TABLE OF CONTENTS Intended Purpose ⚠Warnings and Precautions Product Overview Display Screen Before Testing Perform Blood Glucose Tests Quality Control Tests Review Test Results Data Transmission Maintenance Performance Characteristics System Troubleshooting Symbol Information Specifications Ver. 1.0 2024/08 311-0099100-186...
Dear GlucoRx Q Blue System Owner Thank you for choosing the GlucoRx Q Blue Blood Glucose Monitoring System. This manual provides important information to help you to use the system properly. Before you begin, please read the following contents thoroughly and carefully. If you have other questions regarding this product, please contact the GlucoRx Customer Care or place of purchase.
2. Do not use accessories that are not specified by the manufacturer. 3. The GlucoRx Q Blue Blood Glucose Monitoring System should only be used with GlucoRx Q Blue test strips. Using other test strips with this meter can produce inaccurate results.
Page 7
11. Proper maintenance and control solution tests are essential to the longevity of your device. If you are concerned about your accuracy of measurement, please contact GlucoRx Customer Care or place of purchase for assistance. 12. We do not recommend using this product on severely hypotensive individuals or patients in shock.
Test strips, control solution, sterile lancets and lancing device may be purchased separately. Please make sure you have those items needed for a test beforehand. If any item is missing or damaged, contact GlucoRx Customer Care for assistance. Meter 1.
Page 9
Select or change parameters. 5. Data port Download test results with a cable connection. 6. Battery Compartment 7. Test Strip Ejector Eject the used strip by pressing this ejector. 8. Set Button (S) Enter and confirm the meter settings. 9. Test Strip Slot Insert test strip here to turn the meter on for testing.
Display Screen 1. Test Result 2. High/Low Indicator 3. Error Message 4. Measuring Mode Gen – any time of day AC – before meal PC – after meal 5. Memory Mode 6. QC mode QC – control solution test 7. Date 8.
Page 11
Entering the Setting Mode Start with the meter off (no test strip inserted). Press S. The following settings are set up in sequence: 1. Date With the YEAR / MONTH / DAY flashing in sequence, press M to select the number. Press S. 2.
depth of penetration. 6. Pull the cocking control back until it clicks. You will see a colour change inside the release button when it is ready. If it does not click, the device may have been cocked when the lancet was inserted. Alternative Site Testing (AST) Testing on sites other than the fingertip is referred to as alternative site testing (AST).
Page 13
another drop of blood and use it as the sample for the test. Be careful not to smear the blood sample. 5. Apply the sample: Gently apply the drop of blood to the absorbent hole of the test strip at a tilted angle. Make sure the absorbent hole of the test strip comes into contact with the blood sample but not the skin.
The default target range of blood glucose in each mode Gen: 3.8 to 6.6 mmol/L (70 to 119 mg/dL) AC: 3.8 to 7.1 mmol/L (70 to 129 mg/dL) PC: 3.8 to 9.9 mmol/L (70 to 179 mg/dL) Quality Control Tests If you are concerned that the meter or test strips are not working properly, you can check the performance of the meter, test strips and your technique by comparing the control solution results with the range...
Control solution that is too warm or too cool. Failure to shake the control solution vial vigorously. If you are still unsure of the problem, please contact GlucoRx Customer Care for assistance. Review Test Results To enter memory mode, start with the meter switched off.
You can use the Health Care Management System to view test results on a Windows 10 (or above) computer. Connect to a computer via cable, Bluetooth and visit GlucoRx's website at www.glucorx.co.uk to find the appropriate Health Care Management System for this meter.
Maintenance Meter Storage and handling ►Always store or transport the meter in its original storage case. ►Avoid dropping the meter or heavy impact. ►Avoid direct sunlight or high humidity. ►Store the meter at -20°C to 60°C (-4°F to 140°F), 10% to 93% R.H. (non-condensing).
2°C to 30°C (36°F to 86°F), 10 to 90% R.H. Do not freeze. Performance Characteristics For more information, please refer to the test strip insert. System Troubleshooting If you follow the recommended action but the problem persists, please call GlucoRx Customer Care. Error Messages MESSAGE WHAT IT MEANS WHAT TO DO Replace the battery The battery is too low.
Page 19
E-0, E-A Problem with the strip. If the meter still does not E-E, E-C meter. work, please contact GlucoRx Customer Care for assistance. Review the instructions and Test strip is removed repeat test with a new strip. If while counting down,...
Perform the test with a proper stored warm or too cold. control solution. Defective test strip. Repeat the test with a new test strip. Please contact GlucoRx Customer Meter malfunction. Care. Improper working of meter Please contact GlucoRx Customer and test strip.
Authorised Consult instructions for representative in use or consult electronic the United instructions for use Kingdom Authorised CE Mark representative in the European Community Do not use if package is damaged and consult instructions for use This device does not belong to household waste and must be returned to a collection point for recycling electric and electronic devices according to local laws.
Page 25
TD-4125 TaiDoc Technology Corporation B1-7F, No.127, Wugong 2nd Rd., Wugu Dist., 24888 New Taipei City, Taiwan www.taidoc.com GlucoRx Ltd. Unit 1C Henley Business Park, Pirbright Road, Guildford, Surrey, GU3 2DX, UK www.glucorx.co.uk MediMap Ltd 2 The Drift, Thurston, Suffolk, IP31 3RT, UK For self-testing.