SVC XHR-1024i / SVC HR-1024i
SVC HR-768i / SVC HR-768si
SVC HR-640i
SVC HR-512i
USER MANUAL
Revision 1.13
29 Firemens Way, Poughkeepsie, New York 12603
Voice: (845) 471-7007, Fax: (845) 471-7020, E-mail: svcinfo@spectravista.com
Copyright © © © © 2017 Spectra Vista Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the XHR-1024i and is the answer not in the manual?

Questions and answers

Summary of Contents for SVC XHR-1024i

  • Page 1 SVC XHR-1024i / SVC HR-1024i SVC HR-768i / SVC HR-768si SVC HR-640i SVC HR-512i USER MANUAL Revision 1.13 29 Firemens Way, Poughkeepsie, New York 12603 Voice: (845) 471-7007, Fax: (845) 471-7020, E-mail: svcinfo@spectravista.com Copyright © © © © 2017 Spectra Vista Corporation...
  • Page 3: Table Of Contents

    SINGLE GRAPH ................................. 54 MULTI GRAPH ................................55 SHOW TARGET JPEGS ............................. 56 PLOT SETTINGS … ..............................57 LARGER BUTTONS ..............................59 About Menu ..............................59 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 4 PDA S – D ................... 97 OFTWARE PPLICATION ETAILED SAGE Introduction ..............................97 Data File Format ............................97 Data File Naming Convention (AutoSave) ........................97 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 5 APPENDIX X – INSTRUMENT QUALITY CHECKLIST ................. 120 APPENDIX Y – INSTRUMENT DEFAULT SETTINGS ................122 APPENDIX Z – INSTRUMENT FOV DIAGRAM(S) .................. 126 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 6: Section 1 - Spectroradiometer

    A full range of options are available. These options include alternate foreoptics, fiber optic light guides, irradiance spheres, cosine diffusers, Reflectance Probe, back packs, white reference panels, tripods and other custom engineered field devices. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 7: Hardware Description

    1/4-20 UNC threaded holes that are compatible with most conventional photographic or video tripods. Alternatively, a tripod can be mounted at the bottom of the optical plate (as shown in SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 8: Foreoptic Options

    3 item 1 for the field of view setup. Measurements can be taken at further distances but the target size will be larger. The foreoptics are mounted in lens barrels or adapters that are secured to the unit by SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 9: Battery Pack

    0 volts. The battery can then be charged via the charging connector with the supplied charger. Refer to Section 4 – Battery Charger for charging details. PULL HANDLE COVER PLATE CHARGE JACK ELECTRICAL CONTACTS Figure 3 Battery Pack SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 10: Support Feet

    The hand strap is adjustable and allows holding of the instrument with the right hand whilst actuating the laser scan switch with the right index finger. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 11: Laser Scan Switch

    The instrument has 3 connectors that may be used to communicate with external devices; two of these connections are located on the Control Panel, one is located on the bottom of the unit: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 12: Led Indicators

    (no stylus needed.) Text and icons are high contrast, designed to be read in direct sunlight conditions. Spectral plots are viewable from the display. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 13: Section 2 - Pc Data Acquisition Software

    Use Windows Explorer to explore the CD. • Double-click on the “setup.exe” file located on the CD in order to manually run the setup. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 14: Removing The Software

    Several tools are built into the software that allow SIG format data files to be merged, re-sampled, or re- processed automatically (see the Tools Menu description below.) The Data Acquisition Software’s initial screen is shown below. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 15 The Reference Scan Button causes the instrument to take and graphically display reference data. This button may also be activated by pressing either the “R” or “F1” key on the keyboard. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 16 A data point in a graph is selected by moving the cursor to the point of interest, and clicking the left mouse button on the point. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 17 NORMAL MODE: Click the right mouse button anywhere in the graph area to restore the magnification to normal. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 18: Description

    The following standard Windows file dialog box appears when “New” is selected. Note that the dialog style may vary slightly with different versions of the Windows OS. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 19 This box has no purpose other than to display instructions to the user. Open Accepts the new base file name entered. Cancel Cancels the operation. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 20: Open

