DH Instruments 1-2 IN. FILTER KIT Instruction Sheet

Installation of 1/2 in. filter kit

Advertisement

This kit is used to provide filtration for 1/2 in. molbloc
flowpaths, especially for 1/2 in. molstic-S and molbloc-
S installations. Unlike most DHI molstics, the1/2 in.
molstic-S
does
not
contamination is not a likely problem for the larger
diameter molbloc-S elements that are normally
installed on it.
For customers who expect the
presence of significant particulate contamination, or
want to maintain very clean gas, this 1/2 in. filter kit
can be installed upstream of critical molstic-S
components.
The kit may also be used for extra
filtration on 1/2 in. supply lines to other molstics.

INSTALLATION OF 1/2 IN. FILTER KIT

ONTO 1/2 IN. MOLSTIC-S
(SINGLE OR DUAL)
The filter kit can be installed on the upstream side of a
single or dual channel 1/2 in. molstic-S. If there is no
regulator installed on the molstic-S, the filter will fit
completely over the molstic base.
If there is a regulator installed on the molstic-S, the filter
kit should generally be installed upstream of the
regulator. In this installation, the filter will extend beyond
the upstream side of the molstic. If this is inconvenient,
the filter kit can be installed on a supply line upstream of
the molstic instead.
Figure 1. Dual 1/2 in. molstic-S, with 1/2 in. Filter Kit
molbloc, molbloc-S, molbox, molstic and molstic-s are trademarks, registered and otherwise, of DH Instruments, Inc.
Document 560056
DH Instruments, Inc.
4765 East Beautiful Lane
Phoenix AZ 85044-5318
USA
include
a
filter
because
VCR is a registered trademark of the Swagelok Company.
Installation of 1/2 in. Filter Kit
P/N 401888
Instruction Sheet

To permanently install the filter kit:

Using a pair of properly sized wrenches, remove the
VCR x tube adaptor that is on the inlet end of the
molstic-S. This will leave a female VCR swivel nut
(5) as the molstic interface. Retain the tube adaptor
for future use.
The VCR gaskets used in this
connection are for one-time use and should
therefore be discarded.
Place a new VCR gasket securely against the
sealing surface inside the female VCR swivel nut .
Align the filter downstream fitting (4) in the VCR nut
(5) and thread the nut onto the fitting by hand. The
filter should be installed with the printed arrow
pointing in the direction of the gas flow.
Using two wrenches, tighten the filter's downstream
male VCR connector (4) into the female nut (5) 1/8
turn past finger-tight.
Thread the 8VCR female coupling (2) onto the filter
upstream connector (3). Place a new VCR gasket
securely against the sealing surface of the filter
upstream fitting inside the 8VCR female coupling.
Thread the VCR x tube adaptor fitting (1) from the
original molstic inlet into the 8VCR female coupling
(2). The coupling should be threaded onto both
male fittings so that the VCR gasket is centered in
the coupling.
Using two wrenches, tighten the two male nuts from
the tube adaptor (1) and filter upstream fitting (3)
against each other in the coupling. The connection
should be tightened 1/8 turn past finger-tight.

To temporarily install the filter kit:

Using a pair of properly sized wrenches, remove the
VCR x tube adaptor that is on the inlet end of the
molstic-S. This will leave a female VCR swivel nut
(5) as the molstic interface. Retain the adaptor for
future use.
The VCR gaskets used in this
connection are for one-time use and should
therefore be discarded.
Place a new 8VCR o-ring securely against the
sealing surface inside the female VCR swivel nut.
041215
Tel 602.431.9100
Fax 602.431.9559
dhi@dhinstruments.com
www.dhinstruments.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 1-2 IN. FILTER KIT and is the answer not in the manual?

Questions and answers

Summary of Contents for DH Instruments 1-2 IN. FILTER KIT

  • Page 1 Installation of 1/2 in. Filter Kit P/N 401888 Instruction Sheet This kit is used to provide filtration for 1/2 in. molbloc To permanently install the filter kit: flowpaths, especially for 1/2 in. molstic-S and molbloc- Using a pair of properly sized wrenches, remove the S installations.
  • Page 2 Align the filter downstream fitting (4) in the VCR nut (5) and thread the nut onto the fitting so it is hand- Over-tightening VCR connections will damage the sealing beads and/or o-rings and possibly cause system leakage. Do tight. The filter should be installed with the printed not re-use metallic VCR gaskets.

This manual is also suitable for:

401888

Table of Contents