VIAspire™ Oxygen Portables 300 and 600. Before You Begin Please read this entire manual before you use your VIAspire Liquid Oxygen Portables 300 and/or 600. If you are unable to understand the warnings, cautions, and instructions, contact your medical equipment provider before using this equipment.
Terms, Abbreviations, and Symbols General Dangers and Warnings This guide and the VIAspire Portable product labeling uses In order to ensure the safe installation, assembly, and opera- the following terms, abbreviations, and symbols: tion of the VIAspire Liquid Oxygen Portable, these instruc- tions MUST be followed.
Page 4
Use extreme care when fi lling. Use extreme care WARNING when fi lling your VIAspire Liquid Oxygen Portable. Do not use if... WARNING Do not use the VIAspire Portable 300 or 600 if it: Keep unit is in an upright position. Always WARNING Is not working properly. •...
Instruct the patient on the safe operation of the VIAspire Liquid Oxygen Portable and all associated accessories, and review ALL warnings. Filling a VIAspire Portable 300 or 600 £ Leave a copy of this instruction guide with the patient. After the VIAspire Liquefi er has liquefi ed enough oxygen, you can fi...
Page 6
Instructions Align the coupler on the bottom of the VIAspire Portable with the coupler in the dock area on top of Open the Liquid Oxygen Coupler the Liquefi er. Access Door on the top of the VIAspire Liquefi er. This will stop the liquefi...
Page 7
• Do not touch any frosted parts. oxygen. When you are at rest, the VIAspire Portables 300 and 600 will deliver a resting volume of oxygen. As you • If any liquid is leaking from the liquefi er reservoir or the...
Page 8
Starting Oxygen Delivery from the With each breath, you should now be receiving a set dosage of gaseous oxygen. VIAspire Portable • The green LED battery indicator will light up with each To receive oxygen from your VIAspire Portable, follow breath, signaling that the VIAspire Portable is being trig- these steps.
SmartDose mode. WARNING Using the VIAspire Portable 300 or 600 in continu- ous fl ow (CF) mode will use up oxygen at a faster rate. Monitor the contents indicator as described...
(with the exception of the condensation pad). tube is not blocked. Keep the exterior of the VIAspire Portable 300 or 600 clean Check that the CF/SD switch is in • using a clean, damp, oil-free, lint-free cloth.
Allow the portable to sit for at least 8 hours to allow it to empty. Storage Store the VIAspire Portable 300 or 600 in a cool, clean, dry area when not in use. Avoid storing in direct sunlight. Battery disposal Dispose of all alkaline batteries in accordance with local, state, and federal regulations.
Page 12
VIAspire™ Oxygen Portables 300 and 600 Product Manual Insert This insert applies to the use of the 300 and 600 Liquid Oxygen Portable with PB coupler. The VIAspire 300 and 600 Liquid Oxygen Portable with SmartDose technology is also available with a PB coupler suitable for use with liquid oxygen (LOX) systems which have a male PB Coupler (100% LOX).
Need help?
Do you have a question about the VIAspire 300 and is the answer not in the manual?
Questions and answers