ASCENT OPTI BASIC/ ASCENT OPTI DUCTLESS FUME CABINET USER & SERVICE MANUAL Esco SPB/SPD User & Service Manual Version 4 / Released 1 February 2008 Specifications subject to change without notice...
Page 3
It is a pleasure to welcome you to the exclusive club of our growing number of customers who, over the years have come to not just own and operate their Esco fume cabinets but treasure them too. For your convenience all exposed exterior and interior surfaces of this product have been wiped in a controlled clean environment.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: TABLE OF CONTENTS Table of Contents Safety Warning and Limitation of Liability Declaration of Conformity Freight Claim Information Warranty Terms and Conditions Corporate Profile Fume Filtration Division A.
Page 5
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3.9 Performance Validation/Certification 4. Operating Your Cabinet 4.1 Sentinel® Control System 4.1.1 Software Self-Reset 4.1.2 Hardware Reset 4.2 Menu Options 4.2.1 Settings (Clock, Timer, and Units) 4.2.2 Calibration 4.2.3 Administrator Access and PIN 4.2.4 Setting the Mode...
Page 6
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3. Troubleshooting 3.1 Electrical and Mechanical Troubleshooting 3.1.1 SPD-3/4AX 3.1.2 Software Troubleshooting 3.1.3 SPB-2AX Copyright Information APPENDIX Appendix A Filter Saturation Test Appendix B Measurement of Face Velocity Test Appendix C Estimated Carbon Capacities for Contaminant...
The disposal and / or emission of substances used in connection with this cabinet may be governed by various local regulations. Familiarization and compliance with any such regulation are the sole responsibility of the users of the cabinet. Esco’s liability is limited with respect to user compliance with such regulations.
ANSI / ASHRAE 110-1995 AFNOR NF X 15-211 AS/NZ 2243.9: 2003 More information may be obtained from Esco’s authorized distributors located within the EU. A list of these parties and their necessary contact information is available on request from Esco. Lim Lay Yew...
• Notify Esco Micro Pte Ltd regarding the items/parts that need to be replaced • For claim purposes, Esco will fax to you an order acknowledgment that would include the prices of the • damaged items (less freight cost) Esco will invoice you for replacement items/parts •...
E-mail: ductless@escoglobal.com Website: WARRANTY TERMS AND CONDITIONS Esco warrants that this equipment will perform according to the specifications for 3 years from the date of purchase. With certain exceptions*, this 3 years warranty applies to all parts and components of this...
We were among the both containment and filtration researchers conducting clinical earliest to spot this opportunity. efficacy such as the American biological research in developing Esco was thus founded in 1978 Standard ASHRAE 110, vaccines dangerous to pioneer development in this European Standard EN 14175, diseases –...
Page 12
ISO 9001 Quality Management World-Class R & D System At Esco, we understand that being at the cutting edge of technology is the key to our survival. Hence continuous Esco regularly participates in...
They are suitable for most common chemical work (Note that ductless fume cabinets are NOT suitable for all chemical application. Please consult Esco for advice concerning your individual application). The Esco Ascent Opti Basic/Ascent Opti Ductless Fume Cabinets are our new-improved products which offer the best combination of performance, quality and cost-effective in the market today.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Activated carbon filters are suitable for use with many chemicals but this is dependent on a number of factors. In general, chemicals with a relative molecular weight over 30 and a boiling point higher than C or 140 F can be adsorbed with a relatively high efficiency by activated carbon filtration.
E-mail: ductless@escoglobal.com Website: 1.3 SAFETY At Esco, your safety is of paramount importance to us. Thus Esco Ascent Opti Basic/Ascent Opti Ductless Fume cabinet has been packed with features that would take your safety to an altogether new level. A highlight of these features is as follows: A unique filter mounting system that eliminates bypass leaks inherent in conventional clamping •...
Product Development Laboratory. Esco performs testing in accordance with more than 20 of the world’s most recognized standards of local, regional and international scopes. In particular, testing in our laboratory is most frequently conducted based on: ANSI/AIHA Z9.7: 1998.
FiltraCheck is a trademark service provided by Esco’s fume filtration division. Customers who intend to purchase a ductless fume hood but are unsure whether the cabinet is suitable for their application, can forward a list of chemicals that they will be handling and their pattern of usage to Esco's FiltraCheck service team.
Page 19
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Esco’s FiltraCheck Diagram USER SECTION CHAPTER 1 – BASIC PRODUCT INFORMATION...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: CHAPTER 2 UNPACKING YOUR CABINET This chapter aims to provide relevant information on how to handle the cabinet properly upon receipt. Failure to follow the following instructions may cause damage to the cabinet. We strongly advise you to read this chapter carefully before proceeding further.
