Page 1
Models: Consoles: T402, T403 Modules: TS410, TS420 & SP430 AMERICAS EUROPE ASIA/PACIFIC JAPAN Transonic Systems Inc. Transonic Europe B.V. Transonic Asia Inc. Nipro-Transonic Japan Inc. 34 Dutch Mill Rd Business Park Stein 205 6F-3 No 5 Hangsiang Rd 7th Floor, Maruha Building...
Table of Contents Table of Contents ..................................ii Warnings & Precautions .................................iv I. Introduction ..................................1 II. T402 & T403 Multi-channel Consoles ..........................2 A. T402 Multi-Channel Bench-top Console ............................. 2 B. T403 Multi-Channel Bench-top Console ............................. 2 C. Console Care ..................................... 2 i.
● Transonic ® Research Consoles and Modules, and compatible Probes, Sensors, Transducers and Catheters are designed only for investigative use with animals and are not for use in humans. Contact Transonic Systems Inc. for alternative products for human use applications.
“Sensor” and “Flowsensor” refer to Transonic ® Clamp-on and Inline Tubing Flowsensors. NOTE: “Transducers” refer to Pressure Transducers including Transonic Scisense Pressure Catheters and/or Transpac IV Pressure Transducers. “Catheters” refer to Transonic Scisense Pressure Catheters. ® Mix and match the measurement capabilities you need in a single instrumentation Console. T402 and T403 line cord powered Consoles are multi-channel capacity cases with a shared universal power supply and back-panel analog outputs compatible with most data acquisition systems.
II. T402 & T403 Multi-channel Consoles All 400-Series Consoles are line powered with convenient carrying handles and tiltable front feet. Consoles are not stand alone pieces of equipment; a properly connected, compatible Module is required to make measurements. A. T402 Multi-Channel Bench-top Console Four-Bay Console Holds up to two double-bay wide Modules, one double-bay Module and two single-bay Modules or a combination.
● If the fuses are blown, replace with 0.8 Amp fast blo fuses (Bussman # GMA0.8, 250 VAC). ● If the Console still does not power on, contact your local Transonic ® representative. AU-OPR-400Ser-EN,...
iii. Consoles: Specifications WEIGHT/ SIZE T402 Four Bay Console ● 5.21” h x 9.25” w x 12” d ● 5.8 lbs T403 Six Bay Console ● 5.21” h x 13.46” w x 12” d ● 7.6 lbs Consoles have side panel handles and tiltable front feet for easy viewing MODULE COMPATIBILITY Accepts 400-Series Flow and Pressure Amplifier Modules ELECTRICAL...
96-pin connector. The Module will click into place so that the Module is slightly recessed in the Console. ● Lock into place by tightening the thumbscrew. ● Cover any empty Module slots with a blank panel cover from Transonic ® . Align the cover holes with the console top and bottom horizontal rails and screw into place.
Consoles: Directions For Use 2. SEQUENTIAL TRIGGERING Interconnect 2 or 3 Flow Modules on the rear panel with jumpers between Modules (pins 11 & 12) from “Synch Out” to “Synch In” as shown (Fig. 2.4) & (Fig. 2.5). This wiring avoids the flow offsets that could result from ultrasonic cross- talk between adjacent Probes/...
III. TS410 Tubing Flow Module The TS410 Tubing Flow Module measures a single channel of volume flow using ultrasonic transit-time technology. The Module operates both Inline and Clamp-on Flowsensors for measuring volume flow in flexible plastic tubing circuits. The Flowsensors can be calibrated for and used with most non-aerated liquids including but not limited to: blood, saline, water, cell culture, physiological buffers, and blood analogs such as glycerine/water solutions.
i. TS410 Module: Functions & Controls A. Front Panel Signal Quality LED Display Indicator LCD Display Sensor Status / Meter Status / Scroll Button Scroll Button Menu / Mute / Exit Button Select Button Filter Selection Button Zero Adjust BNC Flow Output Sensor Connection Attachment screw Fig.
