Accu-Chek
Active Control
®
mg/dL
mmol/L
37 – 67 2,1 – 3,7
1
134 – 181 7,4 – 10,0
2
Figures are examples
Control solution
used
After flagging the result as a performance check:
12 Compare the test result with the concentration
table on the label attached to the test strip
Unit of
container.
measurement
The result must be within the specified concentra-
tion range.
Make sure that you compare the test result with the
concentration data that corresponds to the relevant
control solution that you used (1 or 2) and to the unit
of measurement your meter displays (mg/dL or
mmol/L).
Checking your meter
77