RESmart GII BPAP System Y / U Series User Manual V2.1 Table of Contents 1. Symbols ·································································································· 1 1.1 Control Buttons ················································································ 1 1.2 Device Symbols ················································································ 1 2. Warning, Caution and Important Tip ······························································· 3 3. Intended Use ···························································································· 3 4.
Page 4
RESmart GII BPAP System Y / U Series User Manual V2.1 15.4 Cleaning the Enclosure ···································································· 30 15.5 Cleaning the Tube ·········································································· 30 15.6 Replacing the Air Filter ···································································· 31 15.7 Disinfection ·················································································· 31 16. Traveling with the Device ··········································································· 32 17.
RESmart GII BPAP System Y / U Series User Manual V2.1 1. Symbols 1.1 Control Buttons Ramp Button Mute Button Knob 1.2 Device Symbols Follow Instructions for Use Operating Instructions Type BF Applied Part (mask) Class II (Double Insulated) For indoor use only AC Power DC Power IP22...
Page 6
RESmart GII BPAP System Y / U Series User Manual V2.1 Non-Ionizing Radiation SD Card Water Filling Prohibited Here Water Inlet Directional Indicator for Removing the Water Inlet Cap Directional Indicator for Screwing the Water Inlet Cap WEEE Marking Logo of BMC Medical Co., Ltd. 2 / 42...
• The instructions in this manual are not intended to supersede established medical protocols. • To ensure that you receive the safe, effective therapy prescribed for you, use only BMC accessories. • Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the device or MR medical devices.
RESmart GII BPAP System Y / U Series User Manual V2.1 • Cleaning and disinfection can be performed by the patient. IMPORTANT! • Read and understand the entire user manual before operating this system. If you have any questions concerning the use of this system, contact your home care provider or health care professional.
RESmart GII BPAP System Y / U Series User Manual V2.1 5. Specifications Device Size Dimensions: 170 mm × 180 mm × 118 mm, or 290 mm × 180 mm × 134 mm (with the humidifier) Weight: 1.5 kg, or 2.5 kg (with the humidifier) Product Use, Transport and Storage Operation Transport and Storage...
Page 10
RESmart GII BPAP System Y / U Series User Manual V2.1 Pressure Display Accuracy ± (0.8 hPa+4%) Static Pressure Stability ± 0.5 hPa Ramp The ramp time ranges from 0 to 60 minutes. Sound Pressure Level < 30 dB, when the device is working at the pressure of 10 hPa. Sound Power Level <...
RESmart GII BPAP System Y / U Series User Manual V2.1 6. Available Therapies The device delivers the following therapies: CPAP – Delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of pressure throughout the breathing cycle. If your health care professional has prescribed ramp for you, you can press the Ramp Button to reduce the pressure and then gradually increase the pressure to the therapeutic pressure setting so that you can fall asleep...
RESmart GII BPAP System Y / U Series User Manual V2.1 7. Glossary Apnea A condition marked by the cessation of spontaneous breathing. Auto-CPAP Adjust CPAP pressure automatically to improve patient comfort based on monitoring of apnea and snoring events. Auto Off When this feature is enabled, the device automatically discontinues therapy whenever the mask is removed.
Page 13
RESmart GII BPAP System Y / U Series User Manual V2.1 Res Rate Respiratory Rate. Number of breaths per minute. Reslex A therapy feature that is enabled by your home care provider to provide pressure relief during exhalation. Standby State The state of the device when power is applied but the airflow is turned off.
RESmart GII BPAP System Y / U Series User Manual V2.1 8. Model Product Contents Maximum Work Model Work Mode Optional Pressure Main Device Accessory (hPa) Main device Y-20T (3.5-inch CPAP, S, S/T, T LCD) Main device Y-25T (3.5-inch CPAP, S, S/T, T LCD) Main device Y-30T...
• This device should only be used with the mask and accessories manufactured or recommended by BMC or with those recommended by your prescribing physician. The use of inappropriate masks and accessories may affect the performance of the device and impair the effectiveness of therapy.
• When the insulation layer of the SpO probe cable is damaged, do not connect the probe to the patient. • Please contact BMC to buy the SD card if you need it. 10. System Features Humidifier Indicator Display Screen...
