Page 2
• 如需其他語言,請參閱 www.aquafineuv.com。 • 他の言語については、www.aquafineuv.com をご覧ください。 • 대체 언어를 확인하시려면 www.aquafineuv.com 을 방문하십시오 . • Fassungen in anderen Sprachen finden Sie unter www.aquafineuv.com. So, you can see that it's simply not worth the risk! Insist on genuine AQUAFINE replacement lamps.
Safety Information Section 1 Safety Information These are the original instructions. Please read this entire manual before operating this equipment. Pay attention to all danger, warning, and caution statements in this manual. Failure to do so could result in serious personal injury or damage to the equipment.
Safety Information WA R N I N G • During extended periods of no water flow, the water in your chamber can become very hot (Approx. 60 °C) and potentially lead to scalding. It is recommended to run your water until this hot water has been purged from your chamber. Do not allow water to contact your skin during this time. To eliminate this condition, a temperature management valve can be installed at the outlet of your UV system.
General Information your UV system. These water quality parameters can be tested by your local dealer, or by most private analytical laboratories. Proper pre-treatment is essential for the UV system to operate as intended. Section 2 General Information Figure 1 System Components Item Description Part Number...
Installation Section 3 Installation 3.1 UV System C A U T I O N Electronic controller must be connected to a Ground Fault Protected Circuit (GFCI) receptacle and the lamp connector ground wire connected to the stainless steel chamber. The UV system is designed to be mounted either horizontally or vertically. If installing the chamber in horizontal position the outlet port must be pointing upwards to ensure all air is fully purged from the chamber.
Page 7
Installation the system requires a supplementary disinfectant for the distribution system if any water is used during by-pass condition. Min. clearance equal Unit overall length to unit overall length Figure 3 UV System Installation - Vertical and Horizontal 3. Select a suitable location for the UV system and its related components. As it is recommended to install a GFCI, make sure that this is taken into consideration prior to any installation.
Installation a connection point or a “sweating” system. Make sure you allow for a “drip-loop” as shown in Figure 4 on the lamp, sensor, and power cord, again, to prevent any water from potentially entering the controller. Figure 5 Drip Loop 6.
Maintenance Section 4 Maintenance WA R N I N G • Always disconnect power before performing any work on the UV system. • Always shut-off water flow and release water pressure before servicing. • Regularly inspect your UV system to ensure that the power indicators are on and no alarms are present. •...
Page 10
Maintenance 4.2 Cleaning and Replacing Quartz Sleeve Note: Minerals in the water slowly form a coating on the lamp sleeve. This coating must be removed because it reduces the amount of UV light reaching the water, thereby reducing performance. If the sleeve can not be cleaned, it must be replaced. Prerequisites: •...
Page 11
Maintenance 4.3 Cleaning and Replacing UV Sensor WA R N I N G The UV sensor is very delicate instrument. Extreme care is required when handling and cleaning. The sensor window itself is constructed from quartz which is extremely fragile, be careful you do not chip or break this quartz window. Manufacturer’s warranty does not cover damage due to neglect or misuse.
Operation Section 5 Operation WA R N I N G The advanced warning system has been installed to provide the optimum protection against microbiological contamination in water. DO NOT disregard the warning signals. The best way to ensure optimum UV performance is to have the water microbiologically tested by a recognized testing agency on a regular basis.
Page 13
The AQUAFINE monitored systems incorporate a UV sensor which detects the discrete 254 nm wavelength of the UV lamp. This information is relayed to the AQUAFINE controller and is the default display shown in “% UV output”. The system will display the UV output between 80 to 99 percent. When the system drops below 80%, a low UV warning is displayed as and alternately flashes (at 2 second intervals) back to the actual UV level.
Page 14
Refer to Section 5.1.5 for explanation of this feature. Note: On the AQUAFINE systems, the audible tone provided for lamp failure is a continuous alarm, rather than the intermittent (1 second on, 1 second off) condition on the basic AQUAFINE systems.
Page 15
5.2.5 Auxiliary and Solenoid Output Working in conjunction with the UV intensity monitor, the AQUAFINE controller provides a powered (line voltage) IEC solenoid output connector. (Note: this is not a dry contact) This solenoid output is protected with a replaceable 2 amp isolated fuse.
Troubleshooting Procedure: • Attach the “male” end of the • Connect the ‘Y’ cable to the • Disconnect the existing removed sensor cable to the sensor cable from the controller. “female” end of the new “Y” controller. cable. • Appropriately attach the 4-20 mA cable to the applicable equipment and ensure all connections are hand-tighten.
Refer to Figure xx for components that are accessible for maintenance. N O T I C E • Always using genuine Aquafine parts keeps your warranty, UL & CE certifications valid. • Obey all warning and caution statements. Refer to Section 1 on page •...
Page 18
Manufacturer’s Dose Flow Chart Section 7 Manufacturer’s Dose Flow Chart Note: Not performance tested or certified by NSF.
Aquafine will either, at Aquafine's expense, ship a repaired, reworked or new part to the Customer who will install such part as directed by Aquafine or will direct Customer to acquire, at Aquafine's expense, such part from a third party and then install such part as directed by Aquafine.
Need help?
Do you have a question about the VL410 Series and is the answer not in the manual?
Questions and answers