High-Precision Advanced Tuning Fork Balance MG-S Series Manual Operation IMPORTANT • To ensure safe and proper use of the balance, please read this manual carefully. • After reading this manual, store it in a safe place near the balance, so you can review it as needed.
This document describes how to operate the product. Instructions ● The copyright of this document belongs to Star Micronics. Reprinting or duplicating of all or part of this document without notice shall not be allowed. ● Please note that product improvement or modification may cause partial discrepancy between the product and the description of this document.
● The copyright of this document is held and reserved by Star Micronics. Duplicating or disclosing its drawings and engineering materials without prior approval of Star Micronics in writing is not permitted.
How to use this document ■Symbols used in this document Understand the meaning of the following symbols and observe the instructions of this document. Symbols Meaning Used for the situation that invites an imminent risk of death or severe injury if proper precautions are not taken.
Page 6
■About how to read this document This document consists of the following contents: Prior to use Describes about operating precautions, names and functions of each section, etc. Please be sure to read this section when using this product for the first time. Basic usage Describes about basic usage related to weighing such as how to turn on and off the power in addition to the setting procedures to set various...
Contents Preface ............................i Important Notice ........................ii How to use this document ...................... iii Contents ............................ v 1 Prior to use .......................... 1 1-1 Operating precautions ....................1 1-2 For more accurate measurement ................... 3 1-2-1 Precautions related to measuring environment ............3 1-2-2 Precautions related to measuring table ...............
Page 8
3-6-1 Switching the display at “Statistics mode” ..............35 3-7 Animal mode ........................ 36 3-8 Formulation mode ......................38 3-8-1 Check the stored data of each component..............39 3-9 Unit setting ........................40 3-10 Comparator function ....................40 3-10-1 How to perform discrimination ................. 40 3-10-2 Comparator function setting ..................
MG-S series operation manual 1 Prior to use 1 Prior to use 1-1 Operating precautions ■Do not wet the AC adapter. That may cause an electric shock, short-circuiting or failure. ■Do not handle the balance with wet hands. That may cause short-circuiting or failure.
Page 12
1 Prior to use MG-S series operation manual ■Do not mix old and new batteries, or batteries of different types or manufacturers. ■Do not use the batteries that leak. ■Do not apply excessive force to or impact the balance. Doing so could damage or result in failure of the balance. Carefully place samples on the balance.
MG-S series operation manual 1 Prior to use 1-2 For more accurate measurement To make more accurate measurement, it is necessary to lessen error-causing factors in measurement to the extent possible. Error-causing factors include not only an instrument error and performance of the balance itself but also the nature and condition of a specimen, measuring environment (vibration, temperature, humidity, etc.) and the like.
1 Prior to use MG-S series operation manual 1-2-3 Precautions related to a specimen → In general, synthetic resin- and glass-made specimens are high in electric Static electricity insulation, and so easily charged electrically. Weighing an electrically charged specimen makes the indication value unstable, reducing the reproducibility of the test result.
MG-S series operation manual 1 Prior to use 1-3 Check for the articles contained in the box The package box contains the following; If anything missing or broken should be found, please inform the store where you purchased the product.
1 Prior to use MG-S series operation manual 1-4 Name and function of each section Round pan type (MG-S322) Square pan type (MG-S1501, MG-S8200) Windshield Weighing pan Level Adjuster Display Battery case AC adapter jack USB connector (Type B) RS-232C connector (D-sub 9 pin male) Bluetooth module -6-...
MG-S series operation manual 1 Prior to use 1-5 Assembling and installation of the product 1-5-1 Assembling the balance (Round pan type MG-S322) Attach the “Pan base”. "Slider" to check that in the "Unlock” side, then attach to the balance.
1 Prior to use MG-S series operation manual 1-5-2 Assembling the balance (Square pan type MG-S1501, MG-S8200) Attach the “Pan base”. (1) Direct "FRONT" to the display side. (2) Attach to the balance, then turn the “Pan base screw” to fix.
MG-S series operation manual 1 Prior to use 1-5-3 Level At the time of shipment, the adjusters provided Release the transportation lock of the at the four corners of the bottom are locked. adjuster. Turn them in the direction shown in the figure on the left to loosen them.
1 Prior to use MG-S series operation manual 1-6 Description of the operation keys 1-6-1 Basic These keys are called [F] keys or [F1]-[F4] key in this manual as a matter of convenience, while there are not such indications around them.