RESmart GII BPAP System Y / U Series User Manual V2.1 SD Card Slot Communications Port DC Inlet Filter Cap Fig. 10-2 Name Function SD Card Slot Insert the SD card into this slot Connected to SpO Kit, Wi-Fi Kit, Cellular Module, SpO &Wi-Fi Communications Port Kit or SpO...
RESmart GII BPAP System Y / U Series User Manual V2.1 • Tobacco smoke may cause tar build-up within the device, leading to the malfunctioning of the device. • Air must flow freely around the device for it to work properly. 11.2 Installing the Air Filter and Filter Cap (1) Attach the air filter to the filter cap, as shown in Fig.
RESmart GII BPAP System Y / U Series User Manual V2.1 WARNINGS! • The device is powered on for use when the power cord and power adapter is connected. The Knob turns the blower On / Off. • Use of the device at an AC voltage beyond the stated range (see Section 5 “AC Power Consumption”) may damage the device or cause device failure.
- Use the accompanying tube and mask provided by BMC. - Do not wear the mask for more than a few minutes while the device is not operating. - Use only masks with vent holes. Do not block or try to seal the vent holes in the exhalation port.
RESmart GII BPAP System Y / U Series User Manual V2.1 11.6 Using Oxygen with the Device Oxygen may be added at the mask connection. Please observe the instructions listed below when using oxygen with the device. WARNINGS! • Connect the oxygen tube to the oxygen inlet of the mask. •...
RESmart GII BPAP System Y / U Series User Manual V2.1 Fig. 11-9 If the SD card is inserted incorrectly or not inserted, a symbol indicating incorrect insertion or no SD card present will appear in the Main Interface on the screen of the device, as shown in Fig.
• Be sure to follow your physician’s instructions on adjusting the settings! To order any accessories not included with this device, contact your equipment supplier. • DO NOT connect any ancillary equipment to this device unless recommended by BMC or your physician. If you suffer from chest discomfort, shortness of breath, stomach bloating, or severe headache when using the device, contract your physician or qualified medical personnel immediately.
RESmart GII BPAP System Y / U Series User Manual V2.1 CAUTIONS! • You can press the Ramp Button as often as you wish during sleep. • The ramp feature is not prescribed for all users. 12.6 Turning the Device Off Take off the mask and headgear, press and hold the Knob for two seconds, and the device will stop delivering air.
RESmart GII BPAP System Y / U Series User Manual V2.1 Fig. 13-3 The first icon on the left side of the screen indicates the Main Interface, the second icon indicates the Initial Setup Interface, and the third icon indicates the iCode Interface.
RESmart GII BPAP System Y / U Series User Manual V2.1 indicating a lower humidity level. At this moment, the Humidifier option is still displayed in yellow, as shown in Fig. 13-6. Fig. 13-6 13.1.6 Confirming Adjustments Confirm your adjustment to an option by pressing the Knob .
Page 27
RESmart GII BPAP System Y / U Series User Manual V2.1 Fig. 13-9 Press the Knob , the cursor jumps to the second icon on the left side of the screen. The screen displays the Initial Setup Interface, as shown in Fig. 13-10. Fig.
RESmart GII BPAP System Y / U Series User Manual V2.1 13.2 Options of the Patient Menu and Corresponding Descriptions Option Range Description There are five humidity levels available. As the numbering Off, 1 ~ 5 Humidifier increases, the humidity rises accordingly. “Off” means the humidifier is turned off.
RESmart GII BPAP System Y / U Series User Manual V2.1 Show the relevant information about the device, this is only for Model; SN; users to view, can not be modified. Model: the device model; Firmware About SN: Serial Number of the device; Firmware version: Software version;...
RESmart GII BPAP System Y / U Series User Manual V2.1 In accordance with the requirements of the relevant standards, the volume of the audible alarm signal meets the requirements, and the sound pressure range of the measured auditory alarm signal is described as follows: Measured A-weighted sound sound...
Page 31
RESmart GII BPAP System Y / U Series User Manual V2.1 setting is “20 hPa”. Off, 5 ~ 26 hPa applies to Y-25T and U-25T, in 0.5 hPa increments, the default setting is “25 hPa”. Off, 5 ~ 31 hPa applies to Y-30T, U-30T and U-30AT, in 0.5 hPa increments, the default setting is “30 hPa”...
