Do you have a question about the 725 series and is the answer not in the manual?
Questions and answers
Summary of Contents for MIRAK 725 series
Page 1
BARNSTEAD|THERMOLYNE CORPORATION MIRAK™ Hot Plates, Stirrers and Stirring Hot Plates OPERATION MANUAL AND PARTS LIST SERIES 725, 726, 727, 728, 729, 730 Model # Voltage Description HP72620 HP72620-26 220-240 HP72624 HP72625 S72520 S72520-26 220-240 S72524 S72525 SP72720 SP72720-26 220-240 SP72724...
IMPORTANT INFORMATION This manual contains important operating and safety information. The user must carefully read and understand the contents of this manual prior to the use of this equipment. Table of Contents Safety Information.........................3 Alert Boxes ..........................3 Warnings............................3 Introduction..........................5 Unpacking............................5 Installation............................7 Principles of Operation........................7...
Safety Information Your Thermolyne MIRAK™ stirrer, hot plate, or stirring hot plate has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. Alert Signals WARNING Warnings alert you to a possibility of personal injury.
Page 4
Please note the following WARNINGS: WARNING This warning is presented for compliance with California Proposition 65 and other regulatory agencies and only applies to the insulation in this product. This product contains refractory ceramic, refractory ceramic fiber or fiberglass insulation, which can produce respirable dust or fibers during disassembly.
Unpacking Remove your MIRAK hot plate, stirrer or stirring hot plate from the carton. Inspect to ensure that the unit has not been damaged during shipment. If the unit appears to have sustained shipping damage contact the distributor from whom you purchased this product or Barnstead/Thermolyne Customer Service at 800- 553-0039.
The electronic stirring speed control will maintain the speed setpoint when the unit is loaded at 2 RPM as displayed by the LED display on the unit. The motor in the MIRAK stirrers and stirring hot plates produces maximum stirring torque under normal laboratory load conditions, and is combined with a powerful magnet to provide exceptional magnetic coupling with a stirbar.
The Power Switch Your MIRAK unit has a separate ON/OFF power switch which is a black rocker switch located on the lower front corner of the right side as you're facing the unit. This separate power switch allows you to turn your MIRAK hot plate, stirrer or stirring hot plate on or off without disturbing your temperature and/or speed setting.
Your MIRAK hot plate or stirring hot plate will display the temperature in C. The display will always indicate 'OFF' or your last setpoint, and the 'SET' light will be illuminated until you press the heat control knob to initiate operation.
When using a probe with your MIRAK hot plate or stirring hot plate it is recommended that you use a clamp on a support rod to hold the probe in your solution.
To verify the temperature reading of the probe, insert a calibrated thermometer into a solution you are heating on the stirring hot plate. When the thermometer reading has stabilized, record the temperature readings given by both the thermometer and by the probe as indicated on the MIRAK display. To calibrate the probe: Unplug unit, turn unit upside down and remove screws from bottom.
Because the advanced electronic control in the MIRAK is capable of precisely regulating the top plate temperature, metal vessels and sand baths may be heated safely without the danger of the ceramic top breaking.
Page 13
12" x 12" 240 volt MIRAK 540 C 1 liter 1000 ml (water) 17 min. 12" x 12" 240 volt MIRAK 540 C 2 liter 1800 ml (water) 21 min. 12" x 12" 240 volt MIRAK 540 C 4 liter 4000 ml (water) 25 min.
Operating Specifications Top Plate Surface....................Solid Ceramic Heating Specifications Temperature range All 7" x 7" models and the 208 and 240 volt models of the 12" x 12" models. 40 C to 540 C 100 and 120 volt 12" x 12" models. 40 C to 370 C Heat-up time to maximum temperature (unloaded top plate).
Maintenance and Servicing WARNING Disconnect from power supply before servicing. Keep top surface clean. Use a non-abrasive cleaner. Alkali spills, hydrofluoric acid spills or phosphoric acid spills may damage top and lead to thermal failure. Unplug unit and remove spills promptly. Do not immerse unit for cleaning. Replace the top immediately if damaged by etching, scratching or chipping.
Power board for stirrer detailing correct placement of wires and cables. Replacing the Logic Board Unplug unit, turn unit upside down and remove screws from bottom. Open unit. Locate logic board in the center of the unit. Note the position and color of wires and cables connected to the board.
The set-up procedure should be followed whenever the top plate assembly or the logic board is replaced in your MIRAK hot plate or stirring hot plate. The set-up procedure is not necessary when replacing the top plate or the logic board in the stirrers.
Logic Printed Circuit Board Set-up Procedure Place a 500 ml flask with 300 ml of water on the top plate. Place a standard laboratory thermometer (do not use the probe for this procedure) 1/2" above the bottom of the flask. Turn the unit on and set the temperature at 70 C.
Refer to the difference that you recorded in step #3. If that difference was greater than +3, turn the knob CLOCKWISE, counting each click of the knob until you have advanced the knob the same number of positions as the difference you calculated in step #3, then turn the knob one additional click .
Replacement Parts List To insure your safety and for proper operation, the ceramic top plates for hot plates and stir plates are only sold as complete assemblies. This assembly includes the ceramic top, element, thermocouple, insulation, baffle plate, and 2 ceramic top holders. Key Number Part Number Unit Size Unit Voltage...
Barnstead|Thermolyne One Year Limited Warranty Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne and sold by it within the continental United States or Canada proves to be defective in material or construction, Barnstead|Thermolyne will provide you, without charge, for a period of ninety (90) days, the labor, and a period of one (1) year, the parts, necessary to remedy any such defect.