User Guidelines for
VENTURI
AVANTI / VENTURI
COMPACT
®
®
Negative Pressure Wound Therapy
Systems

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VENTURI COMPACT and is the answer not in the manual?

Questions and answers

Summary of Contents for Talley VENTURI COMPACT

  • Page 1 ...reducing avoidable harms User Guidelines for VENTURI AVANTI / VENTURI COMPACT ® ® Negative Pressure Wound Therapy Systems...
  • Page 2 Explanation of Label Symbols and Statements Protect from heat and Caution radioactive sources Refer to instructions of use / Temperature limitation booklet Class 2b Medical Device Humidity limitation - conforms with Directive (93/42/EEC) Annex II (excluding Section 4) Atmospheric pressure limitation North America ETL listed Suitable for connection to type BF applied parts...
  • Page 3: Table Of Contents

    Thank you for choosing to use the VENTURI COMPACT Negative ® ® Pressure Wound Therapy (NPWT) system from Talley. The VENTURI AVANTI and VENTURI ® ® COMPACT systems are intended for use for patients with acute or chronic wounds that...
  • Page 4: List Of Components

    Gauze Wound Care Set - portal drain (standard; large; abdominal) l Foam Wound Care Set - portal drain (small; standard; large; large/abdominal) Note: Other Wound Care Sets may be available; please contact Talley for latest information. ALSO AVAILABLE l 300ml/600ml/1200ml canister with solidifier...
  • Page 5: Cautions And Warnings Information

    flammable.
  • Page 6: How To Apply Npwt

    ® by users. Battery replacement should only be carried out by qualified Talley service personnel. l Wireless equipment such as mobile phones should be kept at least 1 foot or 0.3 metres away from this equipment.
  • Page 7: Operating The Vacuum Power Unit

    How to Apply NPWT (continued) 4. Attach Wound Care Set to the VENTURI AVANTI / VENTURI COMPACT power unit ® ® canister by lining up locator stud on the tubing connector with the notch on the canister tubing port located on top corner of canister, pushing down gently and twisting clockwise to lock.
  • Page 8: User Information

    How to Apply NPWT (continued) 8. Canisters should be replaced as required or weekly. To change canister, make sure power unit is in stand-by mode (if still running, press and hold the UNLOCK button, followed by the RUN/STOP button). Clamp Wound Care Set tubing and remove by turning anticlockwise and lifting out of tubing port (this can be reconnected to new canister and unclamped if wound dressing is not being changed).
  • Page 9: Button Functions

    User Information (continued) BUTTON FUNCTIONS RUN/STOP Press to invoke stand-by mode prior to running power unit. Press again to run power unit. Press and hold whilst power unit is running (and unlocked) to cease operation and return to stand-by mode. In battery operation the power unit will then power off after 1 minute of inactivity.
  • Page 10: Battery Information

    Use only the mains adapter supplied with the system. l When power unit operation times when running from the internal battery are noticeably shorter than normal, it is time to replace the battery pack. Contact Talley or authorised dealer for battery replacement service.
  • Page 11: Care And Maintenance

    Talley personnel to bag the equipment if the device has been in a known or suspected infectious environment. TRANSPORT AND STORAGE Handle with care.
  • Page 12: Operational Conditions

    VENTURI AVANTI only)– indicates battery is not correctly fitted. The power ® unit will only operate using the power adapter whilst this message is displayed. Contact Talley or authorised dealer to check battery and installation. LOW BATTERY (only appears during battery operation) – the power unit will continue to run whilst this warning is displayed.
  • Page 13 ® icon) – indicates that the battery is not functioning correctly and cannot be used. The power unit will only operate using mains power whilst this message is displayed. Contact Talley to order a replacement battery. If any of the following faults are displayed the power unit will cease to operate:- PUMP OPEN;...
  • Page 14: Specification

    4 metres Part Number: 10244 10245 The above mains adapters are considered part of the ME equipment The VENTURI AVANTI / VENTURI COMPACT power units must only be used with the ® ® specific external power adapters as supplied by Talley.
  • Page 15: Emi/Emc Statement And Manufacturer's Declaration

    EMI/EMC Statement and Manufacturer’s Declaration This equipment has been tested and found to comply with the limits of EN 60601-1-2. These limits are designed to provide reasonable protection against harmful interference in both a medical and residential environment. This equipment generates, uses and can radiate radio frequency energy and, if not used in accordance with manufacturer’s instructions, may cause harmful interference to radio communications.
  • Page 16 Every care has been taken to ensure that the information contained in this manual was correct at the time of going to press. However, Talley reserves the right to modify the specification of any product without prior notice in line with a policy of continual product development.