GTI Aerosol Photometer Model 3991 Aerosol Photometer 3991 Model Operation Manual Please read and understand the warning precautions specified in this Manual 01001 carefully before using the product. 23 . 03 Please keep this Manual in good condition for long-term use.
Page 2
User Instructions 1. Please read this Manual carefully before using the instrument and be familiar with relevant terms. 2. Failure to operate according to the requirements of this Manual may damage the instrument, result in fire, electric shock or even serious personal injury. The Company defines the types of warnings used in this Manual as follows.
CONTENTS Chapter I. Instrument Introduction..........................5 1.1 Product Features ............................5 1.2 Application fields ............................6 1.3 Main specifications ............................6 1.4 Packing list ..............................7 1.5 Options ................................7 Chapter II. Configuration and Accessories Introduction ....................8 2.1. Master ................................8 2.2 Probe................................9 2.3 Downstream Quick Connection Kit ......................10 Chapter III.
Page 4
5.2 Company Name Input ..........................19 5.3 ID. Number input .............................20 5. 3—1 Input ID into the host ........................20 5.3—2 ID Scanning via a probe......................20 5.4 Testing.................................20 5.5 Data printing and transmission ........................21 5.5—1 Data printing ..........................21 5.5—2 Data transmission........................22 5.6 Zero Re-establish ............................22 Chapter VI.
Page 5
Chapter VIII. Records ............................32 8.1 Viewing the Log............................33 8.2 Print Out ..............................33 8.3 Output................................33 Chapter IX. DATA ..............................34 9.1 Check DATA...............................34 9.2 Print ................................34 9.3 OUTPUT ..............................34 9.4 DELETE..............................35 Chapter X、About ..............................35 Chapter XI、Maintenance ............................35 Chapter XII、Common Trouble Shooting........................36 Chapter XIII、Warranty and After Service ......................38 13.1 Warranty..............................38 13.2 After Service .............................38...
Chapter I. Instrument Introduction Model 3991 is a linear digital aerosol photometer for scattered light with advanced technology and scientific design, and serves as the optimum test instrument for testing the integrity of the high efficiency particulate air (HEPA) filter or conducting leakage test. Considering the low concentration of atmospheric dust in the upstream of the HEPA filter under test, the aerosol generator should be used to replenish and generate dust to meet the test conditions.
The self-equipped flow monitoring device and sampling pump with an adjustable pulse width, which can achieve excellent performance; Built-in thermal printer, which can support long-term thermal paper printing, real-time reading and printing of test data; Energy saving and consumption reduction, battery powered available; ...
Build-in Li-battery 3991-0020 Charging Adaptor 3991-0030 Round sampling port 3991-0040 Square sampling port 3991-0050 Chapter II. Configuration and Accessories Introduction 2.1. Master Name Functions Remarks ② ① ③ ① Touch screen Display and operation settings ② Handle Non-folded In the summary mode, it can print ③...
Attention Installing a non-specified type of fuse can cause damage to the instrument Fuse Type: 5×20mm, 2A, 250V. Attention Please supply power according to the voltage range indicated on the panel label at the back of the instrument, thus avoiding any damage to the instrument 1) 100~120V 60Hz 0.7A 2) 220~240V 50Hz 0.35A 2.2 Probe...
mode Start/pause data Terminate data transfer/print transmission/print Parameter modification/location moving 2.3 Downstream Quick Connection Kit ① ② ③ Item. Function Air pipe Connect with the probe’s ① plug air pipe socket Small cable Connect with probe’s ② plug cable socket Large cable Connect with host’s cable ③...
Attention Before testing, the operator should check the sampling joint and ensure that there are no fluffs or impurities blocking the filter screen. During the test, if the pump noise suddenly increases, the operator should check of the gas passage is blocked, especially the filter screen at the sampling joint. It’s recommended to wipe clean the filter screen at the sampling with a piece of silk cloth before each use.
In order to ensure the accuracy of the test results, the upstream and downstream sampling air pipe should not be mixed. 3.1-3 Test Connection at the downstream (low aerosol concentration zone) Attention 1) Be sure to connect the probe’s cable connector before the instrument is powered on.
3.2 Loading printing paper into the printer Please use the printing paper that is matched with the printer, the printing paper’s specifications should be 57mm (2 in) in width. If the printing test results need to be saved for a long run, please choose the long-term thermal o-sensitive printing paper.
If the print paper roll loosen, need to be Uncover the adhesive side of the printer paper, straightened again. both sides are level with bouncing the paper off so that the paper can be the paper shaft, then the paper is tightly rolled smoothly drawn when printing.
② Keep pressing the power switch ① Power cable connecting 线 …… 3.3—2 User login If the security authentication function enabled, the user needs to log in to the user system after initialization. 3.4 Power supply mode 3.4-1 AC Power supply Specification:...
3.4-2 Power supply by battery Rechargeable Li-battery packed in the instrument may keep power supply for about 4.5 hours or so. 1)Power Charge There’s an independent charging port for power supply, please use the standard adapter supplied by us to charge.
Chapter IV. Home Page 1)Date/Time:System time display 2)Measure:Concentration and flow tests 3)Log :traceability operation record 4)Data:storage data browsing, delete, exporting and printing. 5)About:underlying information 6)User:add User name and configuration appropriate permissions. 7)Power supply mode:AC power supply or battery, charge level display, low charge alarm. 8)Setup:parameter setting, function selection.
Attention Set the upstream concentration to 100% via either method, The allowable upstream mass concentration scope of instrument for test is 5-120 μg/L, ※ When the upstream concentration is more than 120 μg/L or less than 5 μg/L, an error prompt will pop out.
