Q-RPT Replacement Kit
Instruction sheet
This
kit
includes
Transducer calibrated to a specific range and class.
The kit also contains the following components:
DESCRIPTION
Crimp terminal
Crimp terminal
Crimp terminal
Crimp terminal
Connector
Connector
Connector
Connector
Connector
CRIMP TOOLS
PRODUCT
PPC4
PPCH and PPCH-G
RPM3, molbox, RPM4, PPC3
Introduction
Only DHI or DHI Authorized Service Providers are
authorized
to
install
Replacement or installation by other parties will
void the warranty of the Q-RPT and possibly the
equipment into which it is installed. Measurement
and/or control specifications are also valid only
under these conditions.
Q-RPTs are used in a variety of DHI Controllers,
Monitors and Flow Terminals. Different crimps,
connectors and associated crimp tools are used
with different models. The pages that follow show
the crimps, connectors and crimp tool required for
each equipment type. Refer to the wiring diagrams
on the following sheets and to the detailed
a
replacement
Pressure
QTY
PART NO.
10
2207416
10
3154579
2
3154472
10
3154485
1
2194502
1
3140008
1
1812216
1
3141826
1
3141173
(Shown for reference only.
Not supplied with kit.)
Molex TOOL NO.
63819-0500
63819-0000
63811-8200
replacement
Q-RPTs.
instructions included with Molex hand crimp tools,
to install crimps onto wires. Molex crimp tool
instruction sheets are also available on the Molex
website at
www.molex.com.
http://www.molex.com/pdm_docs/ats/ATS-638190500.pdf
http://www.molex.com/pdm_docs/ats/ATS-638190000.pdf
http://www.molex.com/pdm_docs/ats/ATS-638118200.pdf
The Molex Quality Crimping Handbook 63800-0029 is
also available on Molex website and is a good reference.
http://www.molex.com/pdm_docs/ats/TM-638000029.pdf
DDS
NOTE:
Some
Transducers include a separate wire bundle used
for DIGIQUARTZ DATA STORAGE (DDS).
bundle includes a White/Violet, White/Brown and in
some cases a Yellow wire. The DDS wires are
utilized only on PPCH, PPCH-G, and PPC4
products. For all other DHI products do not disturb
the DDS bundle. Leave it coiled with the wire ends
sealed as supplied by Paroscientific.
Wire Lengths:
Specific wire lengths have been
intentionally
omitted
Preferred method is to measure the wire length on
the existing Transducer being replaced and cut the
wire lengths on the replacement Transducer to
same length.
Q-RPT must be calibrated after installation.
Note
Paroscientific
Pressure
The
from
these
instructions.