Bar Code Calibration Procedure without Remote Display*

The following calibration procedure can be used when the remote scale display is not present. The Multi-Function Button
(see page 36) is used to advance to the next stage of calibration and the LED display notifies the operator which 'calibration stage' is
active.
1.
Temporarily remove the platter and place it in a safe location.
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and Measures statutes and
regulations applicable in your area by checking with the appropriate government agency before placing a newly calibrated
unit into service or removing any official seals.
2.
If this is a currently installed scanner/scale, there will be either an adhesive label seal or wire sealed cover over the scale
calibration button. New installation units will only have a factory-applied adhesive seal covering the scale calibration button
installed.
To remove the adhesive label seal, peel off the label and discard.
To remove the wire seal, cut the wire and discard. Fully loosen the security screw and lift off the cover.
Place the security screw and cover in a safe location.
3.
Enter the scale program mode. Press the calibration button to engage the scale calibration mode. If the system is a dual cable
system, disconnect the host to scale RS232 cable from the unit.
* These procedures are for MS2320 StratosH Scanner/Diva Scale models only
44
Figure 47. Diva Scale security seals
Figure 48. Entering the Scale Program Mode
.