5.1-2 100% concentration self-defined reference value Attention The input range of the value is 5-120. 5.1-3 Set 100% concentration based on the last measured value 5.2 Company Name Input After 100% upstream concentration is established, the operator can input Company Name into the Master and this Name can be traced with the corresponding testing data for printing, data-saving and etc.
5.3 ID. Number input Regarding to the ID number,the operator can input ID into the host, or use the probe to scan the one-dimensional code for automatic identification. If there is no need to input ID number, the operator can directly click "next step" and enter into the test interface 5.
Test interface of Master: Descriptions Icon The probe has been connected Measure scanning time Testing concentration percent (Green for Probe's audio alarm is available qualified/ red for overloading) Probe’s vibration alarm is available Testing status iron USB interface and PC are connected Functions selection Upstream sampling time constant Testing airflow...
5.5—2 Data transmission In Report mode, "Monitoring" or "Continuous" status can be selected, and when connect the USB cable on instrument back panel to PC, the testing data can be kept transferring until the end of setting time. Data transfer stop then return to the Home-page.
Chapter VI. Setting up ATTENTION 1) When setting up on the main unit, the probe will prompt "main unit setting...", and the probe cannot perform any operations until the main unit returns to the home page interface; 2) When operating on the probe, the host will prompt "Probe setting...", and the host cannot perform any operations until the probe returns to the home page interface;...
6.2—2 Coefficient Suppression ※Coefficient suppression: It is the noise suppression function. By setting the sampling time constant, the test results are displayed as the average value within this period, thereby suppressing the influence of aerosol fluctuations on the test data. The sampling time constants for upstream and downstream are set separately. The time constant ranges from 1 to 120s.
6.2—3 Digit Display ※When the leakage rate is less than 1.0%, there are two options for display digits after the decimal point: 3 digits (0. xxx%) or 4 digits (0. xxxx%) 6.2—4 Report Mode ※Report Mode: Summary, Monitor, Continuous High-efficiency filter (HEPA/ULPA) test results are verified and saved on-site, and Through the USB port, output the real- Through the USB port, the percent leak...
※Displaying the alarm is the default function which will be displayed in red if it exceeds the judgment limit, and it will be displayed in green if the qualified data is exceeded. 6.2—6 Time Format Three Modes: Year Month Date, Month Date Year, Date Month Year. 6.2—7 Date &Time ※Only by host...
6.2—10 Low Power Consumption ※Model 3991 has a built-in air pump and battery, and long-term use will reduce the service life. When the low power consumption mode is turned on, the air pump will run in a low power consumption mode, which greatly...
6.2—11 Data Storage ※The 3991 has the data storage function. When the data storage is turned on, the corresponding data will be stored according to the report mode rules. The stored data can be browsed, printed, exported and deleted in the data system.
6.2—13 Online Upgrade ※This function is used to upgrade the system patch, please use it under the technical guidance of GTI, and beware of system damage caused by random operation. Chapter VII. User ※The user system needs to enable the security authentication function to access.
7.1 Permission Setting ※The permission name is limited to 20 characters or less. ※The permission content refers to the functions that can be used under the permission name, and those that are not selected are unusable. 7.2 Registration ※The username is limited to 20 characters or less. ※The user’s name must choose a permission.
7.3 User Modify 7.4 Delete ※The admin system cannot be deleted. 7.5 Data exporting ※Connect the U disk to the USB interface on the front panel. ※The export file format is txt and the name is 3991Users.
7.6 Print 7.7 User Toggle Chapter VIII. Records ※When the security authentication function is turned on, the log function will keep all operation records, which is convenient for instrument traceability. ※The log cannot be cleared. When the storage space reaches the upper limit, it will follow the "first in, first out"...
8.1 Viewing the Log 8.2 Print Out 8.3 Output ※Connect the U disk to the USB interface on the front panel. ※The export file format is txt and the name is 3991OperationLog.
Chapter IX. DATA ※When the data storage function is turned on, you can enter the system for data viewing and processing; ※The stored data can be printed, exported, and deleted in the statistics mode and detail mode. The method is the same, and the introduction will not be repeated.
9.4 DELETE ※Deleted data cannot be recovered, please operate with caution. Chapter X. About Chapter XI. Maintenance After the 3991aerosol photometer is powered on, the internal vacuum air pump runs, and its own fan also rotates at the same time. There is a cooling fan at the rear of the chassis. On-site maintenance is limited to replacing the fuse and cleaning the sampling connector, and to ensure that the instrument does not run continuously with power for a long time, it can be used...
Run 2080 hours a year ATTENTION Users are not allowed to dismantle the shell or repair the internal components without authorization. Problems with any device inside this product must be addressed and repaired at a GTI authorized service facility. Recommended Periodic Maintenance Daily: Clean the port of the aerosol sampling connector, see 3.1-1 for details.
Page 38
The ambient temperature is lower than Ensure that the temperature of the instrument's operating 5°C or higher than 50°C environment is within the range of 5°C to 50°C Symptom: All or part of the host touch is unresponsive Symptom Possible causes / troubleshooting There is an insulating substance on the Wipe the screen with a cleaning cloth surface of the touch area...
5 years. The retention period is the repairable period, please contact our company for details. The annual calibration service is provided by GTI or a partner who has been trained and certified by the GTI service system, or by an authoritative certification body.
Page 40
GTI PRECISION INSTRUMENT (SHENYANG) CO., LTD. 吉泰精密仪器(沈阳)有限公司 Tel:+86-24-81231386 Address: Office: No. 56-39 (2212), North Huanghe Avenue, Hanggu District, Shenyang, Liaoning Province, P.R. China 110034 Factory: 58# Zhengliang 4 Rd., the New District of North Shenyang, Liaoning Province, P.R.China 110136. URL: www.gti-instruments.com...