ROVIDER OF Label Vision Systems 101 Auburn Court RINT UALITY NSPECTION YSTEMS Peachtree City, GA 30269 Phone: +1-770-487-6414 LVS® IS AN ISO 9001:2008 REGISTERED COMPANY Toll Free: +1-800-432-9430 Fax: +1-770-487-0860 E-mail: info@lvs-inc.com Visit us online at www.lvs-inc.com Approved: Ray Vaughan...
Page 2
LVS®. Any use of this publication and related materials beyond the terms of said agreements is prohibited, and LVS® reserves all rights in this publication and related materials. The name LVS® is a registered trademark of Label Vision Systems, Inc. Copyright Copyright 2013 Label Vision Systems, Inc.
Label Vision Systems, Inc. INTEGRA 9505 Product Discontinuation LVS® is discontinuing the INTEGRA 9505 after introduction of the INTEGRA 9510 to the marketplace. The INTEGRA 9505 will be supported for four years until December 31 , 2015.
Page 9
DO NOT discard this shipping carton in case you must ship or store the system for any reason. Failure to use this carton when returning your product to LVS® will void warranty. IMPORTANT: All INTEGRA 95XX systems must use the calibration card and installation CD issued with the system.
Page 10
After installing the INTEGRA 95XX software, an “LVS Computer Desktop Manuals” icon is saved to the computer’s desktop. Double- click the “LVS Manuals” desktop icon to access the system manuals. Note: Manuals Located on Installation CD: When inserting the installation CD into the CD drive with Autorun activated, screens automatically appear prompting you to install the INTEGRA 95XX software.
INTEGRA 95XX software version 3.0.8 or One of the following: higher. EAN/UPC Calibrated Conformance Test Card (LVS® part # CAL002) GS1-128 Calibrated Conformance Test Card (LVS® part # CAL003) Specifications subject to change without notice. Approved: Ray Vaughan...
Calibration Minimum PC Requirements EAN/UPC Calibrated Conformance (PC Supplied by User) Test Card (LVS® part # CAL002) Windows® XP Professional or Windows® 7 (Windows® Vista is not supported) Specifications subject to Intel® Core™ 2 Duo Processor (or equivalent) change without notice.
Page 26
Double-clicking on the term Version (see Figure above) allows you to view a complete list of the latest software versions. After double-clicking on this term, the following page appears (see Figure below). This information is important if it is necessary to contact the LVS® technical helpline. Approved: Ray Vaughan Effective: 03-June-2013 Supersedes: 18-April-2012 LVS®...
This feature is an upgraded feature and is an additional cost to customers. Multiple sectors verification steps are outlined in “Appendix G: Special Features.” Contact LVS® or an LVS® distributor for more information.
It is important to place the Calibration Test Card in a clean place and away from direct sunlight. If the test card is nicked, soiled, or damaged in any way, it must be replaced. Contact LVS® or your LVS® representative to obtain a new Calibration Test Card.
Judge” within the tolerances stated in ISO 15426-2. The purchase of the Data Matrix Calibrated Conformance Test Card is optional. If you wish to purchase a card, please contact Label Vision Systems or an LVS® representative. IMPORTANT: Please read the document entitled “Read Me First” included with the Data Matrix test card as it provides detailed instructions about the test card.
10-Line Grading Button Because the INTEGRA 95XX Series uses camera-based technology, LVS® is able to analyze many lines throughout the height of a code. The software averages all parameters for each line and reports the overall grade as being the lowest of the averages of all parameters.
Application Identifiers and with data semantics. It is not the intention of this manual nor Label Vision Systems, Inc. to educate an operator how to translate the meaning of the structure analysis. It is only LVS®’s intention to divide the information into sections in accordance with industry standards.
Data Structures but we may make an occasional error or may not have a Data Structure defined that you require. If this is the case, please contact LVS® and we will make the needed changes and/or additions in a future release of the software.
INTEGRA 95XX software for the purpose of analyzing and troubleshooting the image. This feature is used for remote troubleshooting assistance when an image needs to be analyzed by an LVS® technician or other employees internal to your company who also have an INTEGRA 95XX system.
Export Image to File The “Export Image to file” button is used for remote troubleshooting assistance when an image needs to be analyzed by an LVS® technician or other employees internal to your company using an INTEGRA 95XX system. Click this button to export an image, which can then be imported into another computer housing the INTEGRA 95XX software for the purpose of analyzing and evaluating the image.
Page 97
Yes, but is recommended because reports, settings and other information not copied to the new database will be lost and some settings require you to contact LVS® for the password of the day to restore the settings. Click No to not copy reports, settings and other information from the old database to the new database.
Replace the Calibrated Conformance Standard Test Card every two years. The INTEGRA 9510 light tray has LED lights which have a life span of approximately 10,000 hours; thus, there should be no replacement concerns with the light tray. Contact LVS® or your distributor with any questions. INTEGRA 9570 Sensor Cleaning Instructions: The sensor window may need to be cleaned daily, depending on use.
Page 101
This streak may cause the bar code grade results to be distorted if the streak is located within the bar code image. When a scratch occurs, you may need to return the system to LVS® (or your LVS® distributor) for sensor replacement. Contact your LVS® representative for more information.
Page 102
The INTEGRA 9500 light tray has LED lights which have a life span of approximately 10,000 hours; thus, there should be no replacement concerns with the light tray. Contact LVS® or your distributor with any questions. Replace the Calibrated Conformance Standard Test Card every two years.