Summarization of Contents

Safety Notices and Recommendations
General Safety Guidelines
Covers explosive atmospheres, filter suitability, chemical limits, and enclosure usage.
Filter Maintenance and Storage
Emphasizes filter breakthrough tests, calibration, and proper storage.
Product Registration and Warranty
Warranty Extension Program
Details on how to get up to 10 years warranty via product registration and connection.
Warranty Conditions
Specifies requirements for warranty applicability, including timely registration and filter replacement.
Start-up Procedures
Calibrating the Electronic Anemometer
Step-by-step guide to calibrate the unit's electronic anemometer.
Control Module Overview
Control Module Components
Identifies key components and indicators of the control module interface.
Alarm Types and Management
Details various alarm types, their signals, events, and how to silence/reset them.
Filtration and Fan Alarms
Filtration Failure and Replacement Alarms
Describes alarms related to filter breakthrough and the need for replacement.
Fan Operation and Fault Alarms
Covers alarms for fan faults and unserviceable fan conditions.
Sensor Replacement Alerts
Alerts indicating the filter breakthrough sensor has reached its end of service life.
eGuard App Connectivity
Connectivity Ecosystem Overview
Overview of the ecosystem for simpler use and safer protection via connectivity.
eGuard App Versions and Features
Embedded Service Features
Direct PC connection for monitoring and controlling.
Mobile App Features
Web connection via mobile for monitoring.
PC App Features
Web/local connection for monitoring and controlling via PC.
Connection Options and Interface Access
Embedded Service Connection Method
Details direct connection via Ethernet port.
Mobile/Computer App Connection Method
Explains connection via Wi-Fi or network for app access.
Accessing the Administrator Interface
Steps to connect and access the unit's settings and parameters.
Administrator Interface Access Guide
Successful Interface Access Example
Shows the interface after successful login with credentials.
Troubleshooting Interface Access Issues
Guides on troubleshooting network parameter issues for access.
Network Configuration for Interface Access
Accessing Windows Network Settings
Navigating Windows settings to manage network connections.
Modifying Network Connection Properties
Steps to open the properties for the network connection.
TCP/IPv4 Network Parameter Configuration
Ethernet Properties Configuration Steps
Selecting and configuring TCP/IPv4 properties for network connection.
Finalizing Network Settings and Verifying Access
Entering Specific IP Address Parameters
Manually setting IP address, subnet mask, and default gateway.
Verifying Administrator Interface Connectivity
Re-opening the browser to confirm successful access to the interface.
Administrator Interface - Status Page
Status Page Gauge Explanations
Explains the elements and gauges on the main status page.
Detailed Status Page Items
Detailed breakdown of each numbered item on the status page.
Administrator Interface - Settings
Date and Time Configuration
How to adjust the device's date and time.
Network Configuration Options
Configuring IP, Hostname, MASK, and GW for network connection.
eGuard App Data Exchange Control
Activating or deactivating data transmission to the eGuard server.
Fan Speed Control Settings
Adjusting fan setpoints and RPM limits.
Anemometer Calibration Settings
Checking and calibrating the air face velocity measurement.
Filter Saturation Alarm Configuration
Configuring Molecode sensor type, sensitivity, and replacement dates.
Filter Expiry Date Management
Tracking filter replacement dates and remaining service life.
Detailed Settings Explained
Device Network Settings Explained
Detailed explanation of network configuration options like Mode, IP, MASK, GW.
Modifying Default Network Settings
Instructions for changing network settings from default DHCP.
eGuard App Information Transmission
How to activate/deactivate information transmission to the eGuard server.
Fan Setpoint Configuration Details
Details on setting fan speed points for optimal performance.
Anemometer Settings Procedure
Procedure for checking and setting anemometer parameters.
Filter Saturation Alarm Setup
Details on sensor type, sensitivity, and replacement for Molecode option.
Filter Replacement Date Tracking
How to input and track filter replacement dates and service life.
Event Log Management
Viewing the Event Log
Shows the device's event log with timestamps and activities.
Downloading the Event Log
Method to download the event log in CSV format.
Acetone Saturation Detection Procedure
Colorimetric Tube Saturation Test
Method using acetone, absorbing cloth, and a colorimetric tube for filter saturation testing.
Materials for Colorimetric Test
Lists necessary items like acetone, cloth, cup, hose, pump, and tube.
Gas Sampling Pump Usage Guidelines
Instructions on using gas sampling pumps with detector tubes from the same brand.
Acetone Saturation Test Setup
European Fan Connection for Test
How to connect the hose to the fan box for European models.
North American Fan Connection for Test
How to connect the hose to the fan box for North American models.
Connecting Fan Boxes to Filtration Columns
Diagrams showing how to connect fan boxes to carbon filters based on unit model.
Detector Tube Preparation and Insertion
Steps for breaking the tip of a detector tube and inserting it into the pump.
PID Device Saturation Detection
Materials for PID Device Test
Lists necessary items like acetone, cloth, cup, hose, and PID device.
PID Device Air Sampling Procedure
Procedure for air sampling and detecting acetone with a PID device.
PID Device Saturation Test Setup
European Fan Connection for PID Test
Hose connection for European models to PID sampling probe.