Page 32
RESmart GII BPAP System Y / U Series User Manual V2.1 When the airflow is on, an audible alert Middle Function will sound if the vents of the mask are Mask Blocked!! blocked; the screen will display “Mask Priority Alarm Blocked!!”...
RESmart GII BPAP System Y / U Series User Manual V2.1 14.6 Reposition of Alarming After the elimination of the alrming faults, the residual alarming information still exists (alarming information is shown on the top of the screen without any visual and auditory alarming), and turn the button leftwards or rightwards to reduce the residual alarming information.
RESmart GII BPAP System Y / U Series User Manual V2.1 15. Cleaning and Disinfection WARNINGS! • Regular cleaning of the device and its accessories is very important for the prevention of respiratory infections. • To avoid electric shock, always unplug the device before cleaning. •...
RESmart GII BPAP System Y / U Series User Manual V2.1 (2) Clean the tube in warm water which contains washing liquid, and then rinse it in clean water thoroughly. (3) After cleaning, air-dry the tube in a cool, well-ventilated area, and avoid direct sunlight. It takes approximately 30 minutes to completely air-dry the tube.
• If the device is used when the atmospheric pressure is out of the stated range (See Section 5), the accuracy of the leakage alert will be affected. (1) Use the BMC carrying case to carry the device and accessories along with you. Do not put them in your checked baggage.
19. Technical Support Please contact BMC directly if you need the circuit diagram of the device and the list of components for certain purposes such as maintenance or connection to other equipment. BMC will provide the circuit diagram and / or other technical documents in whole or in part according to your needs.
RESmart GII BPAP System Y / U Series User Manual V2.1 21. Troubleshooting The table below lists common problems you may have with the device and possible solutions to those problems. If none of the corrective actions solve the problem, contact your home care provider.
Page 39
RESmart GII BPAP System Y / U Series User Manual V2.1 Turn the humidity setting down, or raise the room temperature. When the humidifier is used, Place the tube under the quilt, or the humidified air tends to use the tube cover. Water in mask condense in the cold tube and mask if the room temperature...
RESmart GII BPAP System Y / U Series User Manual V2.1 21.2 Common Problems in the Device and Corresponding Solutions Problem Possible Cause Solution (s) Take a few deep breaths with the The Auto On / Off feature is mask on, and the device will start enabled automatically Ensure that the power cord,...
RESmart GII BPAP System Y / U Series User Manual V2.1 22. EMC Requirements Guidance and manufacturer's declaration - electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The user of the device should ensure that it is used in such an environment. Electromagnetic Environment - Emissions Test Compliance...
Page 42
RESmart GII BPAP System Y / U Series User Manual V2.1 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. Immunity IEC 60601 Compliance...
Page 43
RESmart GII BPAP System Y / U Series User Manual V2.1 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. Immunity Compliance Electromagnetic Environment -...
Page 44
RESmart GII BPAP System Y / U Series User Manual V2.1 Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Page 45
RESmart GII BPAP System Y / U Series User Manual V2.1 Recommended separation distances between RF wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment.
(1) year for main unit and three (3) months for all accessories from the date of sale by BMC Medical Co., Ltd. to the dealer. If the product fails to perform in accordance with the product specifications, BMC Medical Co., Ltd. will repair or replace, at its option, the defective material or part.
Page 48
H60 Heated Humidifier User Manual V2.0 The H60 Heated Humidifier is designed only for use with specific RESmart GII devices. Do not use the H60 Heated Humidifier with any other devices. The humidifier moistens the air delivered by the RESmart GII devices. It is for use in the home or hospital / institutional environment.
H60 Heated Humidifier User Manual V2.0 1. Warning, Caution and Important Tip WARNING! Indicate the possibility of injury to the user or operator. CAUTION! Indicate the possibility of damage to the device. IMPORTANT TIP! Place emphasis on an operating characteristic. Warnings, Cautions, and Important Tips appear throughout this manual as they apply.
Have any damaged parts replaced before continuing use. • Read all instructions before using the humidifier. • Use only with BMC devices whose instructions specify the use of this humidifier. • Please use the mask which meets ISO17510:2015.
WARNINGS! • Use the humidifier only for its intended use as described in this manual. • Use only accessories recommended by BMC. • Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the device or MR medical devices.