Page 21
The work surface should be smooth, nonporous and resistant to those disinfectants and chemicals, to which the cabinet is regularly exposed. Esco is committed to protecting the environment. Please re-use the pallet as far as possible and/or dispose off the packaging material and box in an environmentally responsible manner. Please hammer any exposed nail/sharp projection on the box’s surface prior to disposal.
2) Electrical diagram In case this manual and/or test report are lost or misplaced, you have little to worry about. Esco retains a copy of the same. At any time, you may obtain a replacement copy by contacting Esco and stating the cabinet model, serial number and a brief description of the information you are seeking.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: CHAPTER 3 INSTALLING YOUR CABINET 3.1 Installing SPB-2AX 3.1.1 List of Parts The following parts are included in your packaging: Parts Quantity Plastic Screws 12 pieces Hexagonal Bolt M8 x 20 3 pieces...
Page 24
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 2. Assemble the main top panel joined to back panel and side wall using: - M8 – 5 plastic screws to side wall left and right. - M8 –...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4. Finally lock the top module to the bottom module by using 2 countersink screw M4 3.2 Installing SPD-3/4AX 3.2.1 List of Parts The following parts are included in your packaging: Parts Quantity...
Page 26
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3.2.2 Assembly Guide 1. Join the Baffle Board onto the Back Panel using 4 screws (M8-type) 2. Join both left and right Side Wall (Acrylic) using M8 – 6 White colored plastic screws for left and right side USER SECTION CHAPTER 3 –...
Page 27
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3. Attach the stainless work top table and the drain pan using M8 – 4 white colored plastic screws at left and right side 4.
Page 28
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 5. Assemble the front glass / acrylic using M4 – 2 metal screws. 6. Finally lock the top module to the bottom module by using 2 countersink screw M4 USER SECTION CHAPTER 3 –...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3.3 INSTALLATION LOCATION REQUIREMENT Installation location of the cabinet has a big impact on its overall performance. Installing the cabinet at a location where it is likely to be exposed to a lot of external airflow disturbances might lead to loss of containment.
E-mail: ductless@escoglobal.com Website: 3.4 INSTALLATION LOCATION EXTRACTED FROM BS 7258 The following article is Esco recommendation extracted from British Standard BS 7258- PART 2:1994: To avoid disturbances to the fume cabinet and its operator, consideration should be given to the...
Page 31
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Key to figures 1 and 2 USER SECTION CHAPTER 3 – INSTALLING YOUR CABINET...
Page 32
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: USER SECTION CHAPTER 3 – INSTALLING YOUR CABINET...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3.5 CASE STUDY: CHOOSING THE BEST INSTALLATION LOCATION Examples of the influence on cabinet airflow from room design and ventilation 1. Cabinet 1 is quite appropriately located with respect to avoidance of excessive air movements from surrounding areas 2.
3.6 OPTIONAL RETROFIT KITS 1. Cabinet is available with a transparent glass back-wall, ideal for classrooms and educational demonstrations 2. Optional exhaust collar for exhaust ducting Ascent Opti Basic/Ascent Opti Ductless Cabinets 3. Further customisation is available upon request 3.7 INSTALLATIONS 1.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 3.8 ENVIRONMENTAL/ELECTRICAL CONDITIONS This cabinet has been designed to be used under the following conditions: ENVIRONMENTAL 1. Indoor use 2. Altitudes of up to 2000 meters (6600 ft) 3.
Step 3(b). 3.9.3 DISCLAIMER The perfomance and safety of all Esco cabinets are rigorously evaluated at our factory. However we cannot guarantee similar levels of performance and safety after cabinet has been shipped and installed at the user’s end because of possible wear and tear during transit. Hence we strongly recommend frequent on-site testing.
This chapter is aimed at providing the users with information on basic operations of this cabinet, besides outlining various recommended work practices, so that they are able to achieve optimal containment from their Esco fume cabinets. 4.1 SENTINEL® CONTROL SYSTEM Note: This is only applicable for SPD 1.
Page 38
4.2.3.2 for further instructions. The Admin PIN has higher priority and can be used to control the fan (override the Fan PIN) too. Please contact Esco should you forget your ADMIN PIN. USER SECTION CHAPTER 4 – OPERATING YOUR CABINET...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4.1.1 SOFTWARE SELF-RESET Sentinel® is equipped with watchdog timer. The timer runs in the background and detects any possible incidence of software malfunction. Once a malfunction occurs, the timer will reset the system. During system reset (which takes approximately 30 seconds), the cabinet will undergo warm-up cycle;...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4.2 MENU OPTIONS Note: This is only applicable for SPD When you are entering menu options, the alarm will sound to indicate that the microprocessor is not monitoring the operation of the cabinet and as such will not give airflow alarms.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4.2.1 SETTINGS (CLOCK, TIMER, AND UNITS) Press SET to enter the SETTINGS menu. Use the UP and DOWN button to move through the available options.