    Modifies the current search pattern used for displaying the existing files in the Files Box. Only “SIG” format files may be opened by the Data Acquisition software. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 21: Save As

    Allows the desired filename for data storage to be entered. If no extension or an invalid extension is entered, then one is supplied, based on the current file format. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 22: Close

    Closes all files in the File Stack. PRINT Opens a standard Windows print dialog box. The current graph view may be printed using this dialog. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 23: Data Options

    Provides for the selection of one of the following: • SIG File This is the format compatible with the PC Data Acquisition software; spectral data saved in this SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 24: Exit

    EXIT Exits the PC Data Acquisition Software program. Control Menu The control menu item provides the following drop-down menu of instrument controls: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 25: Setup Instrument

    Configures the Computer Port used for communications with the instrument. Once connected, sets the spectroradiometer operating parameters. When “Setup Instrument…” is selected, the software presents the following setup dialog box: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 26 Note that using this option will significantly slow down spectrometer operation. Disconnect Closes the connection. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 27 Auto The “Auto” check box allows the instrument to automatically select the optimum integration time for each spectrometer. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 28: Setup Overlap/Matching

    SETUP OVERLAP/MATCHING… Configures how to handle overlapped detector data; note that this option is not available in the SVC HR-512i software: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 29 InGaAs detector data ends, and the second InGaAs detector data begins, when the “Remove Overlapped Detector Data” option is selected. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 30: Setup External Data

    (See Matching Type Frame instructions above). SETUP EXTERNAL DATA … Configures the instrument’s optional external data interface. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 31 SIG files and/or displayed in the main GUI display. When selected, the Setup External Data command displays the following dialog: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 32 The 12 digit Bluetooth address to use in this box may be found on the label attached to the bottom of the customer’s external data device (WEDI-16). SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 33 Last Scan: The most recent values/percentages taken during either a Target Scan, or an External Data Scan. The percentage is computed by dividing Last Scan by Reference. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 34: Read Memory

    Reads (downloads) the instrument’s stand-alone scans from the internal memory into the PC Data Acquisition Software. When selected, the Read Memory command displays the following dialog: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 35 0 to 999, and increment as each stand- alone is acquired in the field. Total Scans In Memory: Total number of scans currently stored in the instrument’s internal memory. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 36 Pressing this button will display a standard file open dialog box, which allows the user to select a new Folder and File Base Name. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 37: Timed Scans

    Acquire spectral data automatically using a scan timer. The user may select the type of scan (reference vs. target), how the acquired data is displayed, and whether the data is written to disk files. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 38 Reference and Target on a timed basis. Scan Display: Single Graph Target scans are displayed one at a time in a Single Graph display. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 39: Setup Gps

    Set the parameters that allow the Data Acquisition Software to access a source of NMEA V2.1 GPS data from a COM port. When selected, the Setup GPS menu item displays the following dialog: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 40 “configured” list above. COM Port Determines the COM port number for the NMEA V2.1 GPS data.. Baud Rate / Parity / Data Bits / SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 41: Setup Water Quality Interface

    SETUP WATER QUALITY INTERFACE … Set the parameters that allow the Data Acquisition Software to access water quality data from a Eureka Manta 2 Water Probe. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 42: Long Term Timer

    SIG file base name during timed acquisition. When selected, the Long Term Timer menu item displays the following dialog: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 43 Save / Read Settings To / From File These two buttons allow you to save / read daily acquisition settings to / from a SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 44 Ignore Scan Warnings Checking this box will cause the software to ignore warning that would normally cause time acquisition to halt. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 45: Tools Menu

    SIG files simultaneously. When “SIG File Merge…” is selected, the software presents the following dialog box. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 46: Sig File Overlap/Matching

    Applies the current overlap and matching settings to a set of input SIG format files; note that this option is not available in the SVC HR-512i software. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 47: White Plate Reflectance

