CONTENTS CERTIFICATE OF CONFORMITY....................WARRANTY..........................GENERAL INFORMATION...................... INSTALLATION OF APPLIANCE....................4.1. Delivery 4.2. Location Precautions for use 4.4. Installation 4.5. Technical data 4.6. Construction 4.7. Loading GENERAL USE......................... 5.1. Control panel 5.2. Starting up / stopping the incubator 5.3. Security devices 5.4.
Page 3
ATTENTION: GENERAL INFORMATION AND SAFETY It is necessary to strictly follow the instructions for use of this manual to ensure the proper functioning of the appliances or to exercise any warranty claims. Using this manual: • Read this manual carefully before starting the appliance. •...
Note: These appliances are not designed to operate in explosive environments (ATEX). Moreover they cannot be used to store flam- mable, corrosive or explosive substances WARRANTY FROILABO SAS guarantees optimum operation of these appliances according to the installation and usage conditions indicated in this manual. The duration of the warranty is: 24 months.
Length (mm) door opening Height (mm) at 44°C of 30 secs (min)** Depth (mm) * Not including measure uncertainties, FROILABO procedure : 9 points Shelves (standard/max) 2/10 2/18 caracterisation according to NFX15-140 norm ** 98% of the value Weight per shelf/total (kg)
GENERAL USE Control panel temperature PXR type fault On/Off ventilation Safety temperature indicator valve thermostat regulateur adjustment 5.2. Starting up / stopping the incubator Starting up Connect the appliance to the electricity supply protected by a 30 mA differential circuit breaker. Press the On/Off knob (0/I) to start the appliance.
Low alarm The operating threshold of the low alarm can be change by means of the temperature regulator: adjusting the low alarm 1. Press the SEL key for 1 second to display parameter block no 1 . 2. Display parameter (A2-L), using the C(S2) and D(S3) keys and select by pressing SEL. 3.
Programmed stoppage Serves to stop the heating of the enclosure automatically at the end of a fixed time after starting. The incubator will operate according to the diagram below from switching on (0/I). setting programming procedure BRE Adjustment Temperature setting Slope time (optional) = 0 to 99 hour 59 min SV = setting The incubator will stop by operating the 0/I knob after the programmed TM-5 time.
5.8. Configuration of the parameters Only temperature cycle no 2 can be used for the incubators, cycle 1 is reserved for the sterilisation cycle. Adjusting temperatures, rise time and step time Allocate the following parameters for each step: - Temperature setting (SV-x), - Rise time (slope) (TMxr), - Step time (TMxs).
5.10. Configuration of a temperature cycle Press the SEL key for 3 secs; the (PVOF) programme is displayed 1. Display Su-5 (value of the 1ststep) using the C(S2) and D(S3) keys. Confirm (SEL) 2. Change the flashing value Su-5 using theC(S2) and D(S3) keys. Confirm (SEL) 3.
Step 2: Step 3: Starting the temperature cycle is manual (launched with the PROG parameter) and repeated: MOD: 1 Programming Follow the instructions on page 10, entering the values defined above. Starting the cycle Follow the instructions on page 10. > The temperature cycle is now programmed and ready to operate. Note: This temperature cycle is pre-programmed by default on refrigerated Incubators.
Any serious problem will require intervention by our Maintenance Department, or diagnosis and help by telephone. Depending on the type of contract, FROILABO undertakes to intervene within predetermined times in the event of a breakdown. To take ondenseurs à air out a maintenance contract, please send the contract request fax 9.3.
Adress Post code City YOUR REQUEST: Maintenance contract Renouvellement Number Type of device temperature brand For freezers: Emergency LN2 yes Number of devices Desired number of visits per year You already have a maintenance contract FROILABO If so, contract number...
DISPOSAL TRANSPORT : To move the appliance, you should always wear protective gloves! Two people are required to lift or carry the cabinet. Do not tip or place cabinet horizontally. Prevent unit from excessive vibrations. DISPOSAL: In case the product is to be disposed of, the relevant legal regulations are to be observed. Information on the disposal of electrical and elec- tronic devices in the European Community: Within the European Community, the disposal of electrical devices is regulated by national regu- lations based on EU Directive 2002/96/EC pertaining to waste electrical and electronic equipment (WEEE).
Page 15
Regulator PXF4: Operation and programming for incubators BP & BE...
Page 16
1. PXF4 temperature regulator S4 S1 S5 S3 S2 Functions buttons Mark Designation Selects the parameter blocks and displays the name or value of the chosen parameter Increases the value of the set temperature In case of prolonged pressure, the value increases more rapidly Switch from one parameter to another Increases the value of the set temperature.
Page 17
2. Starting up / Shutdown the incubator Commissioning: 1. Connect the device to a 230V / 50Hz / 10A + Neutral + Earth electrical network, protected by a 30mA differential circuit breaker. 2. Press the ON/OFF (O/I) button to turn on the unit. 3.
Page 18
Setting the high temperature alarm: 1. Press SEL. 2. Display AL2 parameter using (S2) et (S3) keys and select by pressing SEL. 3. Modify parameter AL2, using (S2) et (S3) keys 4. Validate by pressing SEL. 5.
Page 19
Step n°1: Draw the temperature cycle based on this diagram. Ramp: temperature ramp between two instructions. It is defined by its duration. Level: given time during which a temperature setpoint is executed. Ramp 5 Level 5 Ramp 6 Level 6 Ramp 7 Level 7 Ramp 8...
Page 20
To program the regulator: 1. From home menu, press SEL until CH1 Pid is displayed. 2. Then select CH3 using (S2) et (S3) keys. 3. Press SEL 1 time to enter the programming menu. 4. Set the PtN parameter to define the number of used segments. Press SEL, the parameter starts blinking, set the value using (S2) et (S3) keys and validate using SEL.
Designation Parameter Value Designation Parameter Value Set point 1 SV-1 37 °C Set point 3 SV-3 4°C Ramp time 1 TM-1r 1.0 (hr.min) Ramp time 3 TM-3r 1.0 (hr.min) Level time 1 TM-1S 94.0 (hr.min) Level time 3 TM-3S 22.0 (hr.min) Set point 2 SV-2 37 °C...