Page 42
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Press SET button to enter the POST PURGE menu. The minute display will blink. Press UP or DOWN arrow button to choose the correct minute value. Press SET button to confirm the changes.
Press MENU to go back to the previous level. 4.2.2.1 SET CONSTANT Every sensor manufactured by Esco has a specific Sensor Constant which is used for temperature compensation performed by the temperature sensor. For more information on this option, please refer to Microprocessor Calibration section of the Test Report.
Page 44
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Display sequence to set admin PIN 4.2.3.2 NEW FAN PIN FAN PIN restricts access to the fan controls. Once the FAN PIN has been set, one will have to correctly key in the four-digit PIN before being allowed to switch the fan on or off.
Page 45
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Display sequence to set alarm setting feature Note: When the alarm is disabled, the warm up time will be automatically set to zero and cannot be changed.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4.2.4 SETTING THE MODE Press SET to enter the SET MODE menu. Use the UP and DOWN button to move through the available options.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4.3 STARTING THE CABINET 1. Prepare a written checklist of materials/apparatus necessary for immediate usage. 2. (Note: This point is only applicable for SPD)Turn on the blower (default FAN PIN is 0000). Leave the blower on for 3 minutes (default warm-up time is 3 minutes;...
7. Exercise regularly Ergonomics accessories available with Esco include: a. Armrest padding b. Lab chair c. Footrest Please contact your local distributor or Esco for more information. USER SECTION CHAPTER 4 – OPERATING YOUR CABINET...
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 4.5 CLEANING AND SHUTTING DOWN THE CABINET 1. As far as possible, it is recommended that the cabinet be operated continuously in order to achieve optimal containment and cleanliness.
In view of this, proper and timely maintenance is crucial for trouble free functioning of any device and your Esco cabinet is no exception to this rule. We strongly recommend that you follow the maintenance schedule suggested hereunder in order to obtain optimal performance from your Esco cabinet.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 5.2 SELF CHECK TEST This self check test is meant to determine the saturation point of Activated Carbon. This procedures use Detection Tubes (Color Indicator Tubes) Color indicator tubes can be used as a quick pass/ fail method to check for air purity, in other words, also for determining when it is time to change Activated Carbon filters whereby the concentration of chemical used after adsorption is above its threshold values.
Page 52
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Operator doing a Self Check Test Advantages: 1. Once you own the hardware, you can perform the tests on compressor air as well as air in storage banks 2.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: CHAPTER 1 RE-CERTIFICATION AND MAINTENANCE BY SERVICE PERSONNEL 1.1 AIRFLOW VERIFICATION & CABINET RE-CERTIFICATION Re-certification must be performed during any of the following eventualities: 1.
Page 55
Nominal Motor Voltage in the Test Report. Your building supply voltage may differ from the supply voltage at the Esco Test Lab. This supply voltage difference will roughly double the blower voltage difference, creating a more significant airflow discrepancy.
6. You are recommended to perform the following optional tests too: a. Operator comfort tests: light, noise, and vibration Esco conducts training courses on the validation and certification of fume cabinets. If you are interested you are most welcome to contact Esco or your nearest Esco distributor.
1.3 SOFTWARE UPDATING Esco sometimes releases upgraded versions of the Sentinel® control’s embedded software. The upgraded version can be downloaded from our website using a toolkit that can be purchased from Esco at a nominal cost. You do not have to compulsorily upgrade your Sentinel® control’s embedded software unless it is for safety considerations.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: CHAPTER 2 PRODUCT SPECIFICATION CONTENTS: Engineering Diagram of All Models of Different Sizes Technical Specification Summary Table Electrical Schematics of Each Power Rating Variant Microprocessor Specification All features and specifications are subject to change without any prior notice.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: 2.9 SENTINEL CONTROL SYSTEM Note: This applies only for SPD Mains Input Voltage: 100-240VAC Mains Input Frequency: 50/60Hz Sentinel® Control Input Voltage: 7.5VDC Max input current: 500mA Electromagnetic interference Hardware requirement:...
CHAPTER 3 TROUBLESHOOTING This section helps you troubleshoot some of the common problems you might face while operating this cabinet. Should you have any queries left unanswered here, please feel free to contact Esco. 3.1 ELECTRICAL AND MECHANICAL TROUBLESHOOTING 3.1.1 SPD-3/4AX...
