Page 1
A G N E T O S T R I C T I V E T R A N S D U C E R S S E R I E S P C O D E L S : P C S - P C P - P C R - P C M DS EUROPE DS EUROPE DS EUROPE DS EUROPE S.r.l.
Page 2
V 1.0 Rev 3. The information in this manual can be subjected to modifications without prior notice. DS Europe refuses any responsibility for possible errors contained within, or possible direct or indirect accidental damages deriving from supplier, from the operation or the use of this material.
A CURRENT OUTPUT ..................21 10.6.1. CURRENT CONNECTION: 4÷20 mA WITH VOLTAGE READING: 1÷5 V........21 10.6.2. CURRENT CONNECTION ........................21 10.7. SERIAL PORT CONNECTION........................22 10.7.1. TERMINATION RESISTANCE......................23 10.8. GROUNDING ...............................23 Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Page 4
TESTS FOR APPLICATION OF CE MARK....................38 16.1. TEST CENTRE AND LIST OF TESTS......................38 16.2. TEST RESULTS AND LAW ON CONFIDENTIALITY................38 IMPORTANT: always bench check the transducer operational performance before installation. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Supplier. It is clear that under these conditions, DS Europe is forced to refuse any responsibility regarding use and thus limits itself to listing some fundamental and elementary precautions to ensure correct use.
Permanent external magnetic fields generated by other magnets or residual magnetic fields, generated from the cursor coming into occasional contact or proximity of magnetizable materials (paramagnetic materials), can effect the measurement. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
(section. 3.2). The recommended distances (= limit) are indicated in Fig. 2 (mm). A max. = 1/8 of the diameter (outside ) of the cursor Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
According to the laws of physics regarding attraction of opposing poles, the North pole of a magnet attracts the South pole of a compass. When necessary the North pole is marked with a coloured dot. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
IMPORTANT: Cursors, magnets or float, not produced by DS Europe and chosen by the Customer must be sent to DS Europe in order to be checked. DS Europe, if possible, will custom calibrate the transducer with them. Customer must tell to DS Europe the application conditions: used materials , pole position, distance between magnets and probes etc..
The mod. AFR-1E can pass over these two “A” and “B” death zones without causing any damage but It will not generate any signal (analog outputs will be positively or negatively saturated while the RS485 will give a “9999999” output). For performance optimisation, the Customer must supply DS Europe with all the installation conditions 5.4.
It is to be considered the possibility of using a magnetic cursor with a radial slotting system (Mod. ARR-2E) through which It can pass the probe anchorage. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
10 mm per metre will irreversibly damage the wave guide. This kind of damage can be easily found out by DS Europe and is not covered by the warranty. NOTE: during installation or removal of a transducer equipped with threaded union (PCS, PCP, PCM), use only the hexagon for clamping.
(operating system and hardware) and the program structure used to interrogate the transducer. For additional informations about serial communications protocol refer to Chapter 13. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
As soon as the transducer is turned on, it detects how many magnets are present on the probe and automatically sets the analog outputs enabled by the user during the personalisation procedure. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
This setting of 0÷10 V can be inverted to 10÷0 V ( It is possible also to do the personalisation for the current outputs). For direction settings (if not specified in the order) see chapter 13. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
0,05 m/sec, we would have a corresponding significant measuring step of 0,05 mm. When FS= 350 mm there is a corresponding sampling frequency = 1000 samples / sec. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
In the following text, the indications refer to the cable leads (option C, output with cable); in brackets (…) we have indications about connector pins (option P, connector output). Example: WHITE (PIN3) means: WHITE output with cable, while (PIN3) refers to output with connector. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Incorrect connections can be easily found out by DS Europe and can cause serious damage to the transducer. The warranty does not cover this kind of damage. 10.1. CABLE OUTPUT Table 1 indicate the color codes of conductors for the output cable.