H60 Heated Humidifier User Manual V2.0 CAUTION! • When the main device delivers air and the humidity setting is adjusted, if the indicator lights of the humidifier do not light up, it may be that the humidifier and main device are not connected correctly.
H60 Heated Humidifier User Manual V2.0 Humidifier Uncover Button Fig. 4-5 WARNING! • Turn the device off and allow approximately 15 minutes for the heater plate and water to cool. 4.2.2 Overturning the Water Chamber Turn the water chamber over so that it is bottom up, as shown in the figure below. Fig.
H60 Heated Humidifier User Manual V2.0 Water Inlet Cap Maximum Water Level Line 1/4 Maximum Water Level Line Fig. 4-8 WARNING! • Every time before treatment, be sure to drain any residual water out of the water chamber, and ensure the maximum water level line is not submerged by water. CAUTIONS! •...
Page 55
H60 Heated Humidifier User Manual V2.0 chamber base, and pour any remaining water out of the main body of the water chamber. Undo the Water Chamber Buckle, and open the water chamber as shown below. Water Chamber Buckle Fig. 4-10 CAUTION! •...
H60 Heated Humidifier User Manual V2.0 4.4 Setting the Humidity Level After the main device is powered on, turn the Knob to turn on or turn off the humidifier and to adjust the humidity level according to instructions on the screen of the main device. There are five humidity levels available, and the number of blue indicator lights that light up is directly proportional to the humidity level.
H60 Heated Humidifier User Manual V2.0 5. Cleaning Clean the water chamber before first use of the humidifier or at least once every week. If the humidifier has not been in use for a long time, clean the water chamber before reusing it. WARNING! •...
H60 Heated Humidifier User Manual V2.0 5.3 Detaching the Air-intake Assembly After the water chamber is removed, detach the Air-intake Assembly from the main body of the humidifier by pulling it upwards, as shown in the figure below. Air-intake Assembly Fig.
H60 Heated Humidifier User Manual V2.0 (dip the soft cloth in washing liquid if necessary), rinse them throughly, and then wipe them dry with a soft cloth. Fig. 5-5 (3) Assembling the Water Chamber: Place the two parts of the water chamber together as shown in Fig.
H60 Heated Humidifier User Manual V2.0 water chamber (dip the soft cloth in mild scrubbing solutions if necessary), then rinsed thoroughly, and at last wiped with soft cloth. Top Cover Main Body Fig. 5-8 5.7 Reassembling the Humidifier (1) Set up the air-intake assembly: First install the sealing elements to the air intake, as shown in the figure below.
H60 Heated Humidifier User Manual V2.0 Fig. 5-11 (4) Connect the top cover and main body of the humidifier properly, as shown in the figure below. Fig. 5-12 6. Disinfection Generally speaking, if you have strictly followed the above cleaning instructions, you do not have to disinfect the humidifier.
The humidifier does not require routine servicing. If the humidifier malfunctions, contact your home care provider immediately. Never attempt to open the humidifier’s enclosure. If necessary, contact your local authorized dealer or BMC Medical Co., Ltd. for technical support and documents.
H60 Heated Humidifier User Manual V2.0 Pressure Drop with Humidifier < 0.4 hPa at 60 LPM flow Humidifier Performance Humidity Output: No less than 10 mg H Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity Maximum Delivered Gas Temperature < 43°C The Form and the Dimensions of the Patient Connection Port The 22 mm conical air outlet complies with ISO 5356-1.
H60 Heated Humidifier User Manual V2.0 11. EMC Requirements Guidance and manufacturer's declaration - electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The user of the device should ensure that it is used in such an environment. Electromagnetic Environment - Emissions Test Compliance...
Page 65
H60 Heated Humidifier User Manual V2.0 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. 60601 Compliance Electromagnetic...
Page 66
H60 Heated Humidifier User Manual V2.0 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. Immunity IEC 60601 Compliance...
Page 67
H60 Heated Humidifier User Manual V2.0 Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Page 68
H60 Heated Humidifier User Manual V2.0 Recommended separation distances between wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment.
BMC Medical Co., Ltd. will repair or replace, at its option, the defective material or part. BMC Medical Co., Ltd. will pay customary freight charges from BMC Medical Co., Ltd. to the dealer location only.