Page 69
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Possible Problem Corrective Action causes Restart the cabinet several times. If the following conditions persist, you should know that the main board is spoiled: 1.
Page 70
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Possible Problem Corrective Action causes Switch off the cabinet. Short speed control connection by connecting a jumper wire across the 2 cables going to speed control. Speed control is located on electrical Faulty speed panel, type KBWC, covered by stainless steel bracket.
Page 71
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Possible Problem Corrective Action causes Warning: Before checking connection, be sure disconnect main plug. 1. Check for any loose or bad connection between relay board, fluorescent lamp ballast and female connector B on the electrical box.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Possible Problem Corrective Action causes Plug in the multimeter to the Motor Voltage Sampling Port. Turn the Speed Control to increase the blower voltage by approximately 10 VAC until the alarm is turned off and the LCD shows nominal airflow reading.
Page 73
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Possible Problem Corrective Action causes In electrical box, measure AC voltage between LIVE and NEUTRAL terminal block. The voltage should be around either 230Vac or 115Vac +/- 10%. If the voltage is much lower or near zero, check cable connection at connector A.
Page 74
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: Possible Problem Corrective Action causes Ensure that there are no external sources of airflow disturbance like air External air conditioner vent, window or incidences of door opening or people walking fast interference near the cabinet.
The information contained in this manual and that in the accompanying product are copyrighted and all rights are reserved by Esco. Esco reserves the right to make minor design changes periodically without any obligation to notify any person or entity of such changes.
Test for saturation by organic gases (limited usage) • For our purpose, Esco will be using the propan-2-ol challenge and burning sulphur test method. SECTION B: PROPAN-2-OL CHALLENGE TEST The filter, when challenged continuously at 800 ppm +/- 50 ppm of propan-2-ol, shall maintain the concentration of propan-2-ol in the exhaust stream at below 40 ppm throughout the period of evaporation of 1 liter of propan-2-ol and below 400 ppm throughout the period of evaporation of 2 liter of propan-2-ol.
Page 78
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: B.2 PROCEDURE 1. Determine the air volume flow rate through the fume cupboard and record the value as F /min. 2. Pour about 100 ml of propan-2-ol into a preweighted, stoppered, conical flask containing a few anti-bump granules.
Page 79
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: SECTION C: METHODS USING BURNING SULPHUR The rate of release of sulphur dioxide is calculated by weighing the vessel containing the sulphur dioxide before and after it has burning for a measured time.
Page 80
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: C.1 CALCULATION Rate of release of sulphur dioxide = (M ) x 750 / t Now check your result with the table below. Rate of sulphur dioxide relesed (cm Maximum permitted concentration of sulphur dioxide in the exhaust gas (ppm)
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: APPENDIX B MEASUREMENT OF FACE VELOCITY TEST Note: Do note that our factory face velocity test provided in our test report is based upon the ASHRAE standard which uses more extensive equipments.
Page 82
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: SECTION B: CALCULATION Minimum face velocity Record which of a, b, c, d, e, f, g, h, i is the smallest Is the value above or below 0.3 ms ? If it is below, the fume cupboard fails.
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: APPENDIX C ESTIMATED CARBON CAPACITIES FOR CONTAMINANT This section is provided to present a list giving estimated activated carbon capacities for contaminant vapours. The adsorption capacity of activated carbon for the various vapours varies with: concentration, •...
Page 84
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: APPENDIX...
Page 85
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 www.escoglobal.com/ductless E-mail: ductless@escoglobal.com Website: APPENDIX...
BS 7258-4:1994 Laboratory fume cupboards. – Part 4: Method for determination of the containment value of a laboratory fume cupboard. WEBSITES 1. http://www.acgih.org/ American Conference of Governmental Industrial Hygienists (ACGIH) www.escoglobal.com Esco Micro Pte. Ltd. 4. http://www.cdc.gov/niosh/homepage.html National Institute for Occupational Safety and Health (NIOSH) http://www.osha.gov Occupational Safety and Health Administration, USA https://www.aiha.org/...
E-mail: ductless@escoglobal.com Website: DEFECT REPORTING FORM To submit your warranty claim or to obtain a Return Authorization (RA) number, please complete this defect reporting form and return it to Esco by fax or postal mail. PRODUCT INFORMATION Model Number : _____________________________________________________________________________...
Page 88
This page has been intentionally left blank...
Page 89
21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920 E-mail: ductless@escoglobal.com Visit our website at ductless.escoglobal.com...