NAGATA USER MANUAL ELECTRONIC WEIGHING SCALE TABLE TOP SCALE SERIES Item Number: FAT-06 6 keys function WBG10B-SMD(1) FAT08-RF6S10...
Page 2
Thank you to purchase our scale, please read this manual before try to operate the scale. Before operation: 1. Place the scale on a stable surface. 2. Adjust four feet to center of the bubble on the scale. 3. Avoid hot sunshine directly on the scale or near the exhaust port of ventilating system.
Page 3
Key pad function: ZERO Press this key to return the display to a zero reading. To deduct the weight of a container. TARE TARE 1. Ensure that there is no weight on the weighing platform. 2. Place an empty container on the scale platform. 3.
Page 4
ZERO ZERO TARE NET GROSS HI/LO Available capacity and graduation: Item Capacity Graduation Resolution FAT-06 20.0kg 1.0g 1/20000 Back-light function setting:(Option) <1>. Press "TARE" lasting 3 seconds , when displays "0" release it. TARE Back-light mode is off (Originate setting) <2>.
Page 5
NAGATA FAT-06 Calibration Instructions Note: The test weight used to calibrate the scale should be no less than 1/6 the max capacity of the scale. Example: For a 20kg capacity scale, the minimum test weight used to calibrate the scale must be 4kg ...
Page 6
9. Once the screen is displaying the weight of the test weight you are using, place the test weight on the scale. Press the HI/LO key. Note: If the screen display does not match the test weight then remove the test weight. Press the ZERO+ key.