MG-S series operation manual 1 Prior to use 1-6-2 Setting value and numeric value inputting Name of Performance [Menu] Cancel the input value and go back to the setting menu. [Tare] Input a decimal point < . > in “Specific Gravity mode”.
1 Prior to use MG-S series operation manual 1-7 How to interpret the display 1-7-1 Description of segment. Mark Name Description Minus Indicates the negative weight value and numeric. When displayed: The balance is in the stable condition. Stable mark When not displayed: The balance is not in the stable condition.
MG-S series operation manual 1 Prior to use 1-7-2 LCD character font ■7-segment comma point space minus / hyphen ■16-segment asterisk slash left arrow right arrow space plus minus / hyphen comma point percent Degree Celsius -13-...
Basic usage MG-S series operation manual 2 Basic usage 2-1 Turning on/off the power, and checking for the operation Turn on the power for the balance. Connect the included AC adapter to the balance. When the AC adapter is plugged in, the balance enters the standby state and an asterisk <...
MG-S series operation manual Basic usage 2-2 Zero-point adjustment Adjusting the indication to zero is called "Zero-point adjustment". Check the weighing pan. Make sure that nothing is placed on the weighing pan. Execute “Zero-point adjustment”. Push [Zero] key. After the weight indication is stabilized, the indication become zero and the symbol “...
Page 26
Basic usage MG-S series operation manual Check the tare weight. The tare weight can be checked by operating “Free keys” if the <TARE> is assigned to the Free key. Refer to “8 Controlling and adjustment functions” for setting the Free keys.
MG-S series operation manual Basic usage 2-4 Weighing the additional sample Weigh the first sample and the additional sample separately. Place a sample to be weighed. The mass of the sample to be weighed placed is indicated. Perform the tare.
Basic usage MG-S series operation manual 2-5-2 Operation of the setting menu To perform settings for various functions from the state of weighing, chiefly execute the following procedure. ■Go to the menu item to set ■Select the setting value and execute/fix.
MG-S series operation manual Basic usage 2-5-3 Numeric value input Input upper/lower limit, reference weight, preset tare weight, coefficient, specific gravity of the media liquid, water temperature, date/time and ID/password at each mode. Numeric value inputting is limited to eight digits at a maximum.
Basic usage MG-S series operation manual 2-5-4 [ F ] key switching at each measuring mode You can switch the measuring mode, or select and set the function, by operating the [F] keys at each measuring mode. This chapter shows the [F] keys switching by pushing the [F4] key.
MG-S series operation manual 3 Functions related to the operation 3 Functions related to the operation Settings to change the balance operations. 3-1 Hierarchy of functions related to the operation For verified balance: - Gray shaded items ( ) are not indicated;...
3 Functions related to the operation MG-S series operation manual 3-2 Various measuring modes of the balance Refer to “6 External input/output functions” to output the measuring data to other devices. 3-2-1 Weighing mode Weighing mode is the basic mode for weighing.
MG-S series operation manual 3 Functions related to the operation 3-2-2 Counting mode Counting mode can count the number of items by placing the items for which sampling has been completed on the balance and dividing the total weight of those items by the recorded unit weight.
Page 34
3 Functions related to the operation MG-S series operation manual Place the set number of samples on the Place the samples. container. Record the unit weight. Push [F4] key to fix. The unit weight is recorded. When <1C Simple SCS> is valid, Simple SCS Simple SCS method (When enabled).
MG-S series operation manual 3 Functions related to the operation 3-2-2 (2) Numeric value setting method Input numeric value of the unit weight by key operation. This moethod is not available for verified balance. Push [F3/F4] key to select whether or Select whether or not employ the previous not employ the previous data.
3 Functions related to the operation MG-S series operation manual 3-2-2 (3) Switching the display at Counting mode Push [F1-F4] keys to switch the display. For non <ADD> and <TOTL> can be used when <14 ADDITION> is activated. -26-...
MG-S series operation manual 3 Functions related to the operation 3-3 Percentage mode The weight of a sample to be weighed is indicated in percent relative to the reference weight. There are two methods to input the reference weight; - Actual value setting method ([onW]): Place the reference weight on the balance to record the weight.
Page 38
3 Functions related to the operation MG-S series operation manual Save the reference value. Place the reference weight on the balance. In the case of [onW]. Push [F4] key to record. In the case of [NUM]. Input the reference value.
