Mitutoyo DP-1VR User Manual page 59

Digimatic mini-processor
Hide thumbs Also See for DP-1VR:
Table of Contents

Advertisement

Symptom
When printing, the
DP-1VR unexpect-
mistake
edly returns to the
initial state seen
after the power is
is present on the battery pole
turned ON.
Printing is too
used.
light
connected to the measuring
instrument.
(i.e. AC line voltage) is too low;
falls 5% or more than the rated
voltage
same power line with a high-
voltage or large-current device
Printer does not
turned to OFF
print
paper, etc
input mode and the input interval is
set to 0.25 or 1 sec.
Miscounting often
occurs on the
same power line with a high-
measuring
voltage of large-current device.
instrument side
Possible cause
Manganese dry cells are used by
Peeling or poor fitting of the seal
The specified AC adapter is not
The DP-1VR supply power is
Input voltage to the AC adapter
The AC adapter is sharing the
The printing head is not clean
[PRINTER ON/OFF] switch is
The printer is jammed with
The DP-1VR is in the timer
The AC adapter is sharing the
- 51 -
TROUBLESHOOTING
Remedies
Use properly charged size AA
Nickel Hydrogen battery (Ni-Mh)
or Alkali battery (LR6)
Correct the peeling or poor
fitting of the seal
Use the dedicated AC adapter
09EAA088 (100/115V)
09EAA088D (230V)
09EAA088E (230V UK)
supplied with the DP-1VR
The DP-1VR power supply
should not be connected to the
measuring device. Use a separate
power supply for each of them
Adjust the voltage supply of the
power line correctly, and confirm
the result
Connect the AC adapter to a
separate power line
Clean the printing head with a
cotton swab, etc
Press the [PRINTER ON/OFF]
switch again to turn on.
Remove it by tweezers, etc
When the input interval is set to
0.25 or 1 sec, the printer will be
automatically turned off.
Connect the AC adapter to a
separate power line