Download Print this page

Fluke PK-PPC-BG-CVU PN 3070389 DVU Installation Manual

Dual volume unit and interconnections kit

Advertisement

Quick Links

PK-PPC-BG-CVU Dual Volume Unit and
Interconnections Kit
P/N 3070389
Interconnections Kit, P/N 3070389, includes:
DESCRIPTION
Dual Volume Unit (DVU)
Tubing, PFA, 1/4 in. x 160
cm
Adaptor, 1/4 in. tube x O-
seal 1/8 in. NPT M
Adaptor, 1/4 in. tube x 1/4
in. NPT M
Union, 4P
Tee, 1/4 in. tube
Adaptor, 1/4 in. tube x 1/4
in. NPT F
Set, 6 each, Rubber Shoes
To Install the Dual Volume Unit (DVU)
 Identify a location for the DVU™ that minimizes
the
length
of
connections and the PPC BG15K™ TEST(+) and
TEST(-) ports. The PPC BG15K and the DVU are
designed to stack one on top of the other. The
DVU may be rack mounted using the rack
mount kit (P/N 3069019).
configurations are for the PPC and DVU to be
stacked directly one on top of the other or with
their rear panels face to face. For PPC4 attach
rubber shoe set to the legs of the DVU.
To Connect the DVU and PPC BG15K
 Note
You will be making and connecting
two identical assemblies, one for
the TEST(+) port and one for the
TEST(-) port.
FIGURE
QTY
CALLOUT
1
N/A
1
1
2
2
2
3
2
4
2
5
2
6
1
N/A
tubing
between
the
Recommended
 Install a union (4) into a 1/4 in. tube x 1/4 in.
NPT M adaptor (3) (do this twice). Screw the
1/4 in. NPT M end of the adaptors into the two
PART NO.
ports of the DVU using Teflon
3119872
3232378
3140484
3134427
3135104
3134453
3134430
3125710
 Install a 1/4 in. tube tee (5) onto each of the
DVU
union/adaptor
installed into the DVU ports in Step
the tee considering the connection to the PPC
BG15K (schematic above is for PPC stacked on
top of the DVU).
 Install the 1/4 in. tube x O-seal 1/8 in. NPT M (2)
into the PPC BG15K TEST(+) and TEST(-) port.
 Cut lengths of PFA tubing as needed to connect
the 1/4 in. tube tee to the adaptors on the PPC
TEST ports. Take care that the cut is at a right
angle to the tube and only cut the length
needed.
 Optional for a connection "For DUT" of 1/4 in. NPT
M rather than 1/4 in. tube:
Cut a small length of PFA tubing to use as
a union between the 1/4 in. tube tee (5) and
the 1/4 in. tube x 1/4 in. NPT F adaptor (6)
(see schematic above).
®
Figure 1: Interconnections Kit
assemblies
that
tape as a sealer.
were
just
. Orient

Advertisement

loading

Summary of Contents for Fluke PK-PPC-BG-CVU PN 3070389 DVU

  • Page 1 PK-PPC-BG-CVU Dual Volume Unit and Interconnections Kit P/N 3070389 Interconnections Kit, P/N 3070389, includes:  Install a union (4) into a 1/4 in. tube x 1/4 in. NPT M adaptor (3) (do this twice). Screw the 1/4 in. NPT M end of the adaptors into the two FIGURE DESCRIPTION PART NO.
  • Page 2 Fluke. Keeping your world up and running.® Fluke Corporation For more information call: From other countries: +1 (425) 446-5500 or DH Instruments Division In the U.S.A. (800) 443-5853 or Fax +1 (425) 446-5116 4765 East Beautiful Lane Fax (425) 446-5116 Web access: http://www.dhinstruments.com...

This manual is also suitable for:

Pk-ppc-bg-dvu pn 3070389 dual volume unit and interconnections kitPk-ppc-bg-cvu