Introduction SP Series Pressure Transmitter The SP Series of pressure transmitters are designed to measure the level or pressure of various industrial processes. Models can be mounted with various threaded or flanged process connections. All models are available with 4..20mA (2 wire) output (or voltage outputs). The SP/FA line coms with an integrated Zero and Span adjustment.
Mounting Notes 1) Use PTFE tape or o'rings to seal the system. Transmitters with cable gland: Tighten the cable gland as much as possible so that the cable is securely in place and cannot be pulled through the cable gland. This will help insure that no moisture is able to infiltrate into the housing.
Installation Pressure of the Liquid Column Water Column System MKgfs System MKS P(mH2O)= relative density.h(m) P(Kgf/m2)= Specific Gravity(Kg/m3). h(m) P(KPa)= density(Kg/m3). g(m/s). h(m) Max. Principle of Operations 4...20mA The weight of the column of liquid Pressure generates hydrostatic pressure on the diaphragm of the sensor.
Page 12
Fig. 4 Membrane infiltrating into the transmitter (Fig 5). Flange Use a cable with vent tube for all Sitron transmitter Seal models so that the linearity of the output signal and the operation of the unit is not compromised (this special cable can also be provided by Sitron) (Fig.
Calibration SP98FA and SP96FA Before the calibration of the unit is first tested, the process should be free of pressure or (if that is not possible) the unit should first be tested outside of the vessel. Make sure the muliti-meter is correctly connected in series with the source and the instrument.
Handling Seal the thread with Teflon tape before Fig. 1 installation. Do not turn or handle by the housing (Fig. 1). When tightening the sensor, use only use the 316S.S. hexagon fitting to achieve a seal, do not twist with the body of the sensor. (Fig.
Technical Specifications SP98FA Application Pressure and Level measurement for liquids and gas 12...30Vdc Power supply Consumption 22mA max. 4...20mA (2 wires ) Output Accuracy +/- 0.5% Zero & Span Adjustment Ceramic capacitive Sensor type Range -1 to 40 Bar Electrical connection Cable gland ½”NPT BSP or NPT - flange or sanitary Process connection...
Page 16
Technical Specifications SP96FA Application Pressure and Level measurement for liquids and gas 12...30Vdc Power supply Consumption 22mA max. 4...20mA (2 wires ) Output Accuracy +/- 0.5% Zero & Span Adjustment Sensor type 316 SS piezoresisitve silicon Range -1 to 200 Bar Electrical connection Cable gland ½”NPT BSP or NPT - flange or sanitary...
Page 17
Technical Specifications SP68 Red (+) Black (-) Yellow (Out) 3 wires Application Pressure and Level measurement for liquids and gas 12...30Vdc Power supply 22mA max. Consumption Protection Reverse polarity Output 4...20mA (2 wires )/ ,1...5V, 0...5V (3 wires) Accuracy +/- 0.5% 316 SS silicon piezoresistive Sensor type Range...
Page 18
Technical Specifications SP30 DIN 43650 Connector (+) Positivo (-) Negativo Out (V) ( ) GND Application Pressure and Level measurement for liquids and gas 12...30Vdc Power supply Consumption 22mA max. 4...20mA (2 wires ) 0...4,5V (3 wires) upon request Output Reversal polarity Protection Accuracy...
Page 19
Technical Specifications SP10 Red (+) Green (-) Yellow (Out) 3 wires Application Pressure and Level measurement for liquids and gas Power supply 12...30Vdc Consumption 22mA max. Reversal polarity Protection 4...20mA (2 wires ) 0...5 e 1...5V (3 wires) Output Accuracy +/- 0.5%...
Page 20
Technical Specifications SP21 DIN 43650 connector (+) Positive SP21- IP67 SP21 SP21- IP66 36mm (-) Negative 47mm 47mm Out (V) ( ) GND Ø 30mm Ø 30mm M12 connector Ø 30mm (+) Positive Hex. 1.1/16” (-) Negative Out (V) ( ) GND Application Pressure and Level measurement for liquids and gas Power supply...
Page 21
Technical Specifications SP98 / SP96 DIN 43650 connector (+) Positive (-) Negative Out (V) ( ) GND M12 connector (+) Positive (-) Negative Out (V) ( ) GND Pressure and Level measurement for liquids and gas Application Power supply 12...30Vdc Consumption 22mA max.
Page 22
Technical Specifications SP81 M12 connector (+) Positive (-) Negative ( ) GND Application Pressure and Level measurement for liquids and gas Power supply 12...30Vdc Consumption 22mA max. 4...20mA (2 wires ) Output Protection Reverse polarity / Electromagnetic Interference +/- 0.25% Accuracy Sensor Type Piezoresistive ceramic...
Trouble Shooting Cause Solution Fail Call for service Deformed membrane Lack of Linearity Incorrect power supply Check the power supply Process temperature Temperature outside must be within stated range the operating range Output Signal Check the installation, Turbulence in the process Oscilating may need to install a stilling well...
Page 25
Cancellation: Any cancellation of the Contract by the Buyer shall be effective only if made in writing and accepted, in writing by the Sitron. In such a case, Sitron is entitled to reasonable cancellation charges including but not limited to labor, material and other...
Sitron's field technicians or Sitron's authorized Sales Representatives, in order to be covered by Sitron's warranty. Sitron shall be under no liability in respect to any defect from fair wear and tear, willful damage, negligence, abnormal working conditions, failure to follow Sitron's instructions (whether written or verbal), misuse, modification or alteration or attempted repair of the Goods without Sitron's approval.
Page 27
Sitron - Brasil 1800 Prime Place R. Baronesa de Itu, 83 Hauppauge, NY 11788 São Paulo - SP - 01231-001 PH: 516-935-8001 T .: (5511) 3825-2111 www.sitron.com FX: 800-516-1656 F .: (5511) 3825-2171 BRASIL: vendas@sitron.com USA / Other Countries: info@sitron.com...