    For example, if the input SIG file name is: “gr062606.sig”, then the output SIG file name is: “gr062606_moc..sig” WHITE PLATE REFLECTANCE… Corrects for non-ideal white (reference) plate reflectance. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 48 Begin processing the SIG files; each SIG file in turn is read and processed, and the modified data is written to an output file with the string “_wpc” (White Plate SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 49: Resample Spectral Data

    Resample the input SIG format data files, and write the resampled data to new output SIG format data files. When “Resample Spectral Data…” is selected, the software presents the following dialog box: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 50: Instrument Calibration Data

    Note that the information within these tabs will be blank if no instrument is currently connected (See the Control->Setup Instrument menu to connect to an instrument). SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 51 Write Instrument Information To File: The program will prompt the operator to save the displayed information to the PC’s hard disk. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 52 Write Wavelength Calibration Table To Instrument: The program will prompt the operator to upload the current wavelength calibration table to the connected instrument. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 53 6-character familiar name that is displayed on the instrument’s LCD. Each fore-optic calibration also has an “Optic Type” – the radiometric calibration data is either a “Radiance” SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 54: Remote Access Server

    “cal” file. Contact Spectra Vista for further information. REMOTE ACCESS SERVER… Allow remote network access to the instrument’s control and data, using a TCP/IP client/server interface. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 55 Pressing this button causes the PC Data Acquisition program to close down its server function and stop listening for client connections. If not listening, this button will be disabled. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 56: Envi Spectral Library

    SIG files with overlap removed within a single ENVI Spectral Library. When “ENVI Spectral Library…” is selected, the software presents the following dialog box: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 57 Spectral Library Description: Enter the ENVI Spectral Library description into this box. This name will become the name of the Spectral Library within the ENVI application. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 58: Sig File Wavelengths

    This tool may be used to insert a new wavelength vector into a set of existing SIG files. When “SIG File Wavelengths…” is selected, the software presents the following dialog box: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 59: Window Menu

    A single measurement’s Reference and the Target Radiance are scaled against the left Y-axis and the Percent Reflectance (Target/Reference *100) is scaled against the right Y-axis. An example of this is shown below SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 60: Multi Graph

    Displays the current list of open files together in one graph. Files can be displayed as reference radiance, target radiance or percent reflectance. The following is an example of a Multi Graph: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 61: Show Target Jpegs

    SIG file is named “jarosite.sig”, then the associated target photo must have the name “jarosite.sig.jpg” in order to be automatically displayed. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 62: Plot Settings

    Sets the relative width of the line used to graph spectral data. Use Markers If checked, each data point plotted is marked by a square box. Don’t Plot Data / SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 63 SIG files. Cancel Exits this dialog without changing settings. Apply Now Applies changes and keeps the dialog open. Applies changes and exits the dialog. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 64: Larger Buttons

    Note that your software version may be different from this example. The “PROGRAM CHANGE LOG” contains a list of recent changes to the PC Data Acquisition Software in reverse chronological order. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 65: Section 3 - Operation

    (target) should be placed to completely fill the field of view of the instrument. The sighting laser can be actuated by pressing the laser scan switch on the top of the instrument. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 66: Field Of View

    Appendix Z for inserted hardcopies of FOV diagrams supplied with your instrument, when applicable). Figure 6 Setup for FOV map CAUTION: Do NOT look into the sighting laser window or laser beam. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 67: 2. Connecting The Instrument To A Computer

    The display consists of 3 screens with the following functions: SCREEN 0 (Power-On Screen): appears for 10 seconds after power-on, and includes: INSTRUMENT MODEL/SERIAL NUMBER FIRMWARE VERSION LAST CALIBRATION DATE SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 68: Setup Button

    This field is now displaying the percent change in the currently selected external data channel when compared to the reference, in the range -99% to +99%. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 69 GENERIC STATUS / WARNING / ERROR TEXT LINE – a single line of text at the bottom of the main screen which displays status, warning and/or error messages. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 70 Touch the ERASE ALL button to permanently remove all internal stored scans; once this button is pressed, a confirmation button will appear before the scans are actually removed from memory. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 71: Main Button

