7. SPECIFICATIONS
7.1 Base Unit Specifications
Measurement Range:
Resolution:
Accuracy and Linearity: ±1% of full scale at constant temperature, R.H.
Total Accuracy:
Response Time:
Warm-up Time:
Operating Temperature: 15 C - 40 C (59 F - 104 F)
Storage Temperature:
Humidity:
Power Requirements:
Battery Life:
Low Battery Indication: "LOW BAT" icon displayed on LCD
Sensor Type:
Expected Sensor Life: >900,000% O
Alarm System:
Low Alarm Range:
High Alarm Range:
Alarm Accuracy:
Dimensions:
Weight:
Cable Length:
Diverter Fitting:
7.2 Sensor Specifications
Type:
Life:
Interferent
Nitrous Oxide
Carbon Dioxide
Halothane
Enflurane
Isoflurane
Helium
Sevoflurane
Desflurane
0.0-100%
0.1%
and pressure when calibrated at full scale.
±3% Actual Oxygen Level over full operating
temperature range.
90% of final value in approximately 15 seconds
at 23 C
none required
-15 C - 50 C (5 F - 122 F)
0-95% (non-condensing)
2, AA Alkaline batteries (2 X 1.5 Volts)
approximately 3000 hours in typical use.
Maxtec MAX-250E galvanic fuel cell
Hours
2
over 2 years in typical applications
high/low alarms, flashing red LEDs, nominal
2300Hz audible alarm beeper
18%-99% (>1% lower than high alarm)
15%-99% (>1% lower than high alarm)OM-25ME-L
19%-99% (>1% higher than low alarm)
16%-99% (>1% higher than low alarm)OM-25ME-L
exact to displayed alarm value
3.5"(W) x 5.5"(H) x 1.5"(D) [89mm x 140mm x 38mm]
approximately .92 lbs. (417g)
10 ft. (3m) fully extended
fits industry standard, 15 mm "T" adapter
galvanic fuel sensor (0-100%)
2 years in typical applications
Volume % Dry
75%
10%
5%
5%
5%
70%
6%
15%
Interference in O
<2%
<2%
<2%
<2%
<2%
<2%
<2%
<2%
%
2
16