58502A Command Summary - Symmetricom 58502A User Manual

Broadband distribution amplifier
Table of Contents

Advertisement

Keyword/Syntax
*CLS
*IDN?
:ALARm?
DIAGnostic
:CALibration
:ROSCillator
:EFC
:ABSolute
:ABSolute?
:STEP
:STEP?
INPut
:A:QUEStionable?
:B:QUEStionable?
:ALARm?
:SELect
Chapter 4 Command Reference

58502A Command Summary

58502A Command Summary
Table 4-1 lists and briefly defines all of the commands that can be used
to operate the 58502A.
Table 4-1. 58502A Command
Parameters
Description
Clears error queue.
returns unquoted
Returns identification string.
string
returns 0 or 1
0 = alarm is off, 1 = alarm is on. Alarm is on when any of the
outputs are not functional. (There may be additional internal
conditions that generate alarm.)
Diagnostic Subsystem (USED WITH OPTION 010 ONLY)
Node Only
Node Only
Node Only
Send:
Sets the absolute EFC (electronic frequency control) of the
<EFC value> or
oscillator. Use UP or DOWN to adjust the EFC by the increment
UP or DOWN
specified in the STEP command (see next command). For
example, if present EFC value is 1000 and STEP value is 13,
:EFC:ABS:UP will increase EFC value to 1013.
returns:
Query.
<EFC value>
<EFC increment
Increments the EFC value by this amount.
value>
returns <EFC
Query.
increment value>
INPUT Subsystem
returns 0 or 1
Reports if there is a signal present on input A.
0 = signal present
1 = no signal present.
returns 0 or 1
Reports if there is a signal present on input B.
0 = signal present (not questionable),
1 = no signal present.
0,0,0 or 1,0,0 or
Reports the alarm status of devices connected to the input pins
0,1,0
(6, 7) of the ALARM/STATUS port.
0, 0,0 = no alarm exists.
1, 0,0 = alarm from input A devices connected to pin 6.
0,1, 0 = alarm from input B devices connected to pin 7.
A or B
Enables either Input A or Input B. Use the query form to see
which input is currently in use. This command also disables
auto-switching.