LABSTAC OVE13-040 Operating Manual

Thermostatic drying oven

Advertisement

OPERATING MANUAL
THERMOSTATIC DRYING OVEN
OVE13-040
labstac.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OVE13-040 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for LABSTAC OVE13-040

  • Page 1 OPERATING MANUAL THERMOSTATIC DRYING OVEN OVE13-040 labstac.com...
  • Page 2 INDEX 1. Safety and Warning signs, label explication 1.1 Safety operation and Preventive measure 1.3 Safety operation and Preventive measure 1.4 Safety operation and Preventive measure 2. Instruction (Application, Working principle, Technical parameters) 3. Product Structure 4. Installation 5. Preparation before hand 6.
  • Page 3 Do not place equipment near flammable materials and volatile substance. Otherwise it may cause explosion or fire. Do not place equipment in the area where surrounded by acidic or corrosive gas, Otherwise it may cause creepage or electric shock labstac.com...
  • Page 4 Should unplug power plug if it’s not used for long period, or it may lead to electric shock, leakage or fire because of wear and tear of insulator. Keep out of reach of children and the door unsealed if the equipment is not supervised or not used for a long period. labstac.com...
  • Page 5 2. Temp. resolution: 0.1℃ 3. Temp fluctuation range :±1.0℃(50℃~240℃); 4. Temp uniform range :2%(50℃~240℃ ); 5. Power voltage : AC 220V/50Hz; 6. Timing range: 0~99hour, 0~9999min 7. Equipment class:class I 8. Working ambient :ambient temp 10~40℃ relative humidity70% below ; labstac.com...
  • Page 6: Product Structure

    3. Product Structure Components Figure:- 1 Figure:- 2 1. incubator body 2. controller 3.door handle 4. power supply 5.outer door 6. fan cover 7. inner chamber 8 door button 9 mesh board Product Structure: Control panel labstac.com...
  • Page 7 15 water level indicator light flashes when there is a shortage of water (standby); 16 Fan indicator light: turn on when the fan is output; ℃Symbol: When the temperature unit is Celsius℃Bright, ℉Symbol: Fahrenheit℉Bright; Button Explain :Set Metal bolt; :Shift key; :Reduce the key; labstac.com...
  • Page 8: Installation

    4. Idle equipment Before setting equipment aside, empty water in the humidifier and remove internal moisture thoroughly. Be sure the inner chamber is dry and cool before closing the door . labstac.com...
  • Page 9 flashes. When the temperature runs back and forth for several times, the lamp becomes steady on, and the instrument self-setting ends (set the setting time to 0, cancel the timing, and set the timing time again after the end of the self-setting). labstac.com...
  • Page 10 -1999 measured value) Actual measured value6 (Input mercury Target value 6 -1999-9999 thermometer value is the actual measured -1999 value) Actual measured value7 (the input Target value 7 -1999-9999 mercury thermometer value is the actual -1999 measured value) labstac.com...
  • Page 11 The smaller the value of Ar, the stronger Overshing 0~200% the controller's ability to suppress suppression temperature overtot. 1 to 100 Relay output≮20s, SSR and SCR Heating cycle seconds switch≮2s, only on the heating side labstac.com...
  • Page 12 1 to 100 Relay output≮20s, SSR and SCR Heating cycle seconds switch≮2s, only on the heating side Decimal point DP=0 display resolution is 1℃; DP=1 0; 1 Set up display resolution is 0.1℃, labstac.com...
  • Page 13 Long press the set key for 5 seconds, LK=168 will appear to enter: see the following table Initial Prompt Name Set the range Explain value Temperature filtering Filter coefficient Lower range setting -50.0~400.0 Minimum temperature range setting Timing mode See Table 1 Sensor input type Do not operate standby labstac.com...
  • Page 14 0-100. Form 1:™ Time to reach the set value Power-up and zero ST Key to start 0 Deny Deny Deny 1 Deny Deny 2 Deny Deny 3 Deny 4 Yes Deny Deny 5 Yes Deny 6 Yes Deny labstac.com...
  • Page 15 When closing the door, do not slam the door to avoid damage of the locking system. Do not use corrosive solution to wipe the exterior in order to maintain the appearance of the device. Please use dry cloth or alcohol wipe to keep the inner chamber clean. labstac.com...
  • Page 16 1. Data communication failure (1)Check RS-232 port connection whether is correct. (2)Check RS-485 port output connection whether is correct. (3)Check whether the port whether is connected. 2. Data missing or fault Please check data communication equipment rate and format in accordance. labstac.com...
  • Page 17: Troubleshooting

    ·Please check if the power switch is on Screen can not display ·Please check power switch, if it is tripping operation, please check wring layout. 10. Specification Model Thermostatic Drying Oven OVE13-040 External Dimension 570×580×593 Internal Dimension 350×350×350 Effective volume...
  • Page 18: Wiring Diagram

    11. wiring diagram Labstac LLC 82 Wendell Avenue, STE 100, Pittsfield, MA, 01201, USA Email: contact@labstac.com | Website: labstac.com...