To connect 0÷10 V analog voltage output, connect the YELLOW wire (PIN5) to the negative input of the multi-meter, or the data acquisition card (GND) and the WHITE wire (PIN3) to the positive. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Connect the WHITE wire (PIN3) to the positive of the first multi-meter and the GREEN wire (PIN4) to the positive of the second multi-meter. figure 8: Electrical connection for two analog voltage outputs. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
250 ohm resistor. Set the multimeter in order to read the 4÷20 mA current. figure 9: Electrical connection of two 4÷20 mA outputs with a voltage reading of 1÷5 Vcc. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
4. Check that the parameters of the serial communications port (computer) correspond to those indicated in chapter 9.7; 5. Check computer serial port performance using another computer (also fitted with RS,485 MULTI-DROP interface) Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
WARNING : The common of the power supply ( “ - “ power supply ) must be grounded in order to avoid damages due to mains fluctuations. For further details see chapter 11 on electromagnetic interference. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
The blinking is like a “binary” code where the beginning of the status blinking is a long pulse and between X and Y codes there is a pause with LED off. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Both permit you to calibrate transducer in a simple way. Using the serial connection makes everything easier as you have the added advantage of being able to personalise the analog output performance. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Page 26
Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Important: during calibration phase, avoid: • Introducing or touching components with metal tools; • Using or inserting liquid of any kind whatsoever; Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Cursor “1: D/A1 fluctuation for ZERO 7,5 V 15 mA 16 mA Cursor “1”: D/A1 fluctuation for FULL SCALE 7,5 V 15 mA 16 mA 10 V 20 mA 20 mA Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Page 29
It mantaining at least a 10% Full and to activate the just done Scale distance between zero and calibration. Full Scale points. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Casing screws Casing screws figure 14: Lower view of container for housing PC Series electronics figure 15: Detailed view of push buttons and LEDs used for calibration. One of the advantages of using the series PC is the possibility to be able to use only one part of the nominal stroke of the transducer without having to alter the stability and resolution of the transducer.
ASCII characters starting from the fourth position in the message. The end of message must always end with “<cr>” (Carriage Return, 0x0D). Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
In the example, the D/A 2 (1001) converter is configured as on, direct, representing the displacement measure for cursor 1 while the D/A 1 (1000) is configured as on, direct, representing the displacement measurement for cursor 0. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
“R1” when you request the transducer for the second cursor value (cursor 1). Example 2: @ <cr> 14.3.5. COMMAND “V” Command “V” requests revision of firmware installed into transducer. Example: <cr> Example of a reply: PC V.01.00 S/N xxxxxx Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
@0L0Lxxxxxx High Limit 0 Cursor 0 FULL SCALE reference @0L0Hyyyyyy Minimum Calibration 0 Internal count value minimum 0 @0T0Z Maximum Calibration 0 internal count value maximum 0 @0T0F Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Page 35
In the example the contents of the 0 parameter are read (cursor 0 ZERO refernence) Note: “xxxxxx”, “yyyyyy”, “zzzzzz”, “wwwwww” are 6- figure subjects of commands used to program the transducer configuration. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
Eg.: 1X0001000 The parameter value has no empty spaces. Whenever it has fewer characters than those established in the defined format, zeros are used to fill the spaces. Series PC Manual V 1.0 - Rev. 3 – 03/01/01 DS Europe...
F. Russoli, 6 Milan (Italy) Type of equipment: Position and magnetostrictive level transducers Models: PC series ( Mod. PCS – Mod. PCP – Mod. PCR – Mod. PCM ) With analog and digital outputs. Year of marking 1999 Th equipment has been tested with the typical installation configuration, as prescribed by the product instruction manual.
Any units which prove to be defective, despite the above mentioned controls, shall be repaired or replaced free of charge by DS Europe, on the conditions that the units have been subject to normal or correct use and in any accordingly with the use described in the...
Page 39
NOTE Series PC Manual V 1.0 - Rev. 2 - DS Europe...
Need help?
Do you have a question about the PCM and is the answer not in the manual?
Questions and answers