Agilent Technologies E1343A Service Manual page 28

75000 series b relay multiplexers
Hide thumbs Also See for E1343A:
Table of Contents

Advertisement

Ch 00 - 07 and 90
GU Measurements
26 Verification Tests
Figure 2-4. Ch 0-7 and 90 LO Measurement Connections
3. Repeat for Channels 01 - 07 LO and Channel 90 LO
Repeat steps 1 and 2 for channels 01 - 07 LO and 90 LO
Use
card # and
For Channel 90 LO, subtract combined value of R21 + R27
1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM leads as shown in Figure 2-5
Turn mainframe power ON
Figure 2-5. Ch 00-07 and 90 GU Measurements Connections
2. Measure Channel 00 GU Resistance
Send
Send
Trigger the DMM with
Send
Subtract measured value of R22 from DMM reading
Enter the result in Table 2-2 for Channel 00 GU
and
CLOS (@nncc)
OPEN (@nncc)
= channel # (omit leading zeroes in
cc
to multiplexer
*RST
to close chan 00, where
CLOS (@nn00)
TRIG SGL
to open channel 00
OPEN (@nn00)
, where
=
nn
)
nn
= card #
nn
and note reading
Chapter 2