160 Hz 500 Hz D. Analog Outputs Transonic Flow Modules output analog signals in the range of -5 volts to +5 volts that are compatible ® with most A/D (analog-to-digital) converters of Data Acquisition Systems. The back panel analog outputs...
ii. TS410 Module: Specifications ● Calibrate Scale: Sets output to 0 and 1 Volt to GENERAL FEATURES Size: 5.125” h x 4” w x 9.062” d calibrate external recording devices with scale Weight: 2.3 lbs. factor flow. Module fits 2 Console bays (20HP) in T402 or T403 ●...
TS410 Module: Functional Tests These functional tests are suggested to acquaint a new user with Transonic ® TS410 Flow Modules and Transonic Tubing Flowsensors and to check for damage incurred during shipment. If the apparatus does ® not function as described during this initial operation, please call Transonic ®...
TS410 Module: Functional Tests C. User Adjusted Gain The user may adjust the calibration gain if using different fluids, temperatures or variables from the pre- programmed calibrations. 1) Set up the tubing calibration system with the desired parameters. 2) Choose the calibration setting which most closely matches the system setting. 3) Nullify the zero offset as described in “Flowsensor Zero Offset”...
iv. TS410 Module: Directions For Use A. Program Menu Press [Menu] to enter the program, this makes the blue button labels active. Use arrows to [Scroll] through menus and to set values for alarms and gain (pressing and holding scrolls faster). Press [Select] to enter menus or toggle settings on and off.
C. Use of Clamp-on Flowsensors SENSOR SELECTION Flowsensor size and calibration is determined by the OD of the tubing on which it will be used. Each Transonic ® Clamp-on Tubing Flowsensor is custom calibrated for use on particular tubing (silicone rubber, latex, polyurethane, or polyvinyl chloride {PVC}).
Page 19
TS410 Module: Directions For Use SITE SELECTION The Flowsensor should be applied on straight tubing segments, about 10X the ID from side branches to produce measurements within its accuracy specifications. The best application site for the Flowsensor is a point well below the highest elevation of the tubing (where gas bubbles might lodge). The Flowsensor will deform the tube slightly.
TS410 Module: Directions For Use D. Use of Inline Flowsensors SENSOR SELECTION Flowsensor size is determined by the tubing with which it will be used. Inline Flowsensors are sized based on the ID of the system tubing. SENSOR CALIBRATION Flowsensors are precalibrated for a particular liquid at a certain temperature (see Flowsensor’s calibration certificate).
IV. TS420 Perivascular Flow Module The TS420 Perivascular Flow Module measures a single channel of volume flow using ultrasonic transit-time technology. The Module operates a range of Perivascular Flowprobes for measuring volume flow in arteries, veins and ducts. Probes are available for both acute use and chronic implantation.
i. TS420 Module: Functions & Controls A. Front Panel LED Digital Display Signal Quality Analog Display Indicator Filter Selection Button Mode Selection Button Low Flow Range Invert Button Button Calibration Key Port BNC Flow Output Attachment screw Fig. 4.8: TS420 Front Panel Zero Adjust Probe Connection MODE SELECTION BUTTON...
160 Hz 500 Hz D. Analog Outputs Transonic ® Flow Modules output analog signals in the range of -5 volts to +5 volts that are compatible with most A/D (analog-to-digital) converters of Data Acquisition systems. The back panel analog outputs...
ii. TS420 Module: Specifications GENERAL FEATURES ZERO ADJUST Size 5.125” h x 4” w x 9.062” d Recessed momentary push button to zero flow Weight: 2.2 lbs. reading at stopped flow Module fits 2 Console bays (20HP) in T402 or T403 Power: Derives input power from 400-Series LOW RANGE Consoles.
TS420 Module: Functional Tests These functional tests are suggested to acquaint a new user with Transonic ® 400-Series Flow Modules and Transonic Perivascular Flowprobes and to check for damage incurred during shipment. If the apparatus ® does not function as described during this initial operation, please call Transonic ®...