    No photos are taken. Low resolution (320x240) photos are taken with each Target scan. High resolution (640x480) photos are taken with each Target scan. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 72 SATS LOCKED: Number of satellites currently being used to compute position. LATITUDE/LONGITUDE/GPS TIME: Current position and GPS reported time stamp; these fields will remain blank until a position solution has been computed. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 73 Touch DISPLAY CHAN to select the channel (1 to 8) whose real-time data values are displayed in [BRACKETS] ; this allows the operator to monitor external data values for SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 74: Save Settings Button

    Touch SAVE SETTINGS to save all external data settings to non-volatile memory. When the instrument is re-powered, the most recently saved settings are restored, including the most recently acquired dark values. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 75 115200 baud connection speed; however, this may be over-ridden by right-clicking (instead of left- clicking) the “Connect” button within the software’s Control->Setup Instrument dialog. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 76 Touch SAVE SETTINGS to save all AUX port settings to non-volatile memory. When the instrument is re-powered, the most recently saved settings are restored. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 77: Reflectance Measurement Overview

    Press and hold the LASER/SCAN button to activate the sighting laser and verify the placement of the target to be measured relative to the laser spot as required to fill the field of view. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 78: Downloading Stand-Alone Measurements

    5. PDA Data Acquisition See Appendix F – PDA Data Acquisition Software for instructions on how to use the optional PDA device to acquire spectral data. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 79: Section 4 - Battery Charger

    Do not use any DC power supply or charger other than the charger supplied with the instrument, this may cause the battery pack to malfunction. Figure 7 Battery Pack (Left), Fast-Charger (Middle), and Standard Charger (Right) SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 80: Charging

    The charger’s RED LED will be illluminated during charging; when charging is complete, the charger’s GREEN LED will be illuminated. Charging a battery from empty to full capacity will take approximately 5 hours. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 81: Appendix A1 - Hr-1024I Specifications

    Li-Ion Battery operated, 6.5 – 8.0 VDC @ 2.5A Environmental Temperature: -10C to +40C Humidity: <90%, non-condensing Cleaning Wipe down exterior with soft, dry cloth only. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 82: Appendix A2 - Hr-768I Specifications

    Li-Ion Battery operated, 6.5 – 8.0 VDC @ 2.5A Environmental Temperature: -10C to +40C Humidity: <90%, non-condensing Cleaning Wipe down exterior with soft, dry cloth only. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 83: Appendix A3 - Hr-640I Specifications

    Li-Ion Battery operated, 6.5 – 8.0 VDC @ 2.5A Environmental Temperature: -10C to +40C Humidity: <90%, non-condensing Cleaning Wipe down exterior with soft, dry cloth only. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 84: Appendix A4 - Hr-768Si Specifications

    Li-Ion Battery operated, 6.5 – 8.0 VDC @ 2.5A Environmental Temperature: -10C to +40C Humidity: <90%, non-condensing Cleaning Wipe down exterior with soft, dry cloth only. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 85: Appendix A5 - Hr-512I Specifications

    Li-Ion Battery operated, 6.5 – 8.0 VDC @ 2.5A Environmental Temperature: -10C to +40C Humidity: <90%, non-condensing Cleaning Wipe down exterior with soft, dry cloth only. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 86: Appendix A6 - Xhr-1024I Specifications

    Li-Ion Battery operated, 6.5 – 8.0 VDC @ 2.5A Environmental Temperature: -10C to +40C Humidity: <90%, non-condensing Cleaning Wipe down exterior with soft, dry cloth only. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 87: Appendix B - File Descriptions

