Advertisement

Quick Links

Oxygen Concentrator User Manual
FORAS
OXYGEN CONCENTRATOR
USER MANUAL
MODEL : OXY300 – OXY500
CAUTION: Please read this manual before using.
1 / 18

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OXY500 and is the answer not in the manual?

Questions and answers

Summary of Contents for foras OXY500

  • Page 1: Cover

    Oxygen Concentrator User Manual FORAS OXYGEN CONCENTRATOR USER MANUAL MODEL : OXY300 – OXY500 CAUTION: Please read this manual before using. 1 / 18...
  • Page 2: Table Of Contents

    Oxygen Concentrator User Manual Table of Concents Page No Cover Index Introduction, Cautions and General Precautions Indications, Contraindications and Adverse Effects Patient Population and Accessories of Oxygen Concentrator Back View, Screen View and To-Do Before Operating Device Humidifier Bottle Usage Nasal Cannula Usage, Oxygen Concentrator Turn On/Off Maintenance and Controls Calibration and Cleaning...
  • Page 3: Introduction, Cautions And General Precautions

    In the event of a power cut, failure of the Oxygen Concentrator or insufficient oxygen supply to the patient, it is obligatory to carry a spare oxygen provider or oxygen cylinder with you. This is your responsibility. Foras Medical does not accept any responsibility for customers who do not comply with the manufacturer's instructions.
  • Page 4: Indications, Contraindications And Adverse Effects

    Oxygen Concentrator User Manual • If you feel sick or uncomfortable while using the device, consult your physician. • The doctor or caregiver should take protective measures against the risk of cross- infection if the patient has an infection. At the same time, no one at the risk of infection should have contact with the patient.
  • Page 5: Patient Population And Accessories Of Oxygen Concentrator

    Oxygen Concentrator User Manual  Smoking in the environment where the device is located and to keep flammable materials near the device may be dangerous.  If the device is used above the moisture specified in the instructions, oxygen treatment will not be possible because the columns will be damaged. ...
  • Page 6: Back View, Screen View And To-Do Before Operating Device

    Oxygen Concentrator User Manual (Figure 2) Display View (Figure 3) Note: When there is electricty, the icon on the power indicator is green colored. It is red colored when O2 Low and Call Service icons appeared. BEFORE OPERATE THE OXYGEN CONCENTRATOR After removing the device from its box, the instructions for use must be read before operating device.
  • Page 7: Humidifier Bottle Usage

    Oxygen Concentrator User Manual Note: Do not attach any other device to the outlet where the oxygen concentrator is connected. 2. Before operating the device always make sure that the air filter is installed and clean. 3. Attach the recommended Oxygen Concentrator accessories (Humidifier Bottle and Cannula) to the device.
  • Page 8: Nasal Cannula Usage, Oxygen Concentrator Turn On/Off

    Oxygen Concentrator User Manual Nasal Cannula Usage It is a tubing system made of from plastic-derived biocompatible material that placed in the nostrils of the patient who use the device and it provides oxygen delivery to the patient.This tubing system, does not prevent to patient's deating/drinking activities. Patient can continue his/her daily life with nasal cannula.
  • Page 9: Maintenance And Controls

    Oxygen Concentrator User Manual Flowmeter After operating the device, desired setting should be made by adjustment knob shown in Figure 10. When the button is turned to the "+" side, the flow increases. When the button is turned to the "-" side the flow decreases.
  • Page 10: Calibration And Cleaning

    Oxygen Concentrator User Manual Oxygen Humidifier Bottle Checkings If your doctor recommends using the device with humidifier; • Check the humidifier bottle periodically, as sediment and lime will form in the water tank during extended use. • It is recommended to change the humidifier bottle once a month. •...
  • Page 11: Alarms, Technical Specifications

    %82.. Call Service : The device gives an audible and visual alarm in the event of a power failure and when the flowmeter is 0 (zero). TECHNICAL SPECIFICATIONS OXY300 OXY500 Oxygen Level 1-3 LPM Oxygen Level...
  • Page 12: Environmental Conditions And Troubleshooting Tables

    Oxygen Concentrator User Manual ENVIRONMENTAL CONDITIONS • Do not use oxygen concentrator devices in any toxic environment. • Do not use oxygen concentrator devices in environmental conditions where explosive and chemical gases or other flammable anaesthetic agents are present. Keep fire away. •...
  • Page 13: Symbols Explanation

    Oxygen Concentrator User Manual SYMBOLS EXPLANATION Symbols Explanation “WARNING”, indicates the situations that safety of patient may be in WARNING danger. “CAUTION”, indicates that precautions should be taken and maintenance CAUTION instructions should be followed Check the user manual for instructions Do not smoke while using device Keep fire away Chek the user manual for instructions...
  • Page 14: Residual Risks

    Oxygen Concentrator User Manual RESIDUAL RISKS Potential Risk Residual Risk Application of Line Voltage Usage without reading user manual Voltage Leakage To User Failure to perform leakage tests and final checks Electric Shock Failure to perform leakage tests and final checks Magnetic Field/Radiation EMC test values are not declared in the manual Ionized Radiation Energy...
  • Page 15: Related Directives, Electromagnetic Compatibility

    Medical Devices Directive 93/42 / EEC is taken as reference. ELECTROMAGNETIC COMPATIBILITY Foras OXY Oxygen Concentrator complies with the electromagnetic compatibility limitations of the medical device directive 93/42/EEC. (En 55011 BF class and EN 60601- 1-2) IEC 60601-1-2 requirements for EMC have been completed.
  • Page 16 Oxygen Concentrator User Manual Description of Manual and Manufacturer Electromagnetic Immunity: The oxygen concentrator is suitable for use in the electro-magnetic environment which described below. The user should pay attention to the environment when using the concentrator and use it as in the declared environment. Immunity Test IEC 60601 Test Level Compatibility Level Electromagnetic Env.-Guide...
  • Page 17: Compliance With Standarts, 60601 Classification And Software Information

    The difference between them is software and the OXY300 model has the capacity to give 3 liters of oxygen per minute, while the OXY500 model has the ability to give 5 liters of oxygen per minute.
  • Page 18: Manufacturer Information, Notified Body Information

    Oxygen Concentrator User Manual Foras Bilgisayar Elektronik Medikal San. Tic. Ltd. Şti Ostim O.S.B. Mahallesi 1269.Cadde No: 35 Yenimahalle – ANKARA MADE IN TURKEY / TÜRK MALI www.foras.com.tr / info@foras.com.tr Tel: +90 312 395 77 66 / Fax: +90 312 395 77 67 Notified Body : Kiwa Belgelendirme Hizmetleri A.Ş.