BC GROUP PSR-2200-DS PATIENT SIMULATOR REMOTE CONTROL The Model PSR-2200-DS is a Microprocessor based remote control for use with a specific model patient simulator. The remote allows the user to configure the patient simulator using pre-programmed configurations as well as codes.
OVERVIEW This section looks at the layout of the PSR-2200-DS and gives descriptions of the elements that are present. 20 Light Touch Keys LCD Graphical Display with Backlight: Shows configuration and output setup. 25’ Coiled Cable For connection to a PC or...
Page 7
List of Codes Model and Serial Number Label Battery Compartment 1 9V Battery...
KEYS 20 soft touch keys are provided for system operation: – Pressing and holding this key for 3 seconds will turn the unit off. – This key allows the user to input pre-programmed codes using the number keys. – This key returns parameters to their Standard settings. –...
SCREENS The Main Screen shows the current operating mode of the Patient Simulator. The Backlight Time Screen allows for the setting of the Backlight Off Timer. The Backlight can be Off, set to 1-20 seconds or On. The keys are used to choose the desired setting.
If the cables are connected properly, cycle power on the PSR-2200-DS and the Patient Simulator and try the command again. LOW BATTERY – This message indicates that the batteries are low and should be...
To control a patient simulator, the remote control is first plugged into the RS232 connector of the simulator. You must use the adapter provided to convert the PSR-2200-DS 7 pin Mini-Din connector to the RS232 DB9 connector on the simulator.
Page 12
FUNCTION CODE DISPLAY DESCRIPTION Normal Sinus NSR of 80 BPM TACH Sinus Tachycardia NSR of 140 BPM BRAD Sinus Bradycardia NSR of 30 BPM ASYS Asystole Flat line signal Repeating sequence of 5 atrial beats and 1 ventrical beat for twelve seconds, followed by a repeating A-FLTR Atrial Flutter sequence of 3 atrial beats and 1 ventrical beat for six...
Page 13
PRE-DEFINED FUNCTION CODES: A fixed function can be loaded by entering the desired code. The key is pressed and then the Number Keys are used to enter the code. The available codes are located on the label on the back of the remote. make adjustments output...
Page 14
CODE DISPLAY FUNCTION DESCRIPTION PACEMAKER 80 BPM with Pacer Pulse at the start of each P A-PACE Atrial wave 75 BPM with Pacer Pulse at the start of both the AV-PACE AV Sequential P and QRS waves 75 BPM Ventricular Paced beats with every tenth LOSS CAP Sense &...
Page 15
CODE DISPLAY FUNCTION DESCRIPTION BLOOD PRESSURE 0 mmHg Zero Pressure Puts all BP Channels at 0 mmHg PA WEDGE PA Wedge Pulmonary Artery Wedge 10/2 mmHg Right Atrium CVP 15/10 Insert Right Ventricular 25/0 Insert/Pull Back SWANGANZ Swan-Ganz Insertion Pulmonary Artery 25/10 Insert/Pull Back Pulmonary –Arterial Wedge 10/2 Deflate RESPIRATION APNEA...
MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 DT7349DSA Origination Rev 02 DT7349DSB Misc. Edits LIMITED WARRANTY WARRANTY : BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED.
SPECIFICATIONS Physical DISPLAY LCD Graphical 128 X 64 Pixels, Backlit 5.98 x 3.27 x 1.28 Inches ENCLOSURE (152 x 83 x 32.5 mm) ABS Plastic < 1 Lbs WEIGHT (< 0.45 Kg) FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C Electrical...