Digital illuminance meter

Advertisement

Quick Links

DIGITAL ILLUMINANCE METER
INSTRUCTION MANUAL
PLMT68
www.pyleaudio.com

Advertisement

loading

Summary of Contents for Pyle PLMT68

  • Page 1 DIGITAL ILLUMINANCE METER INSTRUCTION MANUAL PLMT68 www.pyleaudio.com...
  • Page 2 Ⅰ INSTRUCTION The digital illuminance meter is a precision instrument  used to measure illuminance (lux, footcandle) in the field. It is meet CIE photopic spectral response.  It is fully cosine corrected for the angular incidence of  light. The illuminance meter is compact, tough and easy to ...
  • Page 3  Meter corrected for spectral relative efficiency. Correction factor need not be manually calculated for  non-standard light sources.  Short rise and fall times. Peak-hold function for tracing the peak signal of light  pulse with least duration 10μs and keep it. ...
  • Page 4  Spectral Response: CIE Photopic. (CIE human eye response curve).  Spectral Accuracy: CIE Vλ function f1’≦6% Cosine Response: f2’ ≦2%  Accuracy: ±5% rdg±0.5% f.s.  Repeatability: ±3%   Sampling Rate: 13.3 times/sec of analog bar-graph indication;1.3times/sec of digital display. ...
  • Page 5  Photo detector Dimensions: 115L×60W×20H(mm);  Meter Dimensions: 170L×80W×40H; Weight: 390g.   Accessories: Carry case, instruction manual, battery. Ⅳ NAME OF PARTS AND POSITIONS...
  • Page 6 LCD Display: 3-3/4 digit displays with a maximum reading of 3999,and the indicating signs of measured values, unit function symbols and decimal points etc are display. Range Selector key: It indicates 40.00 lux, 400.0 lux,4000 lux,40.00 Klux and 300.0 Klux /40.00 fc,400.0 fc,4000 fc,30.00 Kfc total 5 ranges for lux and 4 ranges...
  • Page 7 illuminance meter ON or OFF. MAX/MIN key: Maximum and minimum reading recorder control key. RESET key: Push function reset control key. BACK-LIGHT key: Back light control key. REL key: Relative reading control key. Data-Hold key: Data Hold control key. Peak Hold key: Peak Hold recorder control key. 10.
  • Page 8 in a horizontal position. read the illuminance nominal from the LCD display. Overrange: If the instrument only displays “OL”, the input signal is too strong, and a higher range should be selected. Data-Hold mode: Press the hold key to select Data-Hold mode.
  • Page 9 Relative reading mode: Press REL key to enter Relative mode. The display shown zero value and the current reading will be stored as a zero-in value. Press again to exit this mode. 10. Reset mode: Every time when this key is pressed the pushed function(HOLD,MAX/MIN,REL,PEAK except...
  • Page 10 Ⅶ SPECTRAL SENSITIVITY CHARACTERISTIC  To the detector, the applied photo diode with filters makes the spectral sensitivity characteristic almost meet C.I.E.(INTERNATIONAL COMMISSION ON ILLUMINATION) Photopic curve V (λ) as the following chart described. Ⅷ MAINTENANCE The white plastic disc on the top of the detector should be cleaned with a damp cloth when necessary.
  • Page 11 The calibration interval for the photo detector will vary according to operational conditions, but generally the sensitivity decreases in direct proportion to the product of luminous intensity by the operational time. In order to maintain the basic accuracy of the instrument, periodic calibration is recommended.
  • Page 12 HOTEL Public room, 100~200 9~18 Cloakroom Reception 200~500 18~47 Cashier 750~1,000 70~93 STORE Indoors Stairs 150~200 14~18 Corridor Show window, 750~1,500 70~140 Packing table Forefront of show 1,500~3,000 140~279 window HOSPITAL Sickroom, 100~200 9~18 Warehouse Medical Examination 300~750 28~70 Room Operating room, 750~1,500 70~140...