IOP 019A.1
Effective Date: 1/11/2017
Replaces IOP 019A.0
Page 1 of 11
INSTRUMENT:
Multiparameter Water Quality Instrument
MODEL:
-
Pro Plus Meter (6050000)
-
Quatro Cable (605790)
-
Probes: temperature/conductivity (5560), pH (1001 or 1001A), DO (2003), ammonium (1004)
MANUFACTURER:
Yellow Springs Instrument Co., Inc. (YSI)
PRECAUTIONS:
POTENTIAL INTERFERENCES
General –
-
Greatest accuracy is achieved if the actual samples to be measured are within 10 ˚C of the
calibration solutions.
-
Longer cable lengths introduce more error into the readings.
INSTRUMENT OPERATING PROCEDURE
U.S. Fish and Wildlife Service
Marquette Biological Station
3090 Wright Street
Marquette, Michigan 49855
U.S.A.
and
U.S. Fish and Wildlife Service
Ludington Biological Station
5050 Commerce Drive
Ludington, Michigan 49431
U.S.A.
and
Department of Fisheries and Oceans
Sea Lamprey Control Centre
1219 Queen Street East
Sault Ste. Marie, Ontario P6A 2E5
Canada

Advertisement

loading
Need help?

Need help?

Do you have a question about the 2003 and is the answer not in the manual?

Questions and answers

Summarization of Contents

Procedures and Configuration
Configuration and Instrument Overview
Details the instrument's configuration, including meter and cable components.
Key Pad Functions
Explains the function of each button on the YSI Pro Plus meter keypad.
Meter Preparation and Software Setup
Covers battery installation, software setup, and sensor configuration.
Sensor Installation and Cable Attachment
Provides detailed steps for installing various sensors and attaching the cable.
Calibration Procedures
pH Sensor Calibration
Step-by-step guide for calibrating the pH sensor using buffer solutions.
Ammonium Sensor Calibration
Procedure for calibrating the ammonium sensor using calibration standards.
DO Sensor Calibration
Instructions for performing calibration of the dissolved oxygen sensor.
Measurements and Storage
Taking Measurements
Guidance on how to take accurate water quality measurements with the instrument.
Sensor Storage Guidelines
Procedures for short-term and long-term storage of instrument sensors.