Page 2
Products described in this manual are manufactured under international patents and one or more of the following U.S. patents: 5,142,483, 5,257,640, 5,331,838, 5,445,035. Other U.S. and international patents pending. DH Instruments, DH, DHI, MFC-CB, MFC Switchbox, molbloc and molbox1 are trademarks, registered and otherwise, of DH Instruments, a Fluke Company Document No.
1.4.2.4 MFC SWITCHBOX CONTROL (DIGITAL CONTROL FROM MOLBOX OR MFC-CB) ......4 I N S T A L L A T I O N ..............5 UNPACKING AND INSPECTION ......................5...
The MFC Switchbox™ can be used to connect a molbox1™ MFC control option or an MFC-CB™ control channel to up to five MFCs, or to one MFC and four MFMs. The rear panel of the MFC Switchbox connects to the molbox1 or MFC-CB via a circular 5-pin connector and a 25-pin DSUB connector.
1.3.1 POWER SUPPLY Powers the electronics inside the MFC Switchbox and provides continuous power to up to five connected DUTs. The switching power supply provides +5 volts at 0.5 amps and +15 volts at 4 amps. This 15 volt signal is converted via a DC-to-DC converter to -15 volts at 1.3 amps (leaving approximately 2 amps for the +15 volt signal).
1.4.2 MFC CONTROL FUNCTION SPECIFICATIONS The MFC Switchbox works with any MFC that can be used with the molbox1 or MFC-CB. The switchbox provides the MFCs with continuous power, but the molbox1 or MFC-CB still provides all the control and measurement. For detailed specifications regarding MFC control please refer to specifications section of the molbox1 or MFC-CB Operation and Maintenance Manual.
Use the 25-pin male DSUB to 25-pin male DSUB cable to connect the molbox MFC port on the MFC Switchbox to the MFC port on the molbox1. If using the MFC Switchbox with an MFC-CB, connect to the desired MFC-CB channel (Device1 or Device2).
3. OPERATION OVERVIEW The MFC Switchbox in and of itself does not require any special setup or configuration. It just needs to be connected to the molbox1 or MFC-CB and DUTs as described above, plugged in and powered on by actuating the power switch on the MFC Switchbox rear panel.
The UNIT function selects the molbox1 flow measurement unit. UNITS OF MEASURE When using the MFC Switchbox with molbox1 When the MFC function is active, the molbox1 flow units of measure will always be the units of measure specified in the active MFC profile.
MFM’s that are connected to channels 2..5 with an MFC controlling on channel 1. Remarks The MFC control output can be volts or current. An MFC Switchbox can be connected to the output to switch between multiple devices. Command: “MFCCH=1, mA”...
MFC Switchbox connected to the molbox1. is used. doesn't appear The MFC Switchbox power is not on. Power on the MFC Switchbox. responding when the MFC Switchbox is used. MFC connected to molbox1 doesn't MFC option not implemented.
8. WARRANTY STATEMENT Except to the extent limited or otherwise provided herein, DH Instruments, a Fluke Company (DHI) warrants for one year from purchase, each new product sold by it or one of its authorized distributors, only against defects in workmanship and/or materials under normal service and use. Products which have been changed or altered in any manner from their original design, or which are improperly or defectively installed, serviced or used are not covered by this warranty.
Need help?
Do you have a question about the MFC SWITCHBOX and is the answer not in the manual?
Questions and answers