Troubleshooting

Tab. 2
SYMPTOM
3.
With the switch in ON
position, no LED are ON.
4.
With the switch in ON
position, the LED L1 is
blinking and L2 is OFF.
Ion Pump Battery Power Supply / 87-900-111-01 (E)
POSSIBLE CAUSE
The battery is exhausted.
The current that supplies
the pump is < 50 nA.
Discharge is not activated
and the pump does not
work.
Instructions for Use

Disposal

FIXING
Change the battery.
Increase the pressure inside
the pump (ex. by heating the
pump), discharge is
activated and L1 goes OFF
and L2 starts blinking.
4