MG-S series operation manual 3 Functions related to the operation 3-3-1 Switching the display at percentage mode Push [F1-F4] keys to switch the display. For non <ADD> and <TOTL> can be used when <14 ADDITION> is activated. -29-...
3 Functions related to the operation MG-S series operation manual 3-4 Multiplied by Coefficient mode Measured weight is multiplied by the preset coefficient, and the result be displayed. This mode is not available for verified balance. Select the Multiplied by Coefficient mode.
MG-S series operation manual 3 Functions related to the operation 3-4-1 Switching the display at Multiplied by Coefficient Push [F1-F4] keys to switch the display. <ADD> and <TOTL> can be used when the <14 ADDITION> is activated. -31-...
3 Functions related to the operation MG-S series operation manual 3-5 Specific gravity mode In the specific gravity mode, the ratio of the density of a substance to the density of water at its densest (4°C) for liquids is calculated.
MG-S series operation manual 3 Functions related to the operation Measure the sample weight in the air. Set the net/basket on the balance and push [Tare] key. Load the on the net/basket to measure the weight of the sample in the air, then push [F4] key to record it.
3 Functions related to the operation MG-S series operation manual 3-6 Statistics mode The statistical operation function collects weight data and indicates maximum, average, and other statistical values. This mode is not available for verified balance. (1) Only “mg” or “g” can be used.
MG-S series operation manual 3 Functions related to the operation Collect more weighing data. Store data in the same way as in step 4. Repeat placing samples, storing data, and removing the samples until the required number of data items are collected.
3 Functions related to the operation MG-S series operation manual 3-7 Animal mode The balance can accurately weigh animals and other samples that move during measurement. Even when animals and other samples move during measurement, when weight variations fit within the set value range, the indication is held (hold) and the measurement result can be read.
Page 47
MG-S series operation manual 3 Functions related to the operation a) Weigh the animal. Place the animal on the weighing pan. After the weight variations fit within the set range, the weighing indication is held and < > indication appears.
3 Functions related to the operation MG-S series operation manual 3-8 Formulation mode “Formulation mode” can store and refer the weight of each component compounded. This mode is not available for verified balance. Only “mg” or “g” can be used.
MG-S series operation manual 3 Functions related to the operation Store the sample weight. Put the sample on the tare and push [Output] key to store the sample weight. Repeat the step 6 for all the samples to be compounded.
3 Functions related to the operation MG-S series operation manual 3-9 Unit setting Various units can be selected. Please also refer to "Appendix 1-1 Basic specifications" and "Appendix 3 Unit indication and conversion table". For verified balance, selectable measuring units are limited to: - MG-S322: “gram”, “carat”, “pound”, “ounce”...
MG-S series operation manual 3 Functions related to the operation 3-10-2 Comparator function setting For the setting of the reference value and upper and lower limit values, refer to “5 User information setting”. Select the comparator function. Push [Menu] key, then push [F1-F4] keys to go to <131 ACTIVATE>...
3 Functions related to the operation MG-S series operation manual 3-11 Adding function Weigh a plurality of samples to be weighed in sequence and indicates its total value. The adding function includes two ways of calculating method. - Method of weighing samples to be weighed while replacing the samples: Addition accumulating function.
MG-S series operation manual 3 Functions related to the operation 3-11-1 Weighing by means of the plus side addition When <<ADD>> is assigned to [F1] key and <<TOTL>> is assigned to [F2] key. Place a first sample to be weighed.
3 Functions related to the operation MG-S series operation manual 3-11-2 Weighing by means of the minus side addition When <ADD> is assigned to [F1] key and <TOTL> is assigned to [F2] key. Place a sample to be weighed. Place a sample to be weighed.
MG-S series operation manual 3 Functions related to the operation 3-12 Tare-subtraction reminder function When the “tare-subtraction reminder” is activated, <PUSH TARE> alert is displayed when the tare (container) is loaded. When the zero-point-adjustment reminder operates at the same time, the zero-point adjustment reminder has priority.
3 Functions related to the operation MG-S series operation manual 3-13 Zero-point-adjustment reminder function When the “zero-point-adjustment reminder” is activated, <PUSH ZERO> alert is displayed when the load return to within the “zero-point adjustment range” after the load is once over the range.