    D1,D2,D3,D4,D D1-D8 are the most recent external dark data 5,D6,D7,D8 samples. These values are acquired when the “Dark Scan” button is pressed. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 88 0 (zero). factors= F1,F2,F3 F1/F2/F3 are the matching factors used to match Si and InGaAs1 detector data. F1 is the reference radiance matching factor. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 89 Reference/Target Values’ units depend on which calibration table or fore-optic was in use: Radiance: 10 * W/(cm² * nm * sr) Irradiance: 10 * W/(cm² * nm) RAW DN: counts SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 90 360.9 697.00 532.00 76.33 362.5 676.00 504.00 74.56 364.1 700.00 524.00 74.86 365.7 724.00 544.00 75.14 367.3 744.00 565.00 75.94 368.9 768.00 584.00 76.04 … SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 91: White Plate Reflectance Data File Format

    0.958 0.959 0.960 0.960 0.960 … 2492 0.945 2493 0.945 2494 0.944 2495 0.943 2496 0.946 2497 0.946 2498 0.947 2499 0.948 2500 0.946 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 92: Pda Software Com Port Configuration File

    If the COM ports listed in the INI file do not match the COM ports configured into the PDA operating system, then the PDA will not be able to communicate correctly. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 93: Appendix C - Laser Safety

    (1) Explanatory Label Location – Front panel near fore-optic mount. (2) Warning Label Location – Front panel near fore-optic mount. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 94: Appendix D - Pc Third Party Bluetooth Usage

    Model Bluetooth Password XHR-1024i xhr1024i HR-1024i hr1024i HR-768i hr768i HR-640i hr640i HR-768si hr768si HR-512i hr512i SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 95: Appendix E - Pc Usb Device Drivers

    Windows XP SP 2 is configured to ask before connecting to Windows Update, the screen below is shown. Select "No, not this time" from the options available and then click "Next" to proceed with the installation. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 96 Acquisition Software installation, in a folder called “USB Driver” that may be found in the PC Data Acquisition Software installation folder (typically, “C:\Program Files\Spectra Vista\HR1024i\”). Once the file path has been entered in the box, click “Next” to proceed. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 97 Windows should then display a message indicating that the installation was successful. Click "Finish" to complete the installation. This has installed the serial converter. The COM port emulation driver must be installed after this has completed. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 98 COM number in the PC Software’s Control->Setup” menu. To avoid this, try to plug the instrument into the same physical USB port on the PC each time. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 99: Removing Usb Device Drivers From The Pc

    Open the Add/Remove Programs utility located in "Control Panel\Add/Remove Programs". Select "FTDI USB Serial Converter Drivers" from the list of installed programs. Click the "Change/Remove" button, and follow the instructions to uninstall the drivers. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 100: Appendix F - Pda Data Acquisition Software

    Information regarding the PDA, outside of its specific use with the spectroradiometer may be found in the PDA manufacturer’s User Guide. PDA Requirements PDA Type: Windows Mobile Operating System: Windows Mobile 6.x Interfaces: RS-232, and/or Bluetooth virtual com port serial services. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 101: Pda Software - Quick Reference

    Host PC using the Microsoft ActiveSync software via a USB connection. The default location of SIG files on the PDA is in the folder: PDA Software 1.0: \My Documents\HR1024i Data\ PDA Software 1.1 or later: \My Documents\SVC Data\ SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 102: Pda Software Application - Detailed Usage

    On the PDA press the On/Off button. The PDA application can be started by tapping the Start button in the corner of the PDA power-on screen, and selecting the PDA program. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 103: Operation

    Menu Bar, named from left to right as “File”, “Graph”, and “Instrument”, and “Help”; and two large Command Buttons on the top labeled “Reference” and “Target” (See below.) SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 104: Command Button "Reference

    File->AutoSave option is checked. MENU BUTTON “FILE” The drop down menu list for “File” consists of the following menu items: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 105 AutoSave File Counter (ranges from 0000 to 9999.) Exit Exits the application. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 106: Menu Button "Graph

    Wavelength in nanometers(nm). Reflectance is the ratio of the target scan vector to the reference scan vector. MENU BUTTON “INSTRUMENT” The dropdown menu list for “Instrument” has the following menu items: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 107 The COM Ports listed are controlled by a PDA Software configuration file; see Appendix B – File Descriptions, PDA Software COM Port Configuration File for more information. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 108 The Settings menu is used to modify instrument data collection parameters. When selected a dialog box will appear, presenting various options (See SETTINGS FORM below.) Status This indicates the instrument connection status. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 109: Settings Form

