Guide are subject to regulatory registration, licensing, clearance, or approval. Although equipment service representatives uncrate and install the VITROS ECi/ECiQ Immunodiagnostic System at the laboratory site, the site needs to be prepared according to specifications. This section describes general requirements for installing the ECi/ECiQ System at your laboratory, including physical and environmental requirements.
Page 3
162.6 cm (64 in.) 130.2 cm (51.25 in.) 179.1 cm (70.5 in.) 73.7 cm (29 in.) 61 cm (24 in.) 59.7 cm (23.5 in.) - Left side door clearance 66 cm (26 in.) Operator’s Guide VITROS ECi/ECiQ Immunodiagnostic System Introduction...
Installation and Site Specifications Reference UPS Dimensions: Printer Stand for Epson LQ/Model 570e and Epson LX/Model 300* Printer Dimensions: Printer Stand with Texas Instruments Omni 800/Model 830e Printer Dimensions: *Standard printer for the system. Printers are ordered in accordance with regional specifications.
Page 5
Epson LX/Model 300* Line Voltage: 99-132 volt AC Frequency: 50/60 Hz *Standard printer for the system. Printers are ordered in accordance with regional specifications. Electrical Outlet (Receptacle Type) Epson LX/Model 300* NEMA 5-15R *Standard printer for the system. Printers are ordered in accordance with regional specifications.
Page 6
Refer to Chapter 1, “Introduction” Modem. Refer to Chapter 2, “System Overview” e-Connectivity. Operator’s Guide VITROS ECi/ECiQ Immunodiagnostic System for detailed information on the VPN and for detailed information on Part No. 7B6090 Version 2.0...
The system is mounted on casters to facilitate its relocation within the laboratory. The new site must meet the same space, electrical, and environmental requirements as specified for the original site. Operator’s Guide VITROS ECi/ECiQ Immunodiagnostic System Introduction Relative Humidity...