MG-S series operation manual 3 Functions related to the operation 3-15 Stabilization wait setting Set when to indicate the weighed value after the zero-point adjustment or tare; either after or before the weighed value stabilizes. For verified balance: This setting menu is not available;...
3 Functions related to the operation MG-S series operation manual 3-17 Auto power-off This function is to automatically turn off the power for the balance. Select the auto power-off. Push [Menu] key, then push [F1-F4] keys to go to <1B AUTO OFF>.
MG-S series operation manual 4 Functions related to the performance 4 Functions related to the performance Set the balance indication stability and response speed. 4-1 Hierarchy of functions related to the performance For verified balance, gray shaded items ( ) are not indicated.
4 Functions related to the performance MG-S series operation manual 4-3 Response speed The larger numeric value is set in this setting menu, the more stable the balance indication becomes in unstable conditions. Select the response speed. Push [Menu] key, then push [F1-F4] keys to go to <22 RESPONSE>.
MG-S series operation manual 5 User information setting 5 User information setting Describes about setting items related to the comparator function. 5-1 Hierarchy of user information setting For verified balance, gray shaded items ( ) are not indicated. ★: Initial setting value in a verified balance...
5 User information setting MG-S series operation manual 5-2 Preset tare When using a tare whose tare weight is already known, the tare subtraction can be performed in advance by inputting its tare weight (preset tare weight). Five preset tare weight values can be registered.
MG-S series operation manual 5 User information setting 5-2-2 (1) Actual value setting method Set a preset tare weight value. Push [F3] key to select. onW : Actual value < > is indicated. Place a sample to be weighed that is equivalent to the tare weight value.
5 User information setting MG-S series operation manual 5-3 Setting of the discrimination value of the comparator function There are two ways of inputting a reference value and upper and lower limit values as described below: - Actual value setting method: Weighing a sample with a balance and then making it a setting value.
MG-S series operation manual 5 User information setting 5-3-1 Actual value setting method Set a lower limit value. Push [F3] key to select. onW : Actual value Place a sample to be weighed that is equivalent to the lower limit value.
External input/output functions MG-S series operation manual 6 External input/output functions This function is used for communication through the external peripheral devices. There are RS-232C (D-SUB 9P), USB (Type B) and Bluetooth v4.0 interface as standard equipment. 6-1 Hierarchy of the external input / output functions Gray shaded items ( ) are not indicated for verified balance.
Page 67
MG-S series operation manual External input/output functions -57-...
Page 68
External input/output functions MG-S series operation manual -58-...
MG-S series operation manual External input/output functions 6-2 Standard RS-232C Connecter terminal numbers and their functions The RS-232C connector pin alignment for this product is as shown below: Terminal no Signal name Input/output Function - - - Input Receiving data...
External input/output functions MG-S series operation manual 6-4 Communication format 6-4-1 Basic communication specification Description Items Communication RS-232C: Full-duplex communication method method USB: Half-duplex communication method Bluetooth V4.0 Synchronization Asynchronous communication method method Electrical RS-232C: EIA-232-D/E specification USB: USB2.0 Bluetooth: Class 1...
Page 71
MG-S series operation manual External input/output functions Meaning of the data Symbol Code Description [P1] (one character) Indicates the polarity of data. 0x2B Zero or positive data 0x2D Negative data [D1 to D7/D8/D9] (seven or eight or nine characters) Stores numeric data.
External input/output functions MG-S series operation manual 6-4-3 Generic format Measuring data except Specific Gravity: Composed of 26 characters including a terminator (CR=0x0D/LF=0x0A) (SP) (SP): space (SP) ERROR: Composed of 26 characters including a terminator (CR=0x0D/LF=0x0A) (SP) (SP) (SP): space (SP) Others (Date, Time, Specific Gravity etc.):...
Page 73
MG-S series operation manual External input/output functions Symbol Code Description [U1, U2] (2 characters): Represents the unit of numeric value data. For non milligram 0x6D 0x67 gram (SP) 0x20 0x67 carat 0x63 0x74 momme 0x6D 0x6F ounce 0x6F 0x7A pound...
External input/output functions MG-S series operation manual 6-5 Input command 6-5-1 Transmission procedure Send an input command from an external device to the balance. The table below shows the enable/disable of input commands in each measuring mode. Commands Measuring Zero-point adjustment,...