North American Fan Connection for PID Test
Hose connection for North American models to PID sampling probe.
Connecting Fan Boxes to PID Probe
Diagrams showing fan box connection to the PID sampling probe.
Interpreting PID Test Results for Filter Change
Guidance on changing filters if acetone concentration exceeds 5 ppm.
Molecode S Sensor Saturation Detection
Materials for Molecode S Sensor Test
Lists necessary items like acetone, absorbing cloth, and cup.
Molecode S Sensor Test Procedure
Steps to perform the saturation test using the Molecode S sensor.
Connecting to Embedded Service (Molecode)
How to connect to the unit's embedded service using the default IP address.
Molecode Sensor Settings and Status Monitoring
Adjusting Molecode Sensor Sensitivity
Instructions to set the Molecode sensor sensitivity to 'high'.
Monitoring Filtration Efficiency via Status Page
How to interpret the 'Status' page to check filtration efficiency.
Replacing the Filters
Filter Technology Configuration Options
Summarizes possible Flex™ filter configurations for Captair 321-481 and 392-714 models.
Carbon Filter Types and Applications
Summary of Carbon Filter Types
Table summarizing carbon filter types (AS, BE+, K, F) and their applications.
Replacing HEPA/ULPA Filters
HEPA/ULPA Filter Replacement Prerequisites
Conditions like user knowledge, empty lab, and ventilation before replacing filters.
Essential Protective Equipment for HEPA/ULPA
Essential PPE including overall, gloves, glasses, and mask.
Step-by-Step HEPA/ULPA Filter Replacement
Step-by-step procedure for safely replacing HEPA/ULPA filters.
The « Revolving » System for Filters
Revolving System Molecular Filter Concept
Explains the revolving system for molecular filter replacement, increasing retention capacity.
Filter Replacement Safety Recommendations
Advises on wearing safety equipment during filter replacement operations.
Fan Module Disconnection Procedures
Initial Step: Switch Off Fume Hood
Initial step to turn off the fume hood before maintenance.
Disconnecting European Fan Connection
How to disconnect the power supply and hose for European models.
Disconnecting North American Fan Connection
How to disconnect the power supply and hose for North American models.
Reassembling Filtration Columns and Fan Modules
Reassembling European Fan Connection
Reconnecting the fan module and hose for European models.
Reassembling North American Fan Connection
Reconnecting the fan module and hose for North American models.
Fan Module Connections for European Models
Standard Fan Module Connections (European)
Shows fan module connections for models 392, 483-633, and 714.
Molecode Sensor Fan Module Connections (European)
Shows fan module connections for European units with Molecode sensors.
Fan Module Connections for North American Models
Standard Fan Module Connections (North American)
Shows fan module connections for North American models.
Molecode Sensor Fan Module Connections (North American)
Shows fan module connections for North American units with Molecode sensors.
Filter Storage and Service Life Recommendations
New Filter Shelf Life and Storage Conditions
Guidelines for storing new molecular filters (flat, temperature, humidity).
Filter Types and Maximum Storage Duration
Lists storage durations for different carbon filter types (AS, BE+, BE, F, K).
HEPA/ULPA Filter Storage Guidelines
Storage recommendations for HEPA H14 and ULPA filters (upright, dry).
Predicting Molecular Filter Service Life
Factors affecting molecular filter life and annual replacement recommendations.
Erlab Safety Program and Valipass Certificate
Explains the ESP service, Valiquest questionnaire, and Valipass certificate for filter life prediction.
Filter Replacement Based on Valipass Expiry
Activated carbon filters must be replaced when the Valipass certificate period expires.
General Filter Replacement Recommendations
For unvalidated applications, replace molecular filters every 12 months and HEPA/ULPA every 36 months.
Filtration Breakthrough Sensor Replacement Frequency
Molecode Sensor Replacement Intervals
Specifies replacement intervals for Molecode S (5 years) and Molecode A/F (2 years).
Cleaning and Maintenance Procedures
Mechanical Component Checks
Inspection of hinges and acrylic parts for proper function and cleanliness.
Enclosure Cleaning Methods
Methods for regular cleaning of the enclosure using soap or pH neutral products.
Coated Metal Parts Inspection and Spill Tray
Checking coated metal parts for corrosion and spill tray for stagnant water.
About Erlab Company
Erlab's Commitment to Laboratory Safety
Erlab's history, focus on safety, and 7 ingredients for product quality.
Erlab R&D Laboratory Expertise
Commitment to understanding molecular filtration and developing safe, high-quality products.
Adherence to Strict Safety Standards
Adherence to AFNOR NF X 15-211 and ANSI Z9.5-2012 filtration safety standards.
Importance of the Published Chemical Listing
Importance of the chemical listing for compliance and retention capacity verification.
Independent Filter Testing Assurance
Independent testing guarantees adherence to AFNOR NF X 15-211 standard for emissions.
Valiquest Application Questionnaire
Laboratory specialists recommend hoods, filters, and advice based on user applications.
Chemical Validation Certificate Issuance
Certified chemist issues validation certificate for chemicals used in the hood.
Erlab's Comprehensive Safety Program
Program includes chemical evaluation, installation validation, and safety specialist support.