and Amplifier
Signal Conditioning Plug-on
VT1509A
User's Manual
The VT1509A manual also applies to Agilent/HP E1413s as Agilent/HP E1413 Option 19.
Enclosed is the User's Manual for the VT1509A Signal Conditioning Plug-on. Insert this
manual in your VT1413C and Agilent/HP E1313 manual behind the "Signal Conditioning
Plug-ons" divider.
Copyright © VXI Technology, Inc., 2003
Manual Part Number: 82-0083-000
Printed: June 16, 2003
Printed in U.S.A.

Advertisement

loading

Summary of Contents for VXI VT1509A

  • Page 1 VT1509A User’s Manual The VT1509A manual also applies to Agilent/HP E1413s as Agilent/HP E1413 Option 19. Enclosed is the User’s Manual for the VT1509A Signal Conditioning Plug-on. Insert this manual in your VT1413C and Agilent/HP E1313 manual behind the “Signal Conditioning Plug-ons”...
  • Page 3: Table Of Contents

    Chapters 1 and 2 of your VT1413C or Agilent/HP E1313 manual. Identifying the Plug-on You’ll find the VXI Technology part number on the connector side of the SCP to the left of the serial number bar code. For the VT1509A, the part number is : VT1509A. Introduction...
  • Page 4: Connecting To The Terminal Module

    VT1509A SCPs. The connections are shown in Figure 1. For Agilent/HP E1313 Terminal Moduless, use stickers for VT1509A SCPs. The connections are shown in Figures 2 and 3. For Agilent/HP E1413B and below Terminal Modules, use stickers for HP E1413 Option 19 SCPs. The connections are shown in Figure 4.
  • Page 5 Figure 2 VT1509A B-size Terminal Module Connections (Ch 00-31) Figure 3 VT1509A B-size Terminal Module Connections (Ch 32-63) Figure 4 Agilent/HP E1413B Option 19 Terminal Module Connections Connecting To The Terminal Module VT1509A Fixed Filter+Amplifier SCP 5...
  • Page 6: Programming With Scpi Commands

    6 enter statement here Querying the While the VT1509A does not allow controlling whether the filters are enabled or disabled, this state can be queried. The response to this query Filter State will always be 1. To query any channel to determine if it is enabled or disabled use the INPut:FILTer[:LPASs][:STATe]? (@<channel>)
  • Page 7 2 enter statement here Querying the While the VT1509A’s amplifiers have fixed gain, the channel gain can be queried. The response to this query will always be 64. To query any channel Channel Gain to determine its gain setting use the INPut:GAIN? (@<channel>) command.
  • Page 8: Programming With Register Commands

    Plug-on. When Register Programming an SCP most communication is through the Signal Conditioning Bus. For that you will use the Register Commands: SCBWRITE <regaddr> <regvalue> SCBREAD? <regaddr> 8 VT1509A Fixed Filter+Amplifier SCP Register Based Programming...
  • Page 9 Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 SCP 7 SCP 6 SCP 5 SCP 4 SCP 3 SCP 2 SCP 1 SCP 0 Register Based Programming VT1509A Fixed Filter+Amplifier SCP 9...
  • Page 10: Specifications

    Specifications These specifications for the VT1509A reflect the combined performance of the VT1413C or Agilent/HP E1313 and the VT1509A Signal Conditioning Plug-on. These specifications are not to be added to those presented in the VT1413C or Agilent/HP E1313 User’s Manual.
  • Page 11 0.25°C Type R A/D Filter 0 to 100°C 100 to 200°C 200 to 600°C 600 to 1000°C 1.40°C 0.75°C 0.45°C 0.30°C 1.40°C 0.70°C 0.40°C 0.30°C * [SENSe:]FILTer[:LPASs][:STATe] ON (max scan rate - 100 rdgs/sec/channel) Specifications VT1509A Fixed Filter+Amplifier SCP 11...
  • Page 12 15 to 30°C 30 to 70°C 70 to 85°C 0.011°C 0.013°C 0.017°C 0.010°C 0.013°C 0.017°C 10 k Thermistor A/D Filter 30 to 60°C 60 to 90°C 90 to 115°C 0.015°C 0.016°C 0.022°C 0.014°C 0.015°C 0.020°C 12 VT1509A Fixed Filter+Amplifier SCP Specifications...