MG-S series operation manual External input/output functions 6-5-2 Input command composition 1 Composed of four characters including a terminator (CR=0x0D/LF=0x0A). 6-5-2 (1) Zero-point adjustment/Tare/Output control setting command Please take care not to take alphabetical “O” for Arabic number “0”. Response...
External input/output functions MG-S series operation manual 6-5-3 Input command composition 2 Composed of 15 characters including a terminator (CR=0x0D/LF=0x0A) (1) ‘C3’ is maximum ten-digit (including the polarity +/-, comma and point) numeric data. Example) Upper limit input 120.00g: “LA,120.00”...
MG-S series operation manual External input/output functions 6-6 Response 6-6-1 Response command format (“A00”/“Exx” format) Consists of five characters including terminators. 6-6-1(1) Response command code(A1) code(A2) code(A3) Description 0x41 0x30 0x30 Normal response 0x45 0x30 0x31 Abnormal response 6-6-2 Response command format (“ACK”/“NAK” format) Consists of one character without a terminator.
External input/output functions MG-S series operation manual 6-8 Communication setting 6-8-1 RS232C/USB/Bluetooth For verified balance: Setting menus <412/422/432 FORMAT> are not available. They are fixed to <CBM> (Gereric format), and output formats <6>, <7>, <8>, <CSP6> and <CSP7> are not available;...
Page 79
MG-S series operation manual External input/output functions Select the communication setting. Refer to the step 1 to key operation for setting. Select the output format. Set list 6-digit numeric 7-digit numeric 8-digit numeric format format format CSP6 CSP 6-digit format...
External input/output functions MG-S series operation manual 6-8-2 Bluetooth initialization This function is used to initialize pairing with external Bluetooth device. Select Bluetooth initialization Push [Menu] key, then push [F1-F4] keys to go to <44 BLUETOOTH/BLE>, then push [F4] key to go to <441 INIT>.
MG-S series operation manual Functions related to the lock 7 Functions related to the lock Impose limitations on key operation and accessing the menu items, etc. 7-1 Hierarchy of functions related to the lock ★: Initial setting value 7-2 Total lock release All locks that have been set can be released.
Functions related to the lock MG-S series operation manual 7-3 Key lock function Key operation can be locked. Select the key lock function. Push [Menu] key, then push [F1-F4] keys to go to <52 KEY LOCK>. Push [F4] key to change the setting value.
MG-S series operation manual 8 Controlling and adjustment functions 8 Controlling and adjustment functions Perform setting of the Balance ID (Scale ID), the span calibration/test and the date and time. 8-1 Hierarchy of controlling and adjustment functions For verified balance, gray shaded items ( ) are not available.
Page 84
8 Controlling and adjustment functions MG-S series operation manual -74-...
MG-S series operation manual 8 Controlling and adjustment functions 8-2 Shortcut setting for accessing various measuring modes Shortcuts for various measuring mode can be assigned to <<F1-F3>> which are displayed above [F1-F3] key. For verified balance, <MULT>, <STAT>, <ANIM> and <FORM> are not available.
8 Controlling and adjustment functions MG-S series operation manual 8-3 Free key setting Free key setting is valid only in the weighing mode. For verified balance: - <mg>, <CAL>, <ADD>, <TOTL> and <HOLD> are not available; - <ct> is available only on MG-S322.
MG-S series operation manual 8 Controlling and adjustment functions 8-4 Maintenance settings 8-4-1 Span calibration and span test Span calibration is to “decrease” the difference between an indicated value and the true value (mass), and span test is to “check” the difference between an indicated value and the true value.
Page 88
8 Controlling and adjustment functions MG-S series operation manual The span calibration starts. Place the weight in the center of the weighing pan. Display changes to the order of <on FS> → “blinking of <on FS>”. Start of the span calibration.
MG-S series operation manual 8 Controlling and adjustment functions 8-4-1(2) Span test with external weight Make sure to use the external weight which is equal to the weighing capacity of each model. Select the span test with external weight. Push [Menu] key, then push [F1-F4] keys to go to <632 EX SPAN TEST>.
8 Controlling and adjustment functions MG-S series operation manual 8-5 Balance control setting 8-5-1 Balance ID setting A balance ID (Scale ID) can be set to discriminate the balance. The balance ID is output with GLP header output and external span calibration/test result output. Balance ID is checked by free key <<ID>>.
