Health Solutions for Pets
1.0 PETTEST / GENTEEL PARTS
(1)
Contact Tip
(2)
Nozzle
(3)
Nozzle Slots
(4)
Lancet
(5)
Shuttle
(6)
Front End Tabs
(7)
Front End
(8)
Activation Button
(9)
Hole In Activation Button
(10)
Vacuum Tube
(11)
Push Shaft
(12)
Vacuum Indicator
(13)
Push Cap
(14)
Piston Assembly
7
9
14
4.0 PREPARE GENTEEL FOR LANCING
PREPARATION: Clean the surface of Genteel and other testing supplies.
Verify Contact Tip is securely affixed to Nozzle.
4.1 Prime Genteel: Press the Push Cap to prime the PetTest/Genteel, device
clicks if done correctly.
4.2 Remove Nozzle: Twist counterclockwise and pull gently to remove.
4.3 Insert a Lancet: While the PetTest/Genteel is primed, insert either a
Butterfly
PetTest Maroon
lancet into Shuttle and the #1
or
start the test. Carefully twist off the protective cover from lancet and put aside.
PetTest Maroon Twist Top
(The
lancets are 6 steps more aggressive than the
Blue Butterfly Touch
lancets.)
Note: You want to use the thickest Contact Tip that gets the right amount of
blood with little to no discomfort. Contact Tips going from thickest to thinnest:
#1
Blue
#2 Yellow, #3 Green, #4 Clear, #5 Orange, #6 Violet.
NOTE: Different test sites may require a different Contact Tip.
4.4 Replace Nozzle: Line up the Nozzle Slots with the Front End Tabs.
Gently push together and rotate clockwise into place.
5.0 PERFORM BLOOD DRAW
Wash hands with soap and water. Clean test site. Have your meter and test
strip ready to wick up your blood sample.
5.1 Place Genteel on skin: Pick your test site from the diagram in 3.0 at the
top right. Make sure you have a seal between Contact Tip and skin by gently
and flatly holding the device perpendicular to the skin. Do not press Genteel
forcibly into the skin, as this can cause pain during testing.
NOTE: Hold Genteel like a pen, with index finger on the Activation Button.
4.1
4.2
4.3
4.4
The Pain-Free Blood Glucose Monitoring System
User Instructions for the Genteel
Distributed By: Diabetic Supply of Suncoast, Inc.
PO Box 2102, Vega Alta, PR 00692
1-866-373-2824 / www.ShopPetTest.com
(24 hours a day / 7 days a week)
Follow us on:
Facebook.com/PetTest
2.0 GATHER SUPPLIES NEEDED
1
With clean hands, gather supplies needed
for testing:
(A) PetTest/Genteel Lancing Device
(B) Contact Tip Card (contains 6 Contact Tips)
2
(C) Advocate Twist Top or Butterfly Touch Lancets
(D) Alcohol (not included) to clean Contact Tip
3
(E) Clean, damp cloth (not included) for
cleaning Genteel's surface
(F) Meter and test strips
4
5
6
8
10
12
11
13
5.2 Press Activation Button: Using gentle pressure, hold down the Activation
Button for at least 4 seconds, or until you see a blood drop form.
NOTE: With your finger over the Activation Button, the plunger will rise slowly
to the halfway point, but not fully extend (you should not see the red Vacuum
Indicator).
5.3 Release Activation Button: BEFORE pulling Genteel away from your skin,
lift your finger off the Activation Button. Pull Genteel away from the skin. If done
Blue
out of order, blood drop may smear.
Blue
Contact Tip to
NOTE: Hold Activation Button longer to get a larger blood drop. If no blood is
produced, choose the next thinnest Contact Tip in order depending on the lancet
and body site chosen.
5.4 Test Blood Sugar: Apply the meter's test strip to the blood drop and follow
the instructions for the test strips and meter.
5.5 Remove Lancet: Prime Genteel again (see 4.1) and remove Nozzle (see 4.2).
Carefully insert lancet needle into its protective cap. Once needle is safely in its
protective cap, remove the lancet.
5.6 Lancet Disposal: Safely dispose of the used lancet according to local laws
for Sharps disposal.
6.0 CLEANING GENTEEL
6.1 Surface cleaning: Before every use, wipe the outside of the device with a
clean damp cloth.
IMPORTANT: If the Push Shaft rises past its halfway point—while holding
the Activation Button, this may be a vacuum leak caused by dirt or other
contaminants inside Genteel. If you experience this vacuum loss symptom—
perform (6.2). If symptom persists—perform (6.3). If cleaning doesn't resolve
vacuum leak, call customer support +1 (866) 373-2824.
5.1
5.2
5.3
Lancing Device
®
STOP! If you need troubleshooting assistance or more help
with your Genteel blood draw experience,
PLEASE DO NOT WAIT TO CONTACT US.
WE ARE HERE TO HELP!
For more information visit
Email us at sales@ShopPetTest.com
Call our
24/7 support
3.0 TESTING SITES FOR DRAWING BLOOD
A
Cats have one testing site while dogs have up to six to choose from.
B
C
D
BASE OF TAIL
(shave area to
create a
sufficient seal)
E
F
NEAR NAIL
6.2 Quick Clean (inside):
6.2.1 Remove Piston Assembly: De-prime Genteel by pressing Activation
Button. Twist Piston Assembly base counter-clockwise until you can pull it out
of the Vacuum Tube.
6.2.2 O-ring Wipe: Wipe the O-ring on the Piston Assembly with your clean
finger to remove surface dirt or other contaminants.
6.3 Deep Clean (inside):
6.3.1. Remove Piston Assembly: Repeat step 6.2.1
6.3.2. Flush Vacuum Tube: Run fast moving, warm water through the
Vacuum Tube and out the Nozzle while repeatedly pressing the Activation
Button.
6.3.3. Flush Piston Assembly: While running the Piston Assembly under
fast-moving, warm water repeat 6.2.2.
6.3.4. Dry: Let Genteel air dry thoroughly. Do not reassemble Genteel with
moisture in the Vacuum Tube.
6.3.5. Reassemble: Put Piston Assembly back inside the Vacuum Tube and
twist base clockwise until it locks into place securely. Verify device functions
as intended.
5.5
5.4
5.6
www.ShopPetTest.com/Genteel
line at +1 (866)
373-2824