Display settings
3. In the next dialog box, select option Reset to the latest user
calibration and click OK.
The calibration and the luminance will be reset to the latest
calibrated position.
Note:
When you start up the system, the calibration is also reset to
the latest calibrated position. The luminance, however, is
kept at the latest value.
To view a graph showing the result of the latest calibration:
1. In the Calibration tab dialog, click on Show calibration
chart.
This button is available only if the display has been
calibrated with a sensor before.
2. A graph is shown displaying the applied display function (to
which the display is calibrated) compared to the native panel
behavior.
To revert to the default factory settings:
1. If you have multiple displays connected (multi-head system),
select the display you wish to reset to the default settings.
2. Click the Reset button.
3. In the next dialog box, select option Reset to the factory
default calibration and click OK.
4. As a result, the settings are restored to the default factory
values.