BC Biomedical PSR-2200-DS User Manual

Programmable patient simulator remote control

Advertisement

Quick Links

PROGRAMMABLE
PATIENT SIMULATOR
REMOTE CONTROL
PSR-2200-DS
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PSR-2200-DS and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BC Biomedical PSR-2200-DS

  • Page 1 PROGRAMMABLE PATIENT SIMULATOR REMOTE CONTROL PSR-2200-DS USER MANUAL...
  • Page 3: Table Of Contents

    BC BIOMEDICAL PSR-2200-DS TABLE OF CONTENTS DESCRIPTION......................3 OVERVIEW........................5 KEYS..........................6 SCREENS........................7 MESSAGES ......................... 8 OPERATION ........................ 9 MANUAL REVISIONS....................14 WARRANTY......................... 14 SPECIFICATIONS ....................... 15 NOTES......................... 16...
  • Page 4: Bc Biomedical

    AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES. CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL, INC. PO BOX 25125 9415 GENTRY AVE ST. LOUIS, MO 63125 1-800-242-8428 314-638-3800 www.bcgroupintl.com sales@bcgroupintl.com Manual PSR-2200-DS Copyright © 2006 www.bcgroupintl.com Made in the USA 7/06 Rev 02...
  • Page 5: Psr-2200-Ds

    BC GROUP PSR-2200-DS PATIENT SIMULATOR REMOTE CONTROL The Model PSR-2200-DS is a Microprocessor based remote control for use with a specific model patient simulator. The remote allows the user to configure the patient simulator using pre-programmed configurations as well as codes.
  • Page 6: Overview

    OVERVIEW This section looks at the layout of the PSR-2200-DS and gives descriptions of the elements that are present. 20 Light Touch Keys LCD Graphical Display with Backlight: Shows configuration and output setup. 25’ Coiled Cable For connection to a PC or...
  • Page 7 List of Codes Model and Serial Number Label Battery Compartment 1 9V Battery...
  • Page 8: Keys

    KEYS 20 soft touch keys are provided for system operation: – Pressing and holding this key for 3 seconds will turn the unit off. – This key allows the user to input pre-programmed codes using the number keys. – This key returns parameters to their Standard settings. –...
  • Page 9: Screens

    SCREENS The Main Screen shows the current operating mode of the Patient Simulator. The Backlight Time Screen allows for the setting of the Backlight Off Timer. The Backlight can be Off, set to 1-20 seconds or On. The keys are used to choose the desired setting.
  • Page 10: Messages

    If the cables are connected properly, cycle power on the PSR-2200-DS and the Patient Simulator and try the command again. LOW BATTERY – This message indicates that the batteries are low and should be...
  • Page 11: Operation

    To control a patient simulator, the remote control is first plugged into the RS232 connector of the simulator. You must use the adapter provided to convert the PSR-2200-DS 7 pin Mini-Din connector to the RS232 DB9 connector on the simulator.
  • Page 12 FUNCTION CODE DISPLAY DESCRIPTION Normal Sinus NSR of 80 BPM TACH Sinus Tachycardia NSR of 140 BPM BRAD Sinus Bradycardia NSR of 30 BPM ASYS Asystole Flat line signal Repeating sequence of 5 atrial beats and 1 ventrical beat for twelve seconds, followed by a repeating A-FLTR Atrial Flutter sequence of 3 atrial beats and 1 ventrical beat for six...
  • Page 13 PRE-DEFINED FUNCTION CODES: A fixed function can be loaded by entering the desired code. The key is pressed and then the Number Keys are used to enter the code. The available codes are located on the label on the back of the remote. make adjustments output...
  • Page 14 CODE DISPLAY FUNCTION DESCRIPTION PACEMAKER 80 BPM with Pacer Pulse at the start of each P A-PACE Atrial wave 75 BPM with Pacer Pulse at the start of both the AV-PACE AV Sequential P and QRS waves 75 BPM Ventricular Paced beats with every tenth LOSS CAP Sense &...
  • Page 15 CODE DISPLAY FUNCTION DESCRIPTION BLOOD PRESSURE 0 mmHg Zero Pressure Puts all BP Channels at 0 mmHg PA WEDGE PA Wedge Pulmonary Artery Wedge 10/2 mmHg Right Atrium CVP 15/10 Insert Right Ventricular 25/0 Insert/Pull Back SWANGANZ Swan-Ganz Insertion Pulmonary Artery 25/10 Insert/Pull Back Pulmonary –Arterial Wedge 10/2 Deflate RESPIRATION APNEA...
  • Page 16: Manual Revisions

    MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 DT7349DSA Origination Rev 02 DT7349DSB Misc. Edits LIMITED WARRANTY WARRANTY : BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED.
  • Page 17: Specifications

    SPECIFICATIONS Physical DISPLAY LCD Graphical 128 X 64 Pixels, Backlit 5.98 x 3.27 x 1.28 Inches ENCLOSURE (152 x 83 x 32.5 mm) ABS Plastic < 1 Lbs WEIGHT (< 0.45 Kg) FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C Electrical...
  • Page 18: Notes

    NOTES...