SECTION 7—MAINTENANCE

Nebulizer Change

The nebulizer is designed for single patient use.
The nebulizer MUST be replaced after a long period of inactivity or when 
it is obstructed by dry medication, dust, etc. The reusable nebulizer should 
last from six months to one year with general use. The disposable 
nebulizer should last 30 days. For the nebulizer that is disposable, it is 
recommended that an extra nebulizer be kept at all times.

Filter Change

1.
The filter MUST be replaced after approximately 30 hours of use or 
when it turns grey.
2.
Open the filter holder to replace the filter with a new one.
DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate unit without an air filter.