Put the Flow Module in [Scale] mode to display the value of the Probe’s 1 volt reference signal in the normal or low flow range. This reference signal is used to calibrate the data acquisition system and does not change the factory calibration of the Flowprobe. All Transonic ®...
Probe may exhibit when applied to a vessel. If only chronic measurements are needed, the 10 minute Probe soak may be omitted. The Flowprobe has been factory calibrated to meet Transonic ®...
Page 28
Standard Probe/reflector combinations come factory-prepared with suture holes and eyelets to allow for this suturing. The proper method to secure a Probe in place depends on the application site: Transonic ® Surgical and Technical Application Notes offer some tried and proven methods.
(from the front). Transducers The SP430 Pressure Module is equipped to work with Transonic Scisense solid-state Pressure Catheters. The Module is also compatible with Transpac IV pressure transducers Transonic PART # YS100 when used with the YS100-ADPT. 1. TRANSONIC SCISENSE PRESSURE CATHETERS Transonic Scisense solid-state Pressure Catheters provide high fidelity pressure measurements directly at the source.
Acceptable range: -0.560 V to -0.582 V Acceptable range: -0.560 V to -0.582 V C. Analog Outputs Transonic ® Pressure Amp Modules output analog signals in the range of -5 volts to +5 volts that are compatible with most A/D (analog-to-digital) converters of data acquisition systems. The back panel analog...
2 may result in excessive noise in the signal. CATHETER/TRANSDUCER COMPATIBILITY Accepts Pressure input in 2 front panel HDMI connectors. Each channel has separate controls and outputs. ● Transonic Scisense solid-state Pressure Catheters with HDMI connectors ● Transpac ®...
5) The analog output should be at -0.571 V (Do not adjust the balance control knob). NOTE: If the output at 100 mmHg is significantly different, do not use the Pressure Transducer and contact your local Transonic ® representative for troubleshooting or repair return instructions.
(i.e. a saline filled syringe) and allow a warm up time of 30 minutes while preparing the animal for surgery. Fig. 5.5: Transonic Scisense 5F FDH Pressure Catheter 2) Connect the Catheter to the channel 1 input using the supplied HDMI cable.
® IV Pressure Transducers A disposable Transpac ® IV Pressure Transducer for a fluid-filled arterial or venous catheter can be connected to the Module using a RJ11 to HDMI adaptor (Transonic ® part YS100-ADPT). NOTE: Use of the Transpac ®...
● Consoles & Modules are warranted for a period of 12 months after date of shipment ● Flowprobes & Flowsensors are warranted for a period of 6 months after date of shipment ● Transonic Scisense Pressure Catheters less than two French (<2F) are warranted for a period of 12 months after date of shipment ●...
The Repair Return Form b) Power cord c) Additional Probes, Sensors, cables or other equipment if requested by the Service Technician B. Flowprobes, Flowsensors & Transonic Scisense Pressure Catheters 1) Call your Transonic ® representative or contact your regional office, listed below, with a description of the problem with the Probe, Sensor or Catheter.
Electronic ultrasonic circuitry directs a Flowprobe through the following cycles: Fig. D.1: Schematic views of a Transonic ® perivascular UPSTREAM TRANSIT-TIME MEASUREMENT CYCLE ultrasonic volume Flowsensor.
Appendix B: EMC Tables The 400-Series System complies to the Standard for Safety Electrical Measurement and Test Equipment Part 1: General Requirements UL 3111-1 and Safety Requirements for Measurement, Control and Laboratory Use Part 1: General Requirements CSA C22.2 No. 1001.1-92 in accordance with UL 61010A-1 Electrical Equipment for Laboratory Use, Part 1: General Requirements.
See the Declaration of Conformity accompanying this device for specific directives. ETL Testing Mark Electrical Safety Compliance Certification Authorized Transonic Systems Europe is an authorized representative in the European Representative Community TUV Testing Mark Electrical Safety Compliance Certification This device was manufactured in accordance with RoHS requirements for RoHS Compliant applicable hazardous material content.