    When Auto is checked, the instrument automatically determines the optimal integration settings for all detector arrays (Si, Sw1, and Sw2). When Auto is not checked, the operator may set individual SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 110: Menu Button "Help

    When Cancel is pressed, no changes are made to the settings. MENU BUTTON “HELP” The dropdown menu list for “Help” has the following menu items: About: Displays PDA software version information SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 111: Appendix G - Pda Usage Tips

    Figure 11 Juno T41 PDA Power Menu Screen Note: Performing a Soft Reset simply restarts the PDA operating system. It is similar to re-booting a PC. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 112: Juno T41 Hard Reset Procedure

    Warning: Hard Resetting the PDA will require re-installation of the PDA application, either from internal PDA Backup, or from CD. The Recon 400X PDA may be Hard Reset by: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 113: Pda Power Up

    The PDA Application Software’s “Instrument->Connect” menu displays a list of COM ports that are available to establish the connection to the instrument. A typical list, as displayed, is shown below: SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 114 The other COM port listed is the Bluetooth COM port. The COM port listed must correspond to the COM port configured into the PDA’s Bluetooth driver, or it will not be possible to communicate with the instrument via Bluetooth. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 115: Appendix H - Instrument Remote Access

    Example: “R 0<cr>” – acquire a normal reference scan. Example: “R 1<cr>” – acquire a special reference scan. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 116: Target Scan Command

    Error Response: "E: Parameter Value Error" – The client attempted to set a parameter outside of it’s range of valid values for this command. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 117: Optic Command

    Coadds Command that may have set the instrument’s scan timing to a fixed number of coadds. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 118: Coadds Command

    Busy Error If a new command is sent from the client before the previous command has completed, the server responds with an “E: Busy<cr><lf>” message. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 119: Appendix I - Fiber Optic Installation

    Rotate the retaining ring counterclockwise while holding the fiber optic adapter clock angle until the pin feeds into its locating hole. Hand tighten the retaining ring. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 120: Appendix J - Optional Fan Installation

    If present, remove the four “Filler Fasteners” – these occupy threaded holes that will be used later by the installed fan module. Figure 15 Instrument Optic Plate Prior To Fan Module Installation SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 121 Figure 16 Instrument Optical Plate After Fan Installation SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 122: Appendix K - Aux Port Specification

    Output Host Comm TX RS-232 PC/PDA host connection. Input Host Comm RX RS-232 PC/PDA host connection. Output Host Comm GND RS-232 PC/PDA host connection. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 123: Appendix W - Spectralon® Care / Handling

    The following pages may contain Spectralon® reference panel related information when this manual is supplied in hardcopy form along with a instrument order that includes a reference panel. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 124 A P P E N D I C E S 1 1 9 1 1 9 1 1 9 1 1 9 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 125: Appendix X - Instrument Quality Checklist

    The following pages may contain an instrument-specific quality assurance checklist for a particular serial number instrument, when this manual is supplied in hardcopy form along with a new or re-calibrated instrument. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 126 A P P E N D I C E S 1 2 1 1 2 1 1 2 1 1 2 1 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 127: Appendix Y - Instrument Default Settings

    SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 128 A P P E N D I C E S 1 2 3 1 2 3 1 2 3 1 2 3 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 129 The following pages may contain one or more instrument-specific field-of-view diagrams for a particular serial number instrument, when this manual is supplied in hardcopy form along with a new or re-calibrated instrument. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 130 A P P E N D I C E S 1 2 5 1 2 5 1 2 5 1 2 5 SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...
  • Page 131: Appendix Z - Instrument Fov Diagram(S)

    Updated Section 3 with new PC Data Appendix A6, Acquisition Software controls and Appendix J tools; added new XHR-1024i model information and fan module information. SVC XHR-1024i / HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.13...

This manual is also suitable for:

Hr-1024iHr-768iHr-768siHr-640iHr-512i