MG-S series operation manual 8 Controlling and adjustment functions 8-5-2 (1) Administrator password registration Make sure not to forget the administrator password. In case that the administrator password is lost, please contact the store where you purchased the product. Only one password can be set for administrator.
8 Controlling and adjustment functions MG-S series operation manual 8-5-3 Outputting of the span calibration/ test result After span calibration/test, the result can be output automatically. Make sure to activate <41 RS232C> and/or <42 USB> and/or <43 BLUETOOTH> to output the data.
MG-S series operation manual 8 Controlling and adjustment functions Exit the setting menu. Push [Menu] key to shift to the measuring mode. Push [F2] (<<DATE>>) key. Indication of the date. The date is indicated. Push [F4] key to return to the measuring mode.
8 Controlling and adjustment functions MG-S series operation manual 8-5-7 Direct start setting This is a function to turn on the balance automatically without pushing [On/Off] key when it is connected to the AC power or USB bus powered from PC. You can use this function when the balance is used in conjunction with other devices.
MG-S series operation manual Troubleshooting 9 Troubleshooting If the trouble persists after following the procedures below, please contact the store you purchased. 9-1 Error message Error Message/ Cause Coping method Error Code OVER The weight of the sample to be weighed is in...
Page 96
9 Troubleshooting MG-S series operation manual Error Message/ Cause Coping method Error Code ERR723 Out of Zero adjustment range (1.5% of the Make sure nothing on the weighing pan maximum capacity) while executing zero adjustment. ERR724 Out of Tare subtraction range (0g to the maximum...
Page 97
MG-S series operation manual Troubleshooting Error Message/ Cause Coping method Error Code ERR752 The unit weight of the samples is 0g and under at Choose the samples of which unit Counting mode. weight is larger than the minimum interval of the balance.
10 How to maintain MG-S series operation manual 10 How to maintain Take care that no dust or liquid gets in the balance. 10-1 Simple Method for Maintenance (Round pan type MG-S322) Remove the windshield. (1) Refer to “Appendix 7 Windshield assembly instructions” to remove the windshield.
MG-S series operation manual 10 How to maintain 10-2 Simple Method for Maintenance (Square pan type MG-S1501, MG-S8200) Remove the square pan. Remove the pan-base. Maintenance method. (1) Wipe dirt from the balance with dry and soft cloth. (2) In the case of heavy soil, dismount the weighing pan and/or the pan-base and clean them with a piece of cloth slightly wet with neutral detergent or solvent.
MG-S series operation manual Appendix Indication Accuracy Span Model Capacity Windshield limit Class calibration 320 g 0.01 g 0.01 g 0.03 g 320.09 g 1600 c 0.1 c 0.1 c 0.3 c 1600.9 c MG-S322 0.7 lb 0.0001 lb 0.0001 lb 0.0003 lb...
Page 102
Appendix MG-S series operation manual Item Description Display LCD with backlight 7-segment : Maximum 8-digit/Segment height up to 16.5mm 16-segment : Maximum 20-digit/Segment height up to 8.5mm Bar graph : 30-step Tare range setting Actual weight subtraction with [Tare] key...
MG-S series operation manual Appendix Appendix 4 Installation of batteries This product can operate with four AA batteries. Alkaline, manganese, Nickel-metal hydride batteries can be used. Pull out the battery case. Make sure insert batteries with the positive Put four AA batteries.
Appendix MG-S series operation manual Appendix 5 USB communication and bus power input This product can communicate/power supplied through USB. Download the USB driver on your PC. Go to the Website below and download the USB driver. http://www.star-m.jp/prjump/000082.html Install the USB driver on your PC.
MG-S series operation manual Appendix Appendix 6 Balance operation with password control function This chapter describes how to use the balance with “8-5-2 Password control”. This function is useful for setting different authority for each user/guest. Appendix 6-1 User’s authority setting Power on the balance.
Appendix MG-S series operation manual Set the functions and setting values which are intended to be fixed. Refer to “3 Functions related to the operation”, “4 Functions related to the performance”, “5 User information setting”, “6 External input/output functions” and “8 Controlling and adjustment functions”...
Index of Terms MG-S series operation manual Index of Terms actual value setting method ..23, 27, 52, 54 numeric value setting method ..27, 52, 54 adding function.......... 42 operation keys ........... 10 administrator ..........81 output conditions ........69 animal mode ..........