To Troubleshoot Fast Bus Failures - Agilent Technologies 35670A Service And Operator's Manual

Dynamic signal analyzer
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

4
Service Guide

To troubleshoot fast bus failures

A1
B1
C1
26
New table:
A17 P10 Pin
B14, C14, B15, C15, B16
B22
B24
C12
C19
C23
C24
New illustration showing location of motherboard connector
pins.
2 34
Figure 11 A17 CPU motherboard (P10) connector pin locations
Page in original guide 4-30
Signal Name
TTL Logic State in Test Mode
FA1 to FA5
Toggling
ECLK
Toggling
FSELAn
Toggling
BRESETn
Low
FRW
Toggling
FIFOENn